RIGOL User’s Guide DS1000B Series Digital Oscilloscopes DS1074B,DS1104B,DS1204B Aug. 2016 RIGOL TECHNOLOGIES, INC.
RIGOL Guaranty and Declaration Copyright © 2009 RIGOL TECHNOLOGIES, INC. All Rights Reserved. Trademark Information RIGOL is registered trademark of RIGOL TECHNOLOGIES, INC. Publication Number UGA04125-1210 Notices RIGOL products are covered by P.R.C. and foreign patents, issued and pending. RIGOL reserves the right to modify or change parts of or all the specifications and pricing policies at the company’s sole decision. Information in this publication replaces all previously released materials.
RIGOL Safety Notices Please review the following safety precautions carefully before putting the instrument into operation so as to avoid any personal injury or damage to the instrument and any product connected to it. To prevent potential hazards, please follow the instructions specified in this manual to use the instrument properly. Use Proper Power Cord. Only the exclusive power cord designed for the instrument and authorized for use within the local country could be used. Ground the Instrument.
RIGOL Do Not Insert Anything Into the Air Outlet. Do not insert anything into the air outlet to avoid damage to the instrument. Use Proper Fuse. Please use the specified fuses. Avoid Circuit or Wire Exposure. Do not touch exposed junctions and components when the unit is powered on. Do Not Operate With Suspected Failures. If you suspect that any damage may occur to the instrument, have it inspected by RIGOL authorized personnel before further operations.
RIGOL Keep it out of the reach of children. Improper change of a battery (lithium battery) may cause an explosion. Use the RIGOL specified battery only. Handle with Caution. Please handle with care during transportation to avoid damage to keys, knobs, interfaces, and other parts on the panels. The disturbance test of all the models meet the limit values of A in the standard of EN 61326: 1997+A1+A2+A3, but can’t meet the limit values of B.
RIGOL Safety Notices and Symbols Safety Notices in this Manual: WARNING Indicates a potentially hazardous situation or practice which, if not avoided, will result in serious injury or death. CAUTION Indicates a potentially hazardous situation or practice which, if not avoided, could result in damage to the product or loss of important data. Safety Terms on the Product: DANGER WARNING CAUTION It calls attention to an operation, if not correctly performed, could result in injury or hazard immediately.
RIGOL Allgemeine Sicherheits Informationen Überprüfen Sie diefolgenden Sicherheitshinweise sorgfältigumPersonenschädenoderSchäden am Gerätundan damit verbundenen weiteren Gerätenzu vermeiden. Zur Vermeidung vonGefahren, nutzen Sie bitte das Gerät nur so, wiein diesem Handbuchangegeben. Um Feuer oder Verletzungen zu vermeiden, verwenden Sie ein ordnungsgemäßes Netzkabel. Verwenden Sie für dieses Gerät nur das für ihr Land zugelassene und genehmigte Netzkabel. Erden des Gerätes.
RIGOL entsprechende Öffnungen stecken (Lüfter z.B.) Passende Sicherung verwenden. Setzen Sie nur die spezifikationsgemäßen Sicherungen ein. Vermeiden Sie ungeschützte Verbindungen. Berühren Sie keine unisolierten Verbindungen oder Baugruppen, während das Gerät in Betrieb ist. Betreiben Sie das Gerät nicht im Fehlerfall. Wenn Sie am Gerät einen Defekt vermuten, sorgen Sie dafür, bevor Sie das Gerät wieder betreiben, dass eine Untersuchung durch RIGOL autorisiertem Personal durchgeführt wird.
RIGOL Die richtige Verwendung desAkku. Wenneine Batterieverwendet wird, vermeiden Sie hohe Temperaturen bzw. Feuer ausgesetzt werden. Bewahren Sie es außerhalbder Reichweitevon Kindern auf. UnsachgemäßeÄnderung derBatterie (Anmerkung: Lithium-Batterie) kann zu einer Explosion führen. VerwendenSie nur von RIGOL angegebenenAkkus. Sicherer Transport. Transportieren Sie das Gerät sorgfältig (Verpackung!), um Schäden an Bedienelementen, Anschlüssen und anderen Teilen zu vermeiden.
RIGOL Sicherheits Begriffe und Symbole Begriffe in diesem Guide: WARNING Die Kennzeichnung WARNING beschreibt Gefahrenquellen die leibliche Schäden oder den Tod von Personen zur Folge haben können. CAUTION Die Kennzeichnung Caution (Vorsicht) beschreibt Gefahrenquellen die Schäden am Gerät hervorrufen können. Begriffe auf dem Produkt: DANGER WARNING CAUTION weist auf eine Verletzung oder Gefährdung hin, die sofort geschehen kann.
RIGOL General-Purpose Oscilloscopes The introduction of this document contains three types of DS1000B series oscilloscope: DS1074B, DS1104B and DS1204B RIGOL DS1000B series digital oscilloscopes with four analog channels in connection with an external trigger channel are designed to capture multiple signals and meet the needs of measurement along with its excellent performance, powerful functions.
RIGOL Built-in FFT function. Digital filters, includes LPF, HPF, BPF, BRF. Pass/Fail detection Function, optically isolated Pass/Fail output. Addition, Subtract and Multiply Mathematic Functions of waveforms. Advanced trigger types include: Edge, Video, Pulse width, Pattern, Alternative. Adjustable trigger sensitivity. Multiple Language User Interface. Pop-up menu makes it easy to read and easy to use. Built-in help systems with multinational languages.
RIGOL Document Overview Chapter 1 Quick Start To introduce the panls and menus to help users to realize the base operations. Chapter 2 Operating Your Oscilloscope More details to help users to understand the functions and measurements of instrument deeply. Chapter 3 Application Examples To show the measurement functions directly by application examples. Chapter 4 Troubleshooting Provide you troubleshooting to solve the problems during operating.
RIGOL Contents Guaranty and Declaration ........................................................................ I Safety Notices........................................................................................ II Safety Notices and Symbols .................................................................... V Allgemeine Sicherheits Informationen................................................... VI Sicherheits Begriffe und Symbole ..........................................................
RIGOL To To To To Set up the Sampling System ............................................................ 2-50 Set up the Display System ............................................................... 2-54 Store and Recall ............................................................................. 2-57 Set up the Utility ............................................................................ 2-66 The I/O Setup ..............................................................................
RIGOL Appendix A: DS1000B series Accessories ................................................. 6-1 Appendix B: Warranty ........................................................................... 6-2 Appendix C: Care and Cleaning .............................................................. 6-3 Index ......................................................................................................
RIGOL Chapter 1 Quick Guide This chapter covers the following topics: General Inspect Panels and User Interface Functional Check To Compensate the Probes To Display a Signal Automatically To Understand the Vertical System To Understand the Horizontal System To Understand the Trigger System To Understand the Quick Function User’s Guide for DS1000B Series 1-1
RIGOL General Inspect After receiving a new DS1000B series oscilloscope, please inspect the instrument as follows: 1. Inspect the shipping container for damage. Keep the damaged shipping container or cushioning material until the contents of the shipment have been checked for completeness and the instrument has passed both electrical and mechanical test. The consigner or carrier