Specifications
RIGOL
DS2000 Quick Guide
32
5. No display after pressing RUN/STOP:
Check if the MODE at the trigger panel (TRIGGER) is on “Normal” or “Single”
and if the trigger level exceeds the waveform range. If yes, set the trigger level
to the middle or set the MODE to “Auto”.
Note: using AUTO could automatically finish the above setting.
6. The display of waveform is ladder-like:
(1) The horizontal time base might be too low. Increase the horizontal time
base to increase the horizontal resolution and improve the display.
(2) If Type is “Vectors”, the lines between the sample points may cause
ladder-like display. Set Type to “Dots” to solve the problem.
7. Fail to connect PC or PictBridge through USB:
Check the IO Setting menu in Utility to make sure whether the setting in USB
Device matches the device currently connected. If needed, restart the
oscilloscope.
8. The USB storage can not be recognized:
(1) Check if the USB storage can work normally.
(2) Make sure that the USB storage being used is flash type. This oscilloscope
does not support hardware type USB storage.
(3) Make sure if the capacity of the USB storage is too large. It is recommended
that the capacity of the USB storage being used with this oscilloscope is no
larger than 8 GBytes.
(4) Restart the instrument and then insert the USB storage to check it.
(5) If the USB storage still can not be used normally, please contact RIGOL.