RIGOL Quick Guide DS2000 Series Digital Oscilloscope Jul.
RIGOL Guaranty and Declaration Copyright © 2012 RIGOL Technologies, Inc. All Rights Reserved. Trademark Information RIGOL is a registered trademark of RIGOL Technologies, Inc. Publication Number QGA13004-1110 Notices RIGOL products are protected by patent law in and outside of P.R.C. RIGOL reserves the right to modify or change parts of or all the specifications and pricing policies at company’s sole decision. Information in this publication replaces all previously corresponding material.
RIGOL Safety Requirement General Safety Summary Please review the following safety precautions carefully before putting the instrument into operation so as to avoid any personal injuries or damages to the instrument and any product connected to it. To prevent potential hazards, please use the instrument only specified by this manual. Use Proper Power Cord. Only the power cord designed for the instrument and authorized by local country could be used. Ground The Instrument.
RIGOL Avoid Circuit or Wire Exposure. Do not touch exposed junctions and components when the unit is powered. Do Not Operate With Suspected Failures. If you suspect damage occurs to the instrument, have it inspected by qualified service personnel before further operations. Any maintenance, adjustment or replacement especially to circuits or accessories must be performed by RIGOL authorized personnel. Keep Well Ventilation. Inadequate ventilation may cause increasing of temperature or damages to the device.
RIGOL Safety Terms and Symbols Terms in this Manual. These terms may appear in this manual: WARNING Warning statements indicate the conditions or practices that could result in injury or loss of life. CAUTION Caution statements indicate the conditions or practices that could result in damage to this product or other property. Terms on the Product. These terms may appear on the Product: DANGER WARNING CAUTION indicates an injury or hazard may immediately happen.
RIGOL Measurement Category Measurement Category DS2000 series digital oscilloscopes can make measurements in Measurement Category I. WARNING This oscilloscope can only be used for measurements within its specified measurement categories. Measurement Category Definitions Measurement category I is for measurements performed on circuits not directly connected to MAINS. Examples are measurements on circuits not derived from MAINS, and specially protected (internal) MAINS derived circuits.
RIGOL Ventilation Requirement This oscilloscope uses fan to force cooling. Please make sure that the air intake and exhaust areas are free from obstructions and have free air. When using the oscilloscope in a bench-top or rack setting, provide at least 10 cm clearance beside, above and behind the instrument for adequate ventilation. WARNING Inadequate ventilation may cause temperature increase which would damage the instrument.
RIGOL Working Environment Temperature Operating: 0℃ to +50℃ Non-operating: -20℃ to +70℃ Humidity Under +35℃: ≤90% relative humidity +35℃ to +40℃: ≤60% relative humidity WARNING To avoid short circuit inside the instrument or electric shock, please do not operate in humid environment. Altitude Operating: less than 3 km Non-operating: less than 15 km Installation (overvoltage) Category This product is powered by mains conforming to installation (overvoltage) category II.
RIGOL Pollution Degree Degree 2 Pollution Degree Definitions Pollution degree 1: No pollution or only dry, non-conductive pollution occurs. The pollution has no influence. For example: a clean room or air-conditioned office environment. Pollution degree 2: Normally only dry, non-conductive pollution occurs. Occasionally a temporary conductivity caused by condensation may occur. For example: general indoor environment.
RIGOL General Care and Cleaning General Care: Do not store or leave the instrument at places where the instrument will be exposed to direct sunlight for long periods of time. Cleaning: Clean the instrument regularly according to its operating conditions. To clean the exterior surface: 1. Disconnect the instrument from all power sources. 2. Clean the loose dust on the outside of the instrument with a lint- free cloth (with mild detergent or water). When cleaning the LCD, take care to avoid scarifying it.
RIGOL Environmental Considerations The following symbol indicates that this product complies with the applicable European Union requirements according to Directives 2002/96/EC on waste electrical and electronic equipment (WEEE). Product End-of-Life Handling The equipment may contain substances that could be harmful to the environment or human health.
