User’s Guide RIGOL Publication number UGA04107-1210 July 2008 DS1000B Series Digital Oscilloscopes DS1062/4B,DS1102/4B,DS1202/4B © 2008 RIGOL Technologies, Inc.
RIGOL © 2008 RIGOL Technologies, Inc. All Rights Reserved. RIGOL products are protected by patent law in and outside of P.R. China. Information in this publication replaces all previously corresponding material. RIGOL Technologies, Inc. reserves the right to modify or change part of or all the specifications and pricing policies at company‟s sole decision. NOTE: RIGOL is registered trademark of RIGOL Technologies, Inc. © 2008 RIGOL Technologies, Inc.
RIGOL Safety Notices Review the following safety precautions carefully to avoid any personal injuries or damages to the instrument and any products connected to it. To avoid potential hazards, please use the instrument as specified by this manual only. The instrument should be serviced by Authorized personnel only. To Avoid Fire or Personal Injury. Use Proper Power Cord. Use the power cord designed for the instrument and authorized in your country only. Ground The Instrument.
RIGOL Do not Operate in an Explosive Atmosphere. Keep Product Surfaces Clean and Dry. The disturbance test of all the models meet the limit values of A in the standard of EN 61326: 1997+A1+A2+A3, but can’t meet the limit values of B. Measurement Category The DS1000B series Digital Oscilloscope is intended to be used for measurements in Measurement Category I. Measurement Category Definitions Measurement Category I is for measurements performed on circuits not directly connected to MAINS.
RIGOL Safety Terms and Symbols Terms in This Guide. These terms may appear in this guide: ! WARNING: Warning statements identify conditions or practices that could result in injury or loss of life. ! CAUTION: Caution statements identify conditions or practices that could result in damage to this product or other property. Terms on the Product: These terms may appear on the product. DANGER indicates an injury hazard may be immediately accessible.
RIGOL General-Purpose Oscilloscopes RIGOL DS1000B series digital oscilloscopes offer exceptional waveform viewing and measurements in a compact, lightweight package. The DS1000B series is ideal for production test, field service, research, design, education and training involving applications of analog circuit tests and troubleshooting.
RIGOL Contents Safety Notices ...........................................................................................II General-Purpose Oscilloscopes ................................................................... V Chapter 1 : Getting Started ................................................................. 1-1 The Front Panel and User Interface ......................................................... 1-2 To Inspect the Instrument ..................................................................
RIGOL Pass/Fail ........................................................................................ 2-76 Print Setting ................................................................................... 2-80 Waveform Recorder ........................................................................ 2-82 System information ........................................................................ 2-86 Language ......................................................................................
RIGOL Chapter 1 : Getting Started This chapter covers the following topics: The front panel and user interface To inspect the instrument To perform a functional check To compensate the probes To display a signal automatically To understand the vertical system To understand the horizontal system To trigger the oscilloscope To understand the quick function © 2008 RIGOL Technologies, Inc.
RIGOL The Front Panel and User Interface Being familiar with the front panel of an oscilloscope is the first priority when you get a new type digital oscilloscope. This chapter will bring you a brief introduction and description of the front panel operations of DS1000B series digital oscilloscopes. It is a great help for your understanding to the layout of the knobs and keys and how to use them. Read the chapter carefully before further operations.
RIGOL Figure 1-2 Front Panel Controls Notation definitions in this User’s Guide: Throughout this guide, notation symbols of buttons and knobs are the same of those on front-panel. A box around the name of the key denotes MENU function buttons on front-panel, such as Measure. ( ) denotes the multi-function knob .
RIGOL Location of the current waveform in the memory Running status Trigger position in the memory Trigger position in current waveform windows Trigger symbol Mark of channel1 Menu Mark of channel2 Mark of channel3 Mark of channel4 Waveform display windows Figure 1-3 Display screen 1-4 User‟s Guide for DS1000B series © 2008 RIGOL Technologies, Inc.
RIGOL To Inspect the Instrument After receiving a new DS1000B series oscilloscope, please inspect the instrument as follows: 1. Inspect the shipping container for damage. Keep the damaged shipping container or cushioning material until the contents of the shipment have been checked for completeness and the instrument has been checked mechanically and electrically. 2. Check the accessories. Accessories supplied with the instrument are listed in "Accessories" in this guide.
RIGOL To Perform a Functional Check Perform this quick functional check to verify that the instrument is operating correctly. 1. Turn on the instrument. Use the power cord designed for the oscilloscope only. Use a power source that delivers 100 to 240 V, 45Hz to 440Hz. Turn on the instruments, and wait until the display shows the waveform window. Push the Storage button, select Storage in the top menu box and push the Factory menu box, then continue pressing Load.
RIGOL Probe compensator Figure 1-5 Attach the probe ② Set the probe attenuation of the oscilloscope to 10X. To do this, push CH1→Probe→10X. Probe scale Figure 1-6 Set the probe Figure 1-7 The CH1 menu ③ Push the AUTO button. Within a few seconds, a square wave will be displayed on the screen. ④ Push the OFF button or push the CH1 button again to turn off Channel 1. Push the CH2, CH3, CH4 buttons to turn on channel 2, channel3, channel4, repeat steps 2 and 3.
RIGOL To Compensate the Probes Perform this adjustment to match the characteristics of the probe and the channel input. This should be performed whenever attaching a probe to any input channel at the first time. 1. From CH1 menu, set the Probe attenuation to 10X (press CH1→Probe→10X). Set the switch to 10X on the probe and connect it to CH1 of the oscilloscope. When using the probe hook-tip, inserting the tip into the hook-tip firmly to ensure a tight connection.
RIGOL To Display a Signal Automatically The DS1000B series oscilloscope has an automated feature to display the input signal in best-fit status. The input signal should be 50 Hz or higher and a duty cycle is greater than 1%. Press the AUTO button, the oscilloscope automatically sets up VERTICAL, HORIZONTAL and TRIGGER controls to display the input signal. Adjust the controls manually to get the best results if necessary. Using the automatic setting: 1.
RIGOL To Understand the Vertical System As shown in Figure 1-9, there are several series buttons and knobs in the VERTICAL control area. The following exercises will gradually conduct you to be familiar with the using of the vertical parameters settings. Figure 1-9 The vertical control window 1. Center the signal on the display with the knob. The knob moves the signal vertically, and it is calibrated.
RIGOL Vertical offset back to 0 shortcut key Turn the knob to change the vertical display position of channel and press the knob to set the vertical display position back to 0 as a shortcut key, this is especially helpful when the trace position is far out of the screen and want it to get back to the screen center immediately. 2. Change the vertical setup and notice that each change affects the status bar differently.
RIGOL To Understand the Horizontal System Figure 1-10 shows the HORIZONTAL controls: MENU button, and knobs of horizontal system. Following the exercise to familiarize with the buttons, knobs, and status bar. Figure 1-10 The horizontal system 1. Turn the knob and notice the change in the status bar. The horizontal knob changes the sweep speed in a 1-2-5 step sequence, and displays the value at the upper-left of the display. The time base ranges of the DS1000B series is from 5 ns/div* to 50 s/div.
RIGOL 2. The horizontal on the waveform window. knob moves displayed signal horizontally The horizontal knob adjusts the trigger offset of signal, when turning the knob; you can note that the waveform moves horizontally in conjunction with the knob. Horizontal offset back to 0 shortcut key Press the knob to set the horizontal offset to 0 as a shortcut key, this is especially helpful when the trigger point is far out of the screen and want it to get back to the screen center immediately. 3.
RIGOL To Trigger the Oscilloscope Figure 1-11 shows the TRIGGER control: MODE, MENU, FORCE, 50% and a knob. Following the exercises to familiarize with the buttons, trigger level knob and status bar. Figure 1-11 The trigger control window Three kinds of trigger modes can be switched among Auto, Normal and Single with pressing MODE button. 1. Use knob to change the settings of trigger level.
