Specifications

RIGOL 2 Command System
2-260 DS2000 Programming Guide
2-260
:MEASure:SETup:MID
Syntax
:MEASure:SETup:MID <value>
:MEASure:SETup:MID?
Description
Set the middle value of threshold measurement and the unit is %.
Query the current middle value of threshold measurement.
Parameter
Name Type Range Default
<value> Integer 6 to 94 50
Explanation
The threshold is the vertical level (in percentage) being measured in the analog
channel. Setting the threshold will affect all the time, delay and phase measurement
parameters.
You can use the
:MEASure:SETup:TYPE command to select the threshold
measurement setting.
The middle value set must be lower than the upper limit currently set (refer to the
MEASure:SETup:MAX command) and greater than the lower limit currently set (refer
to the
MEASure:SETup:MIN command).
Return Format
The query returns an integer between 6 and 94.
Example
:MEASure:SETup:MID 94
The query returns 94.