Specifications

RIGOL
DS2000 Quick Guide
16
CLEAR
Press this key to clear all the waveforms on the screen. If
the oscilloscope is in RUNstate, new waveforms will
still be displayed.
RUN/STOP
Press this key to set the
state of the oscilloscope to
RUNor STOP.
In RUNstate, the key is illuminated in yellow.
In STOPstate, the key is illuminated in red.
SINGLE
Press this key to set the trigger mode to Single. In
single trigger mode, press FORCE to generate a trigger
signal immediately.
AUTO
Press this key to enable the waveform auto setting
function. The oscilloscope will automatically adjust the
vertical scale, horizontal time base and trigger mode
according to the input signal to realize optimum
waveform display. Note: auto setting requires that the
frequency of the signal under test should be no lower
than 50 Hz, the duty cycle be greater than 1% and the
amplitude be at least 20 mVpp. If the parameters exceed
these limits, Auto failed !would be displayed after
pressing this key and the quick parameter measurement
menu might not be displayed.