RIGOL 2. Change the trigger setup and notice these changes in the status bar. Press MENU button in the Trigger control area. A soft button menu appears on the display showing the trigger setting choices as shown in Figure 1-12. · Press No.1 menu operating button to choose trigger mode as Edge. · Press No.2 menu operating button to choose trigger source as CH1. · Press No.3 menu operating button to choose slope type as · Press No.4 menu operating button to choose sweep mode as Auto. · Press No.
RIGOL Key point: Holdoff: A time interval before the oscilloscope responses to next trigger signal. During this Holdoff period, the trigger system becomes “blind” to trigger signals. This function helps to view complex signals such as an AM waveform. Press Holdoff button to activate ( ) knob, then turn it to adjust Holdoff time. 1-16 User‟s Guide for DS1000B series © 2008 RIGOL Technologies, Inc.
RIGOL To Understand the Quick Function The following figure shows that there are two buttons in the QUICK function area. The exercise below will gradually conduct you to familiarize the settings of the buttons. Figure 1-13 The quick function menu 1. Firstly, you need to press the button Measure form the main MENU, then press the button associated with QuickMea, you will navigate into the quick measurement setting menu. Three measuring items can be pre-set from the 22 measurements.
RIGOL Chapter 2 : Operating Your Oscilloscope By now, a user should understand the VERTICAL, HORIZONTAL and TRIGGER control systems and know how to determine the system setup from the status bar of a DS1000B series digital oscilloscope. This chapter will go through all groups of front-panel buttons, knobs and menus; and further the knowledge of the operation by hints in this guide.
RIGOL To Set up the Vertical System To Set the Channels Each channel of DS1000B series digital oscilloscope has an independent operation menu and it will pop-up after pressing any button among CH1, CH2, CH3, CH4. The settings of all items in the menu are shown in the table below. Figure 2-1 Table 2-1 The Channel menu (Page 1/2) Menu Settings AC Coupling DC GND ON BW Limit OFF 0.001X . .
RIGOL Figure 2-2 Table 2-2 The Channel menu (Page 2/2) Menu Settings 2/2 Coarse Volts/Div Fine Invert ON OFF Unit V/ A/ W/ U Comments Back to the previous menu page (The followings are the same, no more explanation). To change the Volts/Dive settings in a 1-2-5 sequence from 2mv/div to 10v/div with knob. To change the Volts/Dive settings in small steps between the coarse settings. Turn on the invert function. Restore to original display of the waveform.
RIGOL To set up channel coupling To use Channel 1 as an example, input a square wave signal with DC shift. Press CH1→Coupling→DC to set “DC” coupling. Both the AC component and the DC component of the input signal will pass. The waveform is displayed as Figure 2-3: DC coupling Setup DC coupling Status symbol Figure 2-3 Waveform display Press CH1→Coupling→AC, to set “AC” coupling. It will pass AC component of the input signal with frequency higher than 5 Hz and restraint DC component of the input signal.
RIGOL AC coupling Setup AC coupling Status symbol Figure 2-4 Waveform display Press CH1→Coupling→GND, to set “GND” coupling, it disconnects the input signal. The screen displays as Figure 2-5: GND coupling Setup GND coupling Status symbol Figure 2-5 Screen display © 2008 RIGOL Technologies, Inc.
RIGOL To set up the channel bandwidth limit Take Channel 1 for an example, input a signal containing high frequency component. Press CH1→BW Limit→OFF, to set up bandwidth limit to “OFF” status. The oscilloscope is set to full bandwidth and passing the high frequency component in the signal. The waveform is displayed as Figure 2-6: Turn off the BW limit Figure 2-6 Turn off the BW limit Press CH1→BW Limit→ON, to set up bandwidth limit to “ON” status.
RIGOL 20MHz BW limit Mark of BW limit Figure 2-7 Turn on the BW limit © 2008 RIGOL Technologies, Inc.
RIGOL To set up Probe Attenuation The oscilloscope allows selecting the attenuation factor for the probe. The attenuation factor changes the vertical scaling of the oscilloscope so that the measurement results reflect the actual voltage levels at the probe tip. To change (or check) the probe attenuation setting, press the CH1, CH2, CH3 or CH4 button (according to which channel is in using). Toggle the Probe soft button to match the attenuation factor of the probe.
RIGOL To change the Volts/Div settings The Volts/Div control has Coarse or Fine configuration. The Vertical Sensitivity is 2mv/div - 10V/div. Coarse: It is the default setting of Volts/Div in a 1-2-5-step sequence from 2mV/div, 5mV/div, 10mV/div, to 10 V/div. Fine: This setting changes the vertical scale to small steps between the coarse settings. It will be helpful to adjust the waveform in smooth steps.
RIGOL To invert a waveform Invert turns the displayed waveform 180 degrees with respect to the ground level. When the oscilloscope is triggered on the inverted signal, the trigger is also inverted. Figure 2-10 and Figure 2-11 show the changes before and after the inversion respectively. Invert off Figure 2-10 The waveform before inversion Invert on Figure 2-11 The waveform after inversion 2-10 User‟s Guide for DS1000B series © 2008 RIGOL Technologies, Inc.
RIGOL To configure the Digital Filter Press CH1→ Digital filter, it will display the digital filter menu. Turn the digital filter on or off as shown in the following figure. Turn ( ) knob to adjust the upper and lower limit of frequency after turning on the digital filter. Turn off digital filter Figure 2-12 Turn off digital filter Turn on digital filter Figure 2-13 Turn on digital filter © 2008 RIGOL Technologies, Inc.
RIGOL Figure 2-14 Table 2-4 The Filter menu Menu Settings Digital Filter On Off Comments Turn on the digital filter. Turn off the digital filter. Filter Type Setup Setup Setup Setup Upper limit Turn ( ) knob to set upper limit high limit. Lower limit Turn ( as as as as LPF (Low Pass Filter). HPF (High Pass Filter). BPF (Band Pass Filter). BRF (Band Reject Filter). ) knob to set lower limit. Back to higher level menu (The followings are the same, no more explanation).
RIGOL To Use Math Functions The mathematic functions include “add”, “subtract”, “multiply” and “FFT” for Channel 1, Channel 2, Channel3 and Channel4. The mathematic result can be measured by the grid and the cursor. Math Scale Figure 2-15 The Math function Figure 2-16 Table 2-5 The Math menu (Page 1/2) Menu Operate Source A Source B Invert Settings A+B A-B A× B FFT CH1, CH2, CH3, CH4 CH1, CH2, CH3, CH4 ON OFF Comments Add source A and source B. Subtract source B from source A.
RIGOL Figure 2-17 Table 2-6 The Math menu (Page 2/2) Menu Settings Comments The multifunctional knob ( ) adjusts the vertical position of the Math waveform. The multifunctional knob ( ) adjusts the vertical amplitude of the Math waveform. 2-14 User‟s Guide for DS1000B series © 2008 RIGOL Technologies, Inc.
RIGOL Using FFT function The FFT (Fast Fourier Transform) process converts a time-domain signal into its frequency components mathematically. FFT waveforms are useful in the following applications: Measuring harmonic content and distortion in systems Characterizing noise in DC power supplies Analyzing vibration Figure 2-18 Table 2-7 The FFT menu (Page 1/2) Menu Settings Comments Operate A+B A-B A× B FFT Add source A to source B. Subtract source B from source A. Multiply source B by source A.
RIGOL Figure 2-19 Table 2-8 The FFT menu (Page 2/2) Menu Settings comments The multifunctional knob( ) adjusts the vertical position of Math The multifunctional knob( ) adjusts the vertical amplitude of Math Vertical scale Vrms dBVrms Set Vrms as vertical scale unit Set dBVrms as vertical scale unit Key points for FFT 1. 2. 3. 2-16 Signals that have a DC component or offset can cause incorrect FFT waveform component magnitude values.
RIGOL Selecting an FFT Window DS1000B series oscilloscope provides four FFT windows. Each window is a trade-off between frequency resolution and amplitude accuracy. It depends on the desired measurement and the source signals characteristics to determine the window to use. Use the following guidelines to select the best window. Table 2-9 FFT Windows Window Features Best frequency resolution and worst amplitude accuracy. Rectangle This is essentially the same as no window.
