User manual
RIGOL
© 2008 RIGOL Technologies, Inc.
User‟s Guide for DS1000B series
2-82
Waveform Recorder
Waveform recorder records input waveforms from CH1, CH2, CH3 and CH4 with a
maximum record length of 1000 frames. This performance can also be activated by
the Pass/Fail test output, which makes this function especially useful to capture
abnormal signals in long term without keeping an eye watching it.
Press Utility→Record→Mode→Record to go to the following menu
Waveform recorder: Record the waveforms with specified interval.
Figure 2-104 Table 2-69 The Record menu (Page 1/2)
Menu
Settings
Comments
Mode
OFF
Record
Play back
Storage
Select record mode.
Select play back mode.
Select storage mode.
Turn off all recorder functions.
Source
CH1, CH2
CH3, CH4
P/F-OUT
Select record source channel.
End Frame
<1-1000>
Set number of record frames.
Operate
(Run)
(Stop)
Record stopped, press to Start
recording.
Press to stop recording.
Figure 2-105 Table 2-70 The Record menu (Page 2/2)
Menu
Settings
Comments
Interval
<1.00ms-1000s>
Set time interval between
record frames.