RIGOL Contents Guaranty and Declaration .........................................................................I Safety Requirement ................................................................................ II General Safety Summary ........................................................................... II Safety Terms and Symbols ....................................................................... IV Measurement Category ......................................................................
RIGOL Print ................................................................................................ 19 User Interface......................................................................................... 20 To Use the Security Lock .......................................................................... 24 To Use the Rack Mount Kit ....................................................................... 25 Kit Parts List..........................................................................
RIGOL Quick Start General Inspection 1. Inspect the shipping container for damage. Keep the damaged shipping container or cushioning material until the contents of the shipment have been checked for completeness and the instrument has passed both electrical and mechanical tests. The consigner or carrier shall be liable for the damage to instrument resulting from shipment. RIGOL would not be responsible for free maintenance/rework or replacement of the unit. 2. Inspect the instrument.
RIGOL Appearance and Dimensions 2 Figure 1 Front View Unit: mm Figure 2 Side View Unit: mm DS2000 Quick Guide
RIGOL To Prepare for Operation To Adjust the Supporting Legs Adjust the supporting legs properly to use them as stands to tilt the oscilloscope upwards for stable placement of the instrument as well as easier operation and observation of the instrument.
RIGOL To Connect to AC Power Supply This oscilloscope can accept 100-240 V, 45-440 Hz AC power supply. Please use the power cord supplied with the accessories to connect the oscilloscope to the power source as shown in the figure below. At this point, the oscilloscope is energized and the power key at the lower-left corner of the front panel is in breath state.
RIGOL Power-on Inspection When the oscilloscope is energized, press the power key at the lower-left corner of the front panel to turn on the oscilloscope. During the start-up process, the oscilloscope performs a series of self-test items and you can hear the sound of relay switching.
RIGOL To Connect the Probe RIGOL provides passive probes for the DS2000 series oscilloscopes. For detailed technical information of the probes, please refer to corresponding Probe User’s Guide. The following are the probes recommended for use with this oscilloscope. Model Description RP3300 350 MHz, passive probe, standard RP3500A 500 MHz, passive probe, optional Connect the Probe: 1. Connect the BNC terminal of the probe to a channel BNC connector of the oscilloscope at the front panel. 2.
RIGOL Function Inspection 1. Press Storage Default to restore the oscilloscope to its default configuration. 2. Connect the earth alligator clip of the probe to the “Ground Terminal” under the probe compensation signal output terminal. 3. Use the probe to connect the input terminal of CH1 of the oscilloscope and the “Compensation Signal Output Terminal” of the probe. Compensation Signal Output Terminal Ground Terminal Figure 6 To Use the Compensation Signal 4. Press AUTO. 5.
RIGOL Tip The signal output from the probe compensation connector can only be used for probe compensation adjustment and can not be used for calibration.
RIGOL Probe Compensation When the probes are used for the first time, you should compensate the probes to match the input channels of the oscilloscope. Non-compensated or poorly compensated probes may cause measurement inaccuracy and error. The probe compensation procedures are as follows: 1. Perform step 1, 2, 3 and 4 of “Function Inspection”. 2. Check the displayed waveforms and compare them with the following figures.
RIGOL Front Panel Overview 1 3 2 11 12 13 14 15 4 16 5 6 17 18 7 19 8 9 10 20 Figure 9 Front Panel Overview No. 10 Description No.
RIGOL Rear Panel Overview 1 7 6 2 5 3 4 Figure 10 Rear Panel Overview 1. Handle Pull up the handle vertically for easy carrying of the instrument. When you do not need the handle, press it down. 2. LAN Connect the instrument to the network via this interface for remote control. This oscilloscope conforms to the LXI-C class instrument standards and can quickly build test system with other instruments.