RIGOL To Use REF Function Reference Waveforms are saved waveforms to be selected for display. The reference function will be available after saving the selected waveform to non-volatile memory. During the actual measurement process, you can use DS1000B series digital oscilloscope to observe the waveforms of correlative components. It will help you to determine the cause of malfunction when you compare the measured waveform with the reference waveform.
RIGOL Figure 2-21 Table 2-11 REF menu when using internal/external memory (Page 2/3) Menu Settings Comments Save REF waveform to outer memory location. Save The multifunctional knob ( ) adjusts the vertical position of REF The multifunctional knob ( ) adjusts the vertical amplitude of REF Figure 2-22 Table 2-12 REF menu when using internal memory (Page 3/3) Menu settings comments Reset Reset REF waveform. Imp./Exp. Go to import menu (see table 2-14).
RIGOL To Import and Export Press REF →Imp./Exp. and go to the following menu. Figure 2-24 Table 2-14 The Imp./Exp. Menu Menu Settings Explorer Path Directory File Comments Switch to Path, directory or file. Export Export the REF file from internal memory to external memory (see table 2-15). Import Import the REF file to internal memory. Delete File Delete file.
RIGOL Export Press REF→Imp./Exp.→ Export and go to the following menu. Figure 2-26 Table 2-15 The Export menu Menu Settings Comments Move the input focus point of files‟ name up and down. Move the focus point to next location. To delete chosen letter. Save Execute the operation. The screen of Export is as follows. Figure 2-27 Figure export © 2008 RIGOL Technologies, Inc.
RIGOL Save to External Memory Press REF→Save and navigate to the following menu. Figure 2-28 Table 2-16 The Save menu Menu Settings Path Directory File Explorer Comments Switch among Path, Directory and File. New File (Folder) Set up new file in Path and File. Set up new folder in directory. Delete File (Folder) Delete chosen file (Folder). The screen of Save is as follows: Figure 2-29 The figure of save 2-22 User‟s Guide for DS1000B series © 2008 RIGOL Technologies, Inc.
RIGOL New File (or New Folder) Press REF→Save→New File (or New Folder) and go to the following menu. Figure 2-30 Table 2-17 The New File menu Menu Settings Comments Move the focus point of file name up and down. Move the focus point to the next location. To delete chosen letter. Save Execute the saving operation.
RIGOL Import Press REF→Import and go to the following menu. Figure 2-32 Table 2-18 The Import menu Menu Settings Comments Explorer Path Directory File Switch among Path, Directory and File. Import the REF file into internal memory. Import The screen of Import is as follows: Figure 2-33 The figure of import 2-24 User‟s Guide for DS1000B series © 2008 RIGOL Technologies, Inc.
RIGOL To Display a Reference Waveform Figure 2-34 Reference waveform display Push REF button to show the reference waveform menu. Continue Pressing button No.1 to turn on or turn off one to four reference channels. Select a REF channel as the current reference channel from the channels activated by step 2. Continue pressing button No.3 to select a signal source channel from CH1, CH2, CH3, CH4 and MATH.
RIGOL Turn on or off Channels CH1, CH2, CH3, CH4 and Ext. trigger channels are input channels. All functions applied will base on operating the instrument with channels, so MATH and REF can be regarded as relatively isolated channels. Press the corresponding button (for example: CH1, CH2, CH3, CH4, MATH, REF) on the front panel to turn the channels on. The backlight indicates the channel is currently active.
RIGOL To Apply the Vertical Knobs Use the vertical controls to display signal waveforms by adjusting the vertical knob, and the knob, and setting the input parameters. 1. Using vertical knob. The vertical control changes the vertical position of signal waveforms in all channels (excluding MATH and REF). The resolution changes according to the vertical level set. Pressing this knob will clear the channel‟s vertical offset to zero. 2. Using vertical knob.
RIGOL To Set up the Horizontal System The oscilloscope shows the time per division in the scale readout. With the horizontal control buttons and knobs, you can change the horizontal time base and trigger offset which indicates the horizontal location of trigger in the memory. The horizontal controls change the horizontal scale and position of waveforms. The horizontal center of the screen is the time reference for waveforms.
RIGOL Figure 2-35 Table 2-20 The Horizontal menu Menu Delayed Time Base Settings ON OFF Turn on the Delayed Scan mode. Turn off the Delayed Scan mode. Y-T Show the relative relation between vertical voltage and horizontal time. Show CH1 amplitude value at X axis; show CH2 amplitude value at Y axis. In Roll Mode, the waveform display updates from right to left. X-Y Roll Adjust the trigger position to the center of the memory.
RIGOL Marks Indicator ① The horizontal time base (main time base). ② The position of the current waveform window in the memory. ③ The trigger position in the memory. ④ The trigger position in the current waveform windows. ⑤ The trigger‟s horizontal offset with respect to the center of the window. Key Points Y-T: The conventional oscilloscope display format. It shows that the voltage of a waveform record (on the vertical axis) varies along with time (on the horizontal axis).
RIGOL Delayed Scan: The Delayed Scan is a magnified portion of the main waveform window. Use Delayed Scan to locate and horizontally expand part of the main waveform window for a more detailed (higher horizontal resolution) analysis of signal. The Delayed Scan time base setting cannot be set slower than the Main time base setting.
RIGOL Use the horizontal portion. knob to change the position of the expanded Turn the horizontal To change the main time base, turn off the Delayed Scan mode. Since both the main and Delayed Scan waveform windows are displayed, each waveform window occupies a half of the original waveform window. They are both compressed vertically to a half size, the same as the vertical divisions. knob to adjust the Delayed Scan resolution.
RIGOL X-Y Format This format is useful for studying phase relationships between two signals. Channel 1 in the horizontal axis(X) and channel 2 in the vertical axis(Y), the oscilloscope uses a none-trigger acquisition mode, data is displayed as dots. Figure 2-38 X-Y display format NOTE: The X-Y display format is only suitable for CH1, CH2, CH3; CH4.In the normal condition, the oscilloscope can acquire waveforms with any sampling rate.
RIGOL To Set up the Trigger System The trigger determines when the oscilloscope starts to acquire data and display a waveform. When a trigger is set up properly, it can convert unstable displays or blank screens into meaningful waveforms. When the oscilloscope starts to acquire a waveform, it collects enough data so that it can draw the waveform to the left of the trigger point. The oscilloscope continues to acquire data while waiting for the trigger condition to occur.
RIGOL Trigger Modes The oscilloscope provides five trigger modes: Edge, Pulse, Video, Pattern and Alternative. Edge: An edge trigger occurs when the trigger input passes through a specified voltage level in the specified slope direction. Pulse: Use this trigger mode to catch pulses with certain pulse width. Video: Use video trigger mode on fields or lines for standard video signals. Pattern: Recognising trigger by searching specified code. Alternative: Trigger on non-synchronized signals.
RIGOL Edge Trigger An edge trigger determines whether the oscilloscope finds the trigger point on the rising or the falling edge of a signal. Select Edge trigger Mode to trigger on Rising edge, falling edge or rising & falling edge. Figure 2-40 Table 2-21 The Trigger menu Menu Settings Comments Source CH1 CH2 CH3 CH4 EXT EXT/5 Select CH1 as trigger source. Select CH2 as trigger source. Select CH3 as trigger source. Select CH4 as trigger source. Select EXT TRIG as trigger source.
RIGOL Pulse Width Trigger Pulse trigger occurs according to the width of pulse. The abnormal signals can be detected through setting up the pulse width condition. Figure 2-41 Table 2-22 The Trigger menu (Page 1/2) Menu Source Settings CH1 CH2 CH3 CH4 EXT Comments Select CH1 as trigger source. Select CH2 as trigger source. Select CH3 as trigger source Select CH4 as trigger source l Select EXT TRIG as trigger source Select attenuated EXT TRIG/5 as trigger source.
