Specifications
2 Command System RIGOL
DS2000 Programming Guide
2-143
:CURSor:MANual:TUNit
Syntax
:CURSor:MANual:TUNit <unit>
:CURSor:MANual:TUNit?
Description
Set the horizontal unit in manual cursor measurement.
Query the current horizontal unit in manual cursor measurement.
Parameter
Name Type Range Default
<unit> Discrete {SECond|HZ|PERCentage|DEGRee} SECond
Explanation
SECond: when this unit is selected, in the measurement results, CurA, CurB and △X
are in s and 1/△X is in Hz.
HZ: when this unit is selected, in the measurement results, CurA, CurB and △X are in
Hz and 1/△X is in s.
PERCentage: when this unit is selected, in the measurement results, CurA, CurB and
△X are in %.
DEGRee: when this unit is selected, in the measurement results, CurA, CurB and △X
are in °.
Return Format
The query returns SEC, HZ, PERC or DEGR.
Example
:CURSor:MANual:TUNit DEGRee
The query returns DEGR.