RIGOL 4. Trigger Out The oscilloscope outputs a signal that can reflect the current capture rate of the oscilloscope at each trigger via this interface. 5. Lock Hole You can lock the instrument to a fixed location using the security lock (please buy it yourself) via the lock hole. 6. AC Power Socket AC power input terminal. The power requirement of this oscilloscope is 100-240 V, 45-440 Hz, CAT II. Use the power cord provided with the accessories to connect the instrument to AC power.
RIGOL Front Panel Function Overview VERTICAL CH1, CH2: analog input channels. The 2 channels are marked by different colors which are also used to mark both the corresponding waveforms on the screen and the channel input connectors. Press any key to open the corresponding channel menu and press again to turn off the channel. MATH: press this key to open the math operation menu under which add, subtract, multiply, divide, FFT, logic and advanced operations are provided.
RIGOL Decode1, Decode2: decoding function keys. Press the corresponding key to open the decoding function menu. DS2000 supports parallel decoding and protocol decodings. HORIZONTAL MENU: press this key to open the horizontal control menu under which to turn on or off the delayed sweep function, switch between different time base modes, switch between “Coarse” and “Fine” adjustment of scale as well as modify the horizontal reference setting. HORIZONTAL SCALE: modify the horizontal time base.
RIGOL TRIGGER MODE: press this key to switch the trigger mode to Auto, Normal or Single and the corresponding state backlight of the current trigger mode would be illuminated. LEVEL: modify the trigger level. TRIGGER Turn clockwise to increase the level and turn counterclockwise to reduce the level. During the modification, the trigger level line would move up and down and the value in the trigger level message box (e.g. ) at the lower-left corner of the screen would change accordingly.
RIGOL CLEAR Press this key to clear all the waveforms on the screen. If the oscilloscope is in “RUN” state, new waveforms will still be displayed. RUN/STOP Press this key to set the state of the oscilloscope to “RUN” or “STOP”. In “RUN” state, the key is illuminated in yellow. In “STOP” state, the key is illuminated in red. SINGLE Press this key to set the trigger mode to “Single”. In single trigger mode, press FORCE to generate a trigger signal immediately.
RIGOL Knob Adjust waveform brightness: In non-menu-operation mode (menu is hidden), turn this knob to adjust the brightness of waveform display. The adjustable range is from 0% to 100%. Turn clockwise to increase the brightness and counterclockwise to reduce. Press down this knob to reset the brightness to 50%. You can also press Display WaveIntensity and use the knob to adjust the waveform brightness.
RIGOL MENU Measure: press this key to open the measurement setting menu. You can set the measurement setting, all measure, statistic function etc. Press MENU at the left of the screen to open the measurement menus of 22 waveform parameters. Then, press down the corresponding menu softkey to quickly realize one-key measurement and the measurement result will be displayed at the bottom of the screen.
RIGOL Record Stop Play/Pause Record Record: press this key to start recording the waveform. The backlight is illuminated in red. Besides, when record constant on (Open) is enabled, the backlight will also be illuminated. Play/Pause: in stop or pause state, press this key to play back the waveform and press again to pause the play. The backlight is illuminated in yellow. Stop: press this key to stop the waveform being recorded or being played back. The backlight is illuminated in orange.
RIGOL User Interface DS2000 provides 8.0 inches, WVGA (800*480) 160,000 color TFT LCD. What is worth mentioning is that the 14-grid ultra-wide screen enables you to view “longer” waveform. 1 2 3 12. 12 4 13 5 6 7 1414. 8 15 9 10 11 16. 16 Figure 11 User Interface 1. Auto Measurement Items Provide 12 horizontal (HORIZONTAL) and 12 vertical (VERTICAL) measurement parameters. Press the softkey at the left of the screen to activate the corresponding measurement item.