RIGOL Video Trigger Choose video trigger to trigger on fields or lines of NTSC, PAL, or SECAM standard video signals. Trigger coupling preset to DC. Figure 2-43 Table 2-24 The Video Trigger menu (Page 1/2) Menu Source Settings CH1 CH2 CH3 CH4 EXT EXT/5 Normal polarity Polarity Sync 2-38 Inverted polarity All Lines Line Num Odd field Even field Comments Selects CH1 as trigger source. Selects CH2 as trigger source. Selects CH3 as trigger source. Selects CH4 as trigger source.
RIGOL Figure 2-44 Table 2-25 The Video Trigger menu (Page 2/2 when Sync is set as the specified line) Menu Line Num Standard Settings < Line sync > PAL/SECM NTSC Auto Normal Sweep Single Set Up Comments Select the specified line number for sync. Select Video standard. Force the oscilloscope to trigger in the absence of trigger condition. Lets oscilloscope to trigger in the suitable trigger condition. Lets oscilloscope to trigger one time in the suitable trigger condition, and then stop.
RIGOL Key points: Sync Pulses: When Normal Polarity is selected, the trigger always occurs on negative-going sync pulses. If the video signal has positive-going sync pulses, use the inverted Polarity. Figure 2-46 Video Trigger: Line Synchronization Figure 2-47 Video Trigger: Field Synchronization 2-40 User‟s Guide for DS1000B series © 2008 RIGOL Technologies, Inc.
RIGOL Pattern Trigger Pattern trigger recognises the trigger condition through inspecting the code preset which means the logic values or the logic combination.Every channel has its own logic value,such as,logic high value,logic low value,and the neglect value.
RIGOL Figure 2-49 Pattern trigger: set code as falling edge 2-42 User‟s Guide for DS1000B series © 2008 RIGOL Technologies, Inc.
RIGOL Alternative Trigger When alternative trigger is on, the trigger sources come from two vertical channels. This mode can be used to observe two non-related signals. You can choose two different trigger modes for the two vertical channels. The options are as follows: Edge, Pulse and video. The info of the trigger level of the two channels will be displayed on the upper-right of the screen.
RIGOL Figure 2-52 Table 2-30 The Alternative trigger menu (Page 1/2 Trigger Type: Pulse) Menu Settings Alt Channel Figure 2-53 Comments CH1-CH2 CH1-CH3 CH1-CH4 CH2-CH3 CH2-CH4 CH3-CH4 To specify the current alternative trigger channels as any two channels selected form CH1, CH2, CH3, CH4. select CH1, CH2, CH3, CH4 To select vertical channel among CH1, CH2, CH3, CH4 in alternative trigger mode so as to configure further settings.
RIGOL Figure 2-54 Table 2-32 The Alternative trigger menu (Page 1/2 Trigger Type: Video) Menu Alt Channel Figure 2-55 Settings CH1-CH2 CH1-CH3 CH1-CH4 CH2-CH3 CH2-CH4 CH3-CH4 Comments To specify the current alternative trigger channels as any two channels selected form CH1, CH2, CH3, CH4. select CH1, CH2, CH3, CH4 To select vertical channel among CH1, CH2, CH3, CH4 in alternative trigger mode so as to configure further settings.
RIGOL Trigger Settings Set up different trigger settings according to different trigger modes. When in the mode of Video and Alternative, only trigger sensitivity, trigger holdoff time, holdoff reset is adjustable. Figure 2-56 Table 2-34 The Trigger Set Up menu (Page 1/2 Settings for trigger coupling, trigger sensitivity and Holdoff) Menu Coupling HF reject Sensitivity Holdoff 2-46 Settings Comments DC AC LF Reject Allow all signals pass. Block DC signals. Reject DC and low frequency signals.
RIGOL Figure 2-57 Table 2-35 Restore the trigger holdoff time to default (page 2/2) Menu Holdoff Settings comments Reset Holdoff time to 100ns. When Settings for “sensitivity” and “Hold off” in the Video and Alternative trigger mode, the menu is shown below. Figure 2-58 Table 2-36 The Trigger Set Up menu Menu Sensitivity Holdoff Holdoff Reset Settings Comments Set trigger sensitivity. Set time slot before another trigger event. Reset Holdoff time to 100ns.
RIGOL Trigger Holdoff You can use trigger Holdoff to stabilize a complex waveform, such as a pulse sequence. Holdoff time is the oscilloscope‟s waiting period before starting a new trigger. During Holdoff, oscilloscope will not trigger until Holdoff ends. For instance: To trigger on the first pulse on a group of them, users can set the holdoff time to Pulse cluster width. Trigger site Trigger holdoff Time of holdoff Figure 2-59 Trigger Holdoff To use trigger Holdoff: 1.
RIGOL Trigger Key points 1. Trigger Source: Trigger occurs from several sources: Input channels (CH1, CH2, CH3, CH4), AC Line, Ext, Ext/5. CH1, CH2, CH3, CH4: It is the most commonly used trigger source. The channel works when it is selected as a trigger source whatever displayed. Ext Trig: The instrument can be triggered from an external source while acquiring data from CH1, CH2, CH3, and CH4. For example, be a trigger source with an external clock or a signal from another part of the test circuit.
RIGOL setting), it will force itself to trigger. When forcing invalid triggers, the oscilloscope cannot synchronize the waveform, and the waveform seems to roll across the display. If valid triggers occur, the display becomes stable on the screen. Any factor results in the instability of waveforms can be detected by Auto Trigger, such as the output of Power supply. NOTE: When horizontal control is set under 50 ms/div, Auto mode allows the oscilloscope not to capture trigger signal.
RIGOL surveyed. More data (14div) of pre-trigger and 1s delayed trigger can be surveyed by adjusting the horizontal knob. This feature is very useful to study the events that led up to the trigger point. Everything on the right of the trigger point is called post-trigger information. The delay range (pre-trigger and post-trigger information) depends on the sweep speed selected. 3.
RIGOL To Set up the Sampling System Figure 2-60 shows, the Acquire button at the MENU of the front panel. Acquire setup button Figure 2-60 The Acquire setup button Press the Acquire button, the interface menu shows as follows: Figure 2-61 Table 2-37 The Acquire menu Menu Settings Comments Acquisition Normal Average Peak Detect Normal Acquisition mode. Average Acquisition mode. Peak Detect Acquisition mode. Sampling Real-Time Equ-Time Real-time sampling mode. Equivalent sampling mode.
RIGOL The waveform displayed on the screen will change in conjunction with the setting of Acquire menu. Figure 2-62 Signal that contains noise, and without average sampling Figure 2-63 Display signal after average 256 times sampling Note: Select Real-time acquisition to observe the single-shot or pulse signals. Select Equ-Time to observe high frequency repetitive signals. © 2008 RIGOL Technologies, Inc.
RIGOL To reduce the displayed random noise, select the Average Acquisition. This mode would make the screen refresh slower. To Avoid signal aliasing, select Peak Detect Acquisition. Figure 2-64 Signal with Peak Detect Acquisition The Peak Detect effect is shown as the figure above.
RIGOL Key Points Real-time Sampling: DS1000B has Real-time sampling rate up to 2GSa/s. At the time base 50ns or faster, the oscilloscopes use the sine(x)/x interpolation to expand the horizontal time base. Equivalent sampling: Known as Repetitive sampling to get up to 20ps of horizontal resolution (equivalent 50Gsa/s). This mode is good for observing repetitive signals, and it is not recommended for single-shot or pulse. Normal: Oscilloscope acquires signal by equal time interval.
RIGOL To Set up the Display System The menu button for the display system on the front panel is shown in the following figure. Display setup button Figure 2-65 The display setup button Press the Display button, the interface menu for settings shows as follows. Figure 2-66 Table 2-38 The Display menu (Page 1/3) Menu Type Setting Vectors Dots Infinite Persist OFF 2-56 Display waveforms as vectors. Display waveforms as dots. Clear all existing waveforms from screen.
