Specifications

RIGOL 2 Command System
2-264 DS2000 Programming Guide
2-264
:MEASure:CREGion:CBX
Syntax
:MEASure:CREGion:CBX <val>
:MEASure:CREGion:CBX?
Description
When the measurement range is set to the cursor region, use this command to set
the position of cursor B.
Query the current position of cursor B.
Parameter
Name Type Range Default
<val> Integer
(The current position of cursor A +
6) to 697
400
Explanation
You can use the
MEASure:AREA command to set the measurement range to the
cursor region.
For the position of cursor A, refer to the
:MEASure:CREGion:CAX command.
Return Format
The query returns an integer.
Example
:MEASure:CREGion:CBX 600
The query returns 600.