RIGOL complies with the color of the channel. 3. Status Available states include RUN, STOP, T’D (triggered), WAIT and AUTO. 4. Horizontal Time Base Represent the time per grid on the horizontal axis on the screen. Use HORIZONTAL SCALE to modify this parameter. The range available is from 2.000 ns to 1.000 ks. 5. Sample Rate/Memory Depth Display the current sample rate and memory depth of the oscilloscope. Use HORIZONTAL SCALE to modify this parameter. 6.
RIGOL 12. CH1 Vertical Scale Display the voltage value per grid of CH1 waveform vertically. Use VIRTICAL SCALE to modify this parameter. The following labels will be displayed according to the current channel setting: channel coupling (e.g. ) and bandwidth limit (e.g. ). 13. CH2 Vertical Scale Display the voltage value per grid of CH2 waveform vertically. Use VIRTICAL SCALE to modify this parameter.
RIGOL Note: the following direction keys might appear in the grid at the lower-left corner of the menu bar: Denote that you can open the next page menu. Denote that you can open the previous page menu.
RIGOL To Use the Security Lock If needed, you can use the security lock (please buy it yourself) to lock the oscilloscope to a fixed location. The method is as follows, align the lock with the lock hole and plug it into the lock hole vertically, turn the key clockwise to lock the oscilloscope and then pull the key out. Security Lock Hole Figure 12 To Use the Security Lock Note: please do not insert other articles into the security lock hole to avoid damaging the instrument.
RIGOL To Use the Rack Mount Kit This oscilloscope can be mounted in a standard 19-inch rack cabinet that conforms to the Electric Industrial Association (EIA) standard. If needed, please buy and install the corresponding option. Kit Parts List No.
RIGOL Installation Tool PH2 Phillips Screwdriver (recommended). Installation Space The following requirements must be fulfilled by the machine cabinet in which the instrument is mounted. The machine cabinet must be a standard 19-inch one. At least 6U (222.2 mm) height should be provided by the machine cabinet. The depth inside the machine cabinet should not be less than 400 mm. The dimensions of the instrument after being installed are shown below (the unit is mm).
RIGOL Installation Procedures Only authorized operators can execute the installation operation. Improper installation might result in damage of the instrument or incorrect installation of the instrument on the rack. 1. Install the side plates: fix the side plates (left and right) onto the bottom plate of the rack using four M4 screws. 2. Install the front plate: fix the front plate onto the side plates (left and right) of the rack using six M4 screws.
RIGOL 3. Place the instrument: place the instrument by aligning it with the edges of the side plates (left and right) and front plate of the rack and fix the instrument onto the bottom plate through presser feet using four M4 screws. 4. Load into the machine cabinet: mount the rack with the instrument fixed to it into a standard 19-inch machine cabinet with four M6 screws and four M6 square nuts.
RIGOL 5. Post-installation notice: the rack occupies a height of 5U. The holes pointed out by the arrows are installation holes.
RIGOL To Use the Built-in Help System The help system of this oscilloscope provides instructions for all the function keys (including menu keys) at the front panel. Press Help to open the help interface and press again to close the interface. The help interface mainly consists of two parts. The left is “Help Options” and you can use “Button” or “Index” mode to select. The right is “Help Display Area”. Help Options Help Display Area Figure 13 Help Information Button: Default mode.
RIGOL Troubleshooting The commonly encountered failures and their solutions are listed below. When you encounter those failures, please solve them following the corresponding steps. If the problem remains still, please contact RIGOL and provide your device information (acquisition method: Utility System System Info). 1. The screen is still dark (no display) after power on: (1) Check if the power is correctly connected. (2) Check whether the fuse is burned out.
RIGOL 5. No display after pressing RUN/STOP: Check if the MODE at the trigger panel (TRIGGER) is on “Normal” or “Single” and if the trigger level exceeds the waveform range. If yes, set the trigger level to the middle or set the MODE to “Auto”. Note: using AUTO could automatically finish the above setting. 6. The display of waveform is ladder-like: (1) The horizontal time base might be too low. Increase the horizontal time base to increase the horizontal resolution and improve the display.