RIGOL Figure 2-67 Table 2-39 The Display menu (Page 2/3) Menu Grading Settings On Off Figure 2-68 The waveform is displayed multiple levels of pixel intensity. in Display grids and coordinates on the screen. Turn off the grids. Turn off the grids and coordinates. Grid Menu Display Comments 1s 2s 5s 10s 20s Infinite Set the time before menu fades away. The menu will be hidden after the set time of last button pressing.
RIGOL Key points: Display type: Display type includes Vector and Dot. In vectors type, oscilloscope connects dots through digital interpolation including both linearity and sin(x)/x. Sin(x)/x interpolation is suitable for Real-time sampling and will be more effective at 50ns or faster time base. Refresh rate: It is an important performance of digital oscilloscopes. It means the number of display refreshing per second and it will affect the ability to observe signal.
RIGOL To Store and Recall Figure 2-69 shows the menu button for the storage system on the front panel. Storage setup button Figure 2-69 The Storage setup button Press the Storage button to show the menu for the settings of the storage system. Waveforms and setups can be stored in and recalled from both internal memory and external memory.
RIGOL Waveform and setup menus are as follows: Figure 2-70 Table 2-41 The Storage menu Menu Storage Settings Comments Waveform Setups 8-Bitmap 24-Bitmap PNG CSV Factory Store or recall waveform. Store or recall instrument setups. Create or delete 8bits bitmap files. Create or delete 24bits bitmap files. Create or delete PNG format files. Create or delete CSV files. Recall factory setups. Internal Go to menu for internal memory operation (see Table 2-46).
RIGOL For bitmap, the menu is as follows: Figure 2-72 Table 2-43 The Storage menu Menu Storage Settings Waveform Setups 8-Bitmap 24-Bitmap PNG CSV Factory Para Save On Off External Disk Mana. Comments Store or recall waveform. Store or recall setups. Create or delete 8bits Bitmap files. Create or delete 24bits Bitmap files. Create or delete Portable Network Graphics format files. Create or delete CSV files. Recall factory setups.
RIGOL For PNG, the menu is as follows. Figure 2-74 Table 2-45 The Storage menu Menu Storage Para Save Settings Comments Waveform Setups 8-Bitmap 24-Bitmap PNG CSV Factory On Store or recall waveform. Store or recall instrument setups. Create or delete 8bits bitmap files. Create or delete 24bits bitmap files. Create or delete PNG format files. Create or delete CSV files. Recall factory setups. Save the current oscilloscope settings in different format with the same file name.
RIGOL Internal Memory Press Storage→Internal to go to the following menu. Figure 2-75 Table 2-46 The Memory menu Menu Settings Comments Internal Int_00 (N) . . . Int_09 (N) Set up the location of files in internal memory. Load Recall waveform files and setup files from the internal memory location. Save Save waveform files and setup files to the internal memory location. Delete Files Delete selected files. © 2008 RIGOL Technologies, Inc.
RIGOL External Memory Press Storage→ External to go to the following menu. Figure 2-76 Table 2-47 The Memory menu Menu Settings Path Directory File Explorer Comments Switch among Path, Directory and File. New File (Folder) To create new file or folder.(see table 2-13) Delete File (Folder) Delete file (Folder). Load Recall waveform and setup from USB storage device. File system is as follows: Path Back to previous folder Files Current directory File info.
RIGOL U-disk Selection: Figure 2-78 U-disk selection DS1000B series have two USB Host ports on the front and rear panel. When both ports were inserted USB flash drive, the screen will appear a U-disk selection interface as Figure 2-78. Turn the multifunction knob to choose the driver. The front one marked “F:”, the rear one “G:”. © 2008 RIGOL Technologies, Inc.
RIGOL Disk Management Press Storage→Disk Mana. to go to the following menu. Figure 2-79 Table 2-48 The Disk Management menu (Page 1/2) Menu Settings Path Directory File Explorer Figure 2-80 Switch among Path, Directory and File. New folder To create new folder (same as new files, see table 2-13). Delete File Delete selected file. Load Recall waveform, setup, recorded waveform, Pass/Fail file.
RIGOL Rename Press Storage→Disk Mana.→Rename to go to the following menu. Figure 2-81 Table 2-50 The Rename menu Menu Settings Comments To move the input focus point of file name up and down. To move the focus point to the next location. To delete chosen letter. OK Rename the file. The Rename system screen is as follows: Figure 2-82 Rename the file © 2008 RIGOL Technologies, Inc.
RIGOL Factory The oscilloscope has default settings and can be recalled at anytime by user. Memory location Specify the memory location to save/recall the waveforms and setups. Load Recall saved waveforms, setups and default settings. Save Save waveforms and setups. NOTE: 1. Select Save stores not only the waveforms, but also the current settings of the oscilloscope 2.
RIGOL To Set up the Utility The following figure shows the menu button for the Utility on the front panel. System function setup button Figure 2-83 The Utility setup button Press the Utility button to show the menu of the settings in the Utility system. Figure 2-84 Table 2-51 The Utility menu (Page 1/3) Menu Setting I/O setting Sound Counter Setup I/O configuration. (ON) (OFF) OFF ON 简体中文 繁体中文 Language Comments English Japanese Français Turn beeper sound on/off. Turn off Frequency Counter.
RIGOL Figure 2-85 Table 2-52 The Utility menu (Page 2/3) Menu Figure 2-86 Settings Comments Pass/Fail Setup Pass/Fail test. Record Setup Waveform Recorder. Print set Setup printing. Table 2-53 The Utility menu (Page 3/3) Menu Settings Comments Self-Cal Execute Self-calibration. System Info Show the following information: Serial number, software version, installed module. Preference Go to preference menu. Date/Time Set the real date and time for system.
RIGOL The I/O Setup Press Utility→IO Setting to go to the following menu. Figure 2-87 Table 2-54 The I/O Setup menu Menu Display Set network function, the local area network interface corresponds with the LXI standard. LAN Set USB Device GPIB# Comments Auto Detect Computer PictBridge
Set USB device function, connect USB interface to the needed device. Set the GPIB address, the address range is from 0 to 30. © 2008 RIGOL Technologies, Inc.RIGOL LAN Setting Press LAN Set to enter the following interface. Figure 2-88 LAN Setting interface The black parts in the setting area are the current items can be set. Turn the multifunction knob to set the item on the cursor. IP information can be modify by using the multifunction knob.
RIGOL Press Load to load the LAN initial settings. Figure 2-90 Table 2-56 LAN Setting menus Menu Display Comments OK Load the LAN initial settings Cancel Cancel the operation Key points: DHCP: Dynamic Host Configuration Protocol Auto IP: Auto IP Manual: Manual IP MAC: Medium Access Control Layer protocol VISA: Virtual Instrument System Architecture IP Configure Mode IP Address Subnet Mask Gateway DNS Server © 2008 RIGOL Technologies, Inc.
RIGOL Preference Press Utility→Preference to go to the following menu Figure 2-91 Table 2-57 The Preference menu Menu Display Comments Screen saver 5 hour . . 1 min OFF Set up screen saver timer. Expand Ground Center Set up waveform vertical expand reference. Stickykey Turn the sticky status of CH1, CH2, CH3, CH4, Math, and REF on or off. Default load Use the last setup in the next power on. Use the default setup of the system in the next power on.
RIGOL Self-Calibration The Self-Calibration adjusts the internal circuitry to get the best accuracy. Use these functions to calibrate the vertical and horizontal systems. For maximum accuracy at any time, run this calibration if the ambient temperature changes by 5℃ or more. Before running this procedure, do these steps: 1. 2. Disconnect any probes or cables from all channel inputs, otherwise failure or damage to the oscilloscope may occur. Push the Utility button and select Self-Cal.
RIGOL Pass/Fail The Pass/Fail function monitors changes of signals and output pass or fail signals by comparing the input signal which is within the pre-defined mask. Press Utility→Pass/Fail to go to the following menu. Figure 2-93 Table 2-58 The Pass/Fail menu (Page 1/2) Menu Setting Enable Test ON OFF Turn on Pass/Fail test. Turn off Pass/Fail test. Source CH1, CH2, CH3, CH4 Select Pass/Fail test on CH1, CH2, CH3 or CH4.
RIGOL Mask setting Press Utility→Pass/Fail→Mask Setting to go to the following menu. Figure 2-95 Table 2-60 The Mask setting menu (Page 1/2) Menu X Mask Y Mask Settings Set horizontal clearance to waveform (0.04div-4.00div). Set vertical clearance to the waveform (0.04div-4.00div). Create Mask Location Comments the Create a test mask according to the above clearance. Internal External Set the memory location of the mask files.
RIGOL When using the external memory to save, the operation menu is shown below. Figure 2-97 Table 2-62 The Mask setting menu (Page 2/2) Menu 2-78 Settings Comments Save Go to save menu (same as REF save menu, see table 2-10). Load Go to load menu (see Table 2-57). Import Go to import menu. (Same as REF import menu, see table 2-12). User‟s Guide for DS1000B series © 2008 RIGOL Technologies, Inc.
RIGOL Load Press Utility→Pass/Fail→Mask Setting→Load to go to the following menu. Figure 2-98 Table 2-63 The Load menu Menu Explorer Load Settings Path Directory File Comments Switch among Path, Directory and File. Recall the specified file. NOTE: Pass/Fail function is unavailable in X-Y mode. © 2008 RIGOL Technologies, Inc.
RIGOL Print Setting DS1000B series oscilloscopes support external printers. Press Utility→ Print set to enter to the following menu. Normal mode Press PrintMod button,choose the print mode as Normal ,and navigate into the following menu. Figure 2-99 Table 2-64 The Print Set menu Menu Settings Comments PrintMod Normal PictBridge Set the print mode as normal. Set the print mode as PictBridge. Print Execute the print operation. Inverted ON OFF Invert the color for print. Print original color.
RIGOL Figure 2-101 Table 2-66 The print setup menu (Page 2/4) Menu Paper size File type copies Figure 2-102 1~999 Comments Select the paper size for printing. To specify the image type for printing as Exif/Jpeg or Bmp format. To set the print copies from 1to 999.
RIGOL Waveform Recorder Waveform recorder records input waveforms from CH1, CH2, CH3 and CH4 with a maximum record length of 1000 frames. This performance can also be activated by the Pass/Fail test output, which makes this function especially useful to capture abnormal signals in long term without keeping an eye watching it. Press Utility→Record→Mode→Record to go to the following menu Waveform recorder: Record the waveforms with specified interval.
RIGOL Play back: Play back the recorded waveforms. Figure 2-106 Table 2-71 The Record menu (Page 1/2) Menu Settings (Run) Operate (Stop) Figure 2-107 Comments Play stopped, press to Start playback. Press to stop playing. Play mode Set repeat play mode. Set single time play mode. Interval Set up interval value between frames. <1.00ms-20s> Table 2-72 The Record menu (Page 2/2) Menu Start Frame Current Frame End Frame Settings <1-1000> <1-1000> <1-1000> Comments Set start frame.
RIGOL Storage: Store recorded waveforms in non-volatile memory according to the setup frames. Figure 2-108 Table 2-73 The Storage menu (Page 1/2) Menu Start Frame End Frame Settings Comments Set first frame to be saved. <1-1000> Set last frame to be saved. <1-1000> Internal external Location Set up Store location. When using the internal memory to save, the operation menu is shown below.
RIGOL When using the external memory to save, the operation menu is shown below. Figure 2-110 Table 2-75 The Storage menu Menu Settings Comments Save Go to save menu (same as REF save menu. see table 2-10). Load Go to load menu (see Table 2-57). Import Go to import menu. (Same as REF import menu, see table 2-12). © 2008 RIGOL Technologies, Inc.
RIGOL System information Press Utility, go the following menu and Press System info. Figure 2-111 Operation menu System Info: The system information includes: Product Model, Serial Number, Software Version and Installed Module of the oscilloscope. Follow the prompting message “<>" to exit the information display interface. 2-86 User‟s Guide for DS1000B series © 2008 RIGOL Technologies, Inc.
RIGOL Language The DS1000B series oscilloscopes have multi-language user menu, choose as your desire. Press Utility→Language to select the language. Press button No.4 until the desired language appears. Figure 2-112 The language selection screen © 2008 RIGOL Technologies, Inc.
RIGOL Date and Time DS1000B Series provide build-in system real-time date and time. Users can set the date and time as shown in the following figure. Time display Figure 2-113 Date and time setting Figure 2-114 Table 2-76 Time setting menu Menu Display Settings OFF ON Comments Close time display Open time display Move up/down the setting focus Move left/right the setting focus OK 2-88 Save the current settings User‟s Guide for DS1000B series © 2008 RIGOL Technologies, Inc.
RIGOL To Measure Automatically The Measure button in the menu area activates the automatic measurement function. The instruction below shows the powerful measurement function of DS1000B series oscilloscopes. Auto Measurement Figure 2-115 The Measure setup button Menu explanation: Press the Measure button to display the menu of the Automatic Measurements settings.
RIGOL Figure 2-117 Table 2-78 The Measure menu (Page 2/2) Menu Settings Display All OFF ON QuickMea Delay/Phas e Setup 2-90 Comments Turn off all measurements result. Turn on all measurements result. To specify three quick measure items which you can choose among the above 22 measurements. To configure the delay and phase channels, in order to measure delay and phase value relative to time measurement between any two vertical channels form CH1 to CH4.
RIGOL Quick Measurement Setup Press Measure→QuickMea button,go to the quick measure menu。 Figure 2-118 Table 2-79 The quick measure menu Menu Item1 Item2 Item3 Settings Comments To specify Item1, Item2 and Item3 respectively, each item can be chosen from 22 kind‟s measure items. © 2008 RIGOL Technologies, Inc.
RIGOL Delay/Phase Setup Press Measure→Delay/Phase Setup button,and go to the delay/phase setup menu. Figure 2-119 Table 2-80 The delay/phase setup menu. Menu Settings DelayA DelayB PhaseA PhaseB 2-92 CH1 CH2 CH3 CH4 Comments To set CHA of Delay measurement as one of CH1,CH2,CH3,CH4 To set CHA of Phase measurement as one of CH1,CH2,CH3,CH4 User‟s Guide for DS1000B series © 2008 RIGOL Technologies, Inc.
RIGOL Voltage Measurements The DS1000B series oscilloscopes provide automatic voltage measurements including Vpp, Vmax, Vmin, Vavg, Vamp, Vrms, Vtop, Vbase, Overshoot and Preshoot. Figure 2-120 below shows a pulse with some of the voltage measurement points. Overshoot Vmax Vtop Vpp Vamp Vbase Vmin Preshoot Figure 2-120 Voltage parameters Vpp: Peak-to-Peak Voltage. Vmax: The maximum amplitude. The most positive peak voltage measured over the entire waveform. Vmin: The minimum amplitude.
RIGOL Press Measure→Voltage button,and go to the following menu. Figure 2-121 Table 2-81 The Voltage Measurement menu (Page 1/3) Menu Figure 2-122 Measure maximum voltage of a waveform. Vmin Measure minimum voltage of a waveform. Vpp Measure Peak-to-Peak Voltage. Vtop Measure a flat top voltage of a square waveform. Table 2-82 The Voltage Measurement menu (Page 2/3) Settings Comments Vbase Measure a flat base voltage of a square waveform. Vamp Measure voltage between Top and Base.
RIGOL Time Measurements The DS1000B series oscilloscopes provide 12 kinds timing parameters auto-measurements; Frequency, Period, Rise Time, Fall Time, +Width, -Width, Delay A→B , Delay A→B , +Duty and –Duty, Phase A→B ,Phase A→B . Figure 2-124 shows a pulse with some of the time measurement points. Fall Time Rise Time 90% 50% 10% +Width -Width Figure 2-124 The time parameters Rise Time: Time that the rising edge of the first pulse in the waveform takes to rise from 10% to 90% of its amplitude.
RIGOL Press Measure→Time button,and go to the following menu Figure 2-125 Table 2-84 The Time Measurement menu (Page 1/3) Menu Figure 2-126 Settings Period Measure Period of a waveform. Freq Measure Frequency of a waveform. Rise time Measure Rise Time of a rising edge. Fall time Measure Fall Time of a falling edge. Table 2-85 The Time Measurement menu (Page 2/3) Menu Figure 2-127 Settings Comments +Width Measure +Pulse Width of a pulse wave. -Width Measure –Pulse Width of a pulse wave.
RIGOL NOTE: The 18 kind's results of the automatic measurements will be displayed on the bottom of the screen. Maximum 3 results could be displayed at the same time. When there is no room, the next new measurement result will make the previous results moving left out of screen. © 2008 RIGOL Technologies, Inc.
RIGOL Using Automatic Measurement as the following steps demonstration: 1. Select the signal channel for measuring. CH1, CH2, CH3, CH4 according to the interest. Press soft buttons as follows: Measure→Source→CH1, CH2, CH3 , CH4. 2. To see all measurement values, Press No.2 button to set the Display All to ON. 18 measurement parameters will be displayed on the screen. 3. Select parameters page for measuring; press No.2 or No.
RIGOL To Measure with Cursors Figure 2-124 shows the Cursor button on the front-panel. Cursor Measurement Figure 2-129 The Cursor setup button The cursor measurement has three modes: Manual, Track and Auto Measure. 1. Manual In this mode, the screen displays two parallel cursors. Move the cursors to make custom voltage or time measurements of the signal. The values are displayed on the boxes below the menu. Before using cursors, make sure to set the Signal Source as the channel for measuring. 2.
RIGOL Manual Mode Figure 2-130 Table 2-87 The Cursors menu Menu Mode Settings Comments Manual Adjust the cursor to measure X/Y parameters manually. X Type Y Source CH1 CH2 CH3 CH4 MATH Shown as vertical line to measure the horizontal parameters. Shown as horizontal line to measure the vertical parameters. Select the measurement signal source. In this mode, the oscilloscope measures the Y or X coordinate values of the cursors, and the increments between the two cursors.
RIGOL Table 2-88 The Cursor menu Cursor Increment X Cursor A Y X Cursor B Y Operation Turn the multifunctional knob ( cursor A horizontally. Turn the multifunctional knob ( cursor A vertically. ) to move Turn the multifunctional knob ( cursor B horizontally. Turn the multifunctional knob ( cursor B vertically. ) to move ) to move ) to move NOTE: Cursor could be moved only when the curse function menu is displayed. 5.
RIGOL Track Mode Figure 2-131 Table 2-89 Cursor descriptions Menu Settings Comments Mode Track CH, CH2, CH, CH4, MATH None CH, CH2, CH, CH4, MATH None Set Track mode in cursor measurement. Cursor A Cursor B Set Cursor A in conjunction with CH1, CH2 or turn off Cursor A. Set Cursor B in conjunction with CH1, CH2 or turn off Cursor B. CurA (Cursor A) Turn the multifunctional Knob ( move cursor A Horizontally. ) to CurB (Cursor B) Turn the multifunctional knob ( move cursor B horizontally.
RIGOL NOTE: Moving cursor horizontally is not allowed when other (not tracking cursor) menu is activated. (4) To get measurement values: Position of Cursor 1 (Time cursor centered on the midpoint of screen; Voltage cursor centered on channel ground level). Position of Cursor 2 (Time cursor centered on the midpoint of screen; Voltage cursor centered on channel ground level). Read the horizontal space between Cursor 1 and 2 (△X): Time between cursors, units in seconds. (1/△X), units in Hz, kHz, MHz, GHz.
RIGOL Auto mode Figure 2-132 Table 2-91 The Mode setting Menu Mode Settings Comments Auto Display the cursors for the current automatic measuring. (See the following figure). Figure 2-133 Auto Measure Mode of Cursor Measurement There will be no cursor display if no parameters are chosen in Measure menu. The oscilloscope could move cursor automatically to measure 22 parameters in Measure menu. 2-104 User‟s Guide for DS1000B series © 2008 RIGOL Technologies, Inc.
RIGOL To Use Run Control Buttons The RUN control buttons include AUTO (auto setting) and Run/Stop. Auto: The AUTO features automatic adjustments to produce a stable display of the input signal. Press AUTO button, the following menu appears. Figure 2-134 Table 2-92 The Auto menu Menu Multi-Cycle Single Cycle Rise Edge Fall Edge Undo Settings Comments Press to display multi-cycle waveform on the screen. Press to display single waveform on the screen.
RIGOL Auto-Set functions When the AUTO is pressed the oscilloscope is configured to the following defaults: Table 2-93 The Auto menu Menu Settings Display format Y-T Acquire mode Normal Vertical coupling Adjust to AC or DC according to the signal. Vertical “V/div” Adjusted Volts/Div Coarse Bandwidth limit Full Signal Invert OFF Horizontal position Center Horizontal “S/div” Adjust to right position. Trigger type Edge Trigger source Find the channel with input signal automatically.
RIGOL Chapter 3 : Application & Examples Example 1: Taking Simple Measurements The function is used to observe an unknown signal; to display, measure frequency, and peak-to-peak amplitude. To quickly display a signal, please do the steps as follows: 1. Set the probe and the channel attenuations to 10X 2. Connect signal to CH1 with the probe 3. Press the AUTO button The oscilloscope sets the vertical, horizontal, and trigger controls at the best status automatically.
RIGOL Example 2: View a Signal Delay Caused by a Circuit This example is to test the input and output signals of a circuit and observe the signal delay. First, set the probe and the channel attenuation to 10X and connect CH1 probe to the input, CH2 to the output of the circuit. Do these steps as follow: 1. Display the signals (CH1 and CH2): 1) Press the AUTO button 2) 3) 4) Adjust the vertical and the horizontal scale by turning the set appropriate ranges for display.
RIGOL Example 3: Capture a Single-Shot Signal To capture a single event, it needs to gather some pre-test knowledge of the signal in order to set up the trigger level and slope correctly. For example, if the event is derived from TTL logic, a trigger level of 2 volts should work on a rising edge. The following steps show how to use the oscilloscope to capture a single event. 1. Set the probe and the channel attenuations to 10X. 2. Set up the trigger.
RIGOL Example 4: To Reduce the Random Noise on a Signal If the signal is noisy (Figure 3-2), set up the oscilloscope to reduce the noise on the waveform and avoid its interference to the signal. Figure 3-2 Waveform display 1. Set the probe and the channel attenuations to 10X. 2. Connect a signal to the oscilloscope and obtain a stable display. 3. Improve the trigger by setting the Coupling 1) Press the MENU in the Trigger control area.
RIGOL 1) If there is noise within the signal and the waveform looks too wide, in this case, choose average acquisition. In this mode the waveform will be thin and easy to observe and measure. To use average acquisition follows these steps. Press soft button as Acquire→Acquisition→Average Toggle the Averages soft button to select the number of averages that best eliminates the noise from the displayed waveform. It can be adjusted from 2256.
RIGOL Example 5: Making Cursor Measurements There are 22 build-in automatic measurements. They can also be conducted using cursors to make time and voltage measurements of a waveform quickly. Measure the Peak Frequency of the First Sinc Waveform To measure the rising frequency at the rising edge of a signal, do these steps: 1. Press Cursor key to see the Cursor menu. 2. Press Mode to set Manual mode. 3. Press Type to select X. 4. Turn ( ) knob to place cursor A on the first peak of the wave. 5.
RIGOL Measure the Amplitude of the First Waveform Peak of the Sinc. Please follow these steps: 1. Press Cursor key to see the Cursor menu. 2. Press Mode to set Manual mode 3. Press Type to select Y. 4. Turn ( ) knob to place cursor A on the first peak of the wave. 5. Turn ( ) knob to place cursor B on the second peak of the wave.
RIGOL Example 6: The Application of the X-Y operation Viewing Phase Changes through a Network Theme: Connect the oscilloscope to monitor the input and output of the circuit and capture the phase changes. To view the input and output of the circuit in an X-Y display, do these steps: 1. From the probe menu set the attenuation to 10X. Set the switch to 10X on the probes. 2. Connect the CH 1 probe to the input of the network, and connect the CH 2 probe to the output. 3.
RIGOL Signal must be centered in X D A B C Figure 3-6 Ellipse method to observe the phase difference Sin = A/B or C/D, where = phase shift (in degrees) between the two signals. From the formula above, you could get: = arcsine (A/B) or arcsine (C/D) If the main axis of the ellipse is at I and III quadrant, must be in the range of (0~π/2) or (3π/2~2π). If the main axis is at II and IV quadrant, must be in the range of (π/2~π) or (π~3π/2). © 2008 RIGOL Technologies, Inc.
RIGOL Example 7: Triggering on a Video Signal Test a video circuit in the DVD set. Use video trigger to obtain a stable display. Triggering on Video Fields To trigger on the video fields, please do as the following steps: 1 Press the MENU key in the Trigger control area to see the Trigger menu. 2 Press Mode to select Video. 3 Press Source to select CH1 as trigger source. 4 Press Polarity to select 5 Press Sync as Odd Field or Even Field. 6 Adjust the .
RIGOL Triggering on the Video Lines 1. Press the MENU key in the Trigger control area to see the trigger menu. 2. Press Mode to select Video. 3. Press Source to select CH1 as trigger source. 4. Press Polarity to select 5. Press Sync to select Line Num. 6. Turn ( . ) knob to trigger on a specified line number. 7. Adjust the to set trigger level at the video sync pulse to get a stable trigger. 8. Turn the horizontal knob to observe a complete waveform on the screen.
RIGOL Example 8: FFT Cursor Measurement FFT measurements include: Amplitude measurement (Vrms or dBVrms) and Frequency measurement (Hz). Do these steps as follows: 1. Press Cursor→Press No.1 button to set cursor mode to Manual. 2. Press No.2 button associated with Type to select X or Y. 3. Press button MATH→set operate mode as FFT. 4. Press Cursor →press No.3 button to specify source as MATH. 5. Turn ( ) knob to move the cursor A and cursor B to a point of interest.
RIGOL Figure 3-10 Cursor Measurement (Type X) © 2008 RIGOL Technologies, Inc.
RIGOL Example 9: Pass/Fail Test The Pass/Fail Test is one of enhanced special functions based on DS1000B series. In this test function the oscilloscope compares the input signal to the established waveform mask. If the waveform “touches” the mask, a “Fail” signal occurs, otherwise the test passes. When needed, a programmable output can be used for external automatic control applications. The output is built in as a standard feature, and is photo electricity isolated. Execute the steps as following: 1.
RIGOL Chapter 4 : Troubleshooting 1. (1) (2) (3) (4) After the oscilloscopes is powered on, the screen remains dark (no display): Check the power cord connection. Ensure the power switch is turned on. After the above inspection, restart the oscilloscope. If the problem still remains, please contact RIGOL for help. 2. (1) (2) (3) (4) (5) After the signal acquisition the waveform does not appear: Check the probes connected with the signals. Check the probes connected to the channels firmly.
RIGOL 7. The signal is displayed as ladder like waveform. (1) The time base setting maybe is too slow. Turn the horizontal knob to increase horizontal resolution to improve the display. (2) Maybe the display Type is set to “Vectors”, and set it to “Dots” mode to improve the display. 8. Unable to connect to PC or the PictBridge printer by USB: (1) When connected to PC, if the message “Unknown device” appears or prompting scanner or camera was found.
RIGOL Chapter 5 : Specifications All specifications apply to the DS1000B Series Oscilloscopes and a probe with the Attenuation switch set to 10X unless noted otherwise. To meet these specifications, two conditions must first be met: The instrument must have been operating continuously for thirty minutes within the specified operating temperature. Must perform the Self Calibration operation, accessible through the Utility menu, if the operating temperature changes by more than 5℃.
RIGOL Specifications Acquisition Sampling Modes Real-Time Equivalent-time [1] 50 GSa/s[2] Sampling Rate 2 GSa/s (half channel ) 1 GSa/s (each channel) Averages N time acquisitions, all channels simultaneously, N is selectable from 2, 4, 8, 16, 32, 64, 128 and 256 Inputs Input Coupling DC, AC, GND Input Impedance 1MΩ± 2%, in parallel with 18pF± 3pF Probe Attenuation Factors 0.001X, 0.01X, 0.
RIGOL Delta Time Measurement Accuracy (Full Bandwidth) Single-shot: ± (1 sample interval + 50ppm × reading + 0.6 ns) >16 averages: ± (1sample interval + 50ppm × reading + 0.
RIGOL Trigger Trigger Sensitivity 0.1div-1.0div (adjustable) Trigger Level Range Internal ± 6 divisions from center of screen EXT ± 0.6V EXT/5 ± 3V Internal ± (0.
RIGOL Reciprocal of ∆T in Hertz (1/∆T) Auto Measure Track Voltage value for Y-axis waveform Time value for X-axis waveform Auto Cursors are visible for Automatic Measurement Vpp, Vamp, Vmax, Vmin, Vtop, Vbase, Vavg, Vrms, Overshoot, Preshoot, Freq, Period, Rise Time, Fall Time, +Width, -Width, +Duty, -Duty, Delay A→B , Delay A→B ,Phase A→B ,Phase A →B [1] Half channel indicates selecting one of the channels in CH1 and CH2, or in CH3 and CH4.
RIGOL General Specifications Display Display Type 5.7 inch.
RIGOL IP Degree IP2X Calibration Interval The recommended calibration interval is one year © 2008 RIGOL Technologies, Inc.
RIGOL Chapter 6 : Appendixes Appendix A: Accessories Probe× 4 (1.5m), 1:1, (10:1) Passive Probes When the switch of probe is toggled to 1X attenuation setting, the probe has 6MHz bandwidth, according with 150V CAT II. When the switch of probe is set to 10X attenuation setting ,the probe‟s bandwidth equals to the upper limit of the oscilloscope, according with 300V CAT II. A Power Cord that fits the standard of destination country.
RIGOL Appendix B: Warranty RIGOL Technologies, Inc. warrants its products‟ mainframe and accessories in materials and techniques within the warranty period. During the period concerned, RIGOL guarantees the free replacement or repair of products which are approved defective. To get repair service or obtain a copy of the whole warranty statement, please contact with your nearest RIGOL sales and service office.
RIGOL Appendix C: Maintenance General Care Do not store or leave the instrument in which the LCD display will be exposed to direct sunlight for long periods of time. ! CAUTION: To avoid damage to the instrument or probes, do not expose them to sprays, liquids, or solvents. Cleaning If this instrument requires cleaning, disconnect it from all power sources and clean it with a mild detergent and water. Make sure the instrument is completely dry before reconnecting it to a power source.
RIGOL Appendix D: Contact RIGOL If you have any problem or requirement during using our products, please contact RIGOL Technologies, Inc. or the local distributors. Domestic: Please call Tel: (86-10) 8070 6688 Fax: (86-10) 8070 5070 Service & Support Hotline: 800 810 0002 9:00 am –5: 00 pm from Monday to Friday Or by e-mail: service@rigol.com Or mail to: RIGOL Technologies, Inc.
RIGOL Index 50% ........................................ 1-15 AC Coupling ............................... 2-4 AC Line .................................... 2-49 Accessories ................................ 6-2 Acquire .................................... 2-52 Acquisition ................................. 5-2 Alternative Trigger .................... 2-43 Appendix ................................... 6-1 AUTO .......................... 2-105, 2-106 Auto Trigger.............................
RIGOL Trigger Control ......................... 1-14 Trigger System ......................... 2-34 Troubleshooting .......................... 4-5 UTILITY ................................... 2-69 Vertical ...................................... 5-3 Vertical System ........................... 2-2 ii Vertical Window ....................... 1-10 Video Signal ............................. 3-10 Video Trigger ........................... 2-38 X-Y .......................................... 2-30 Y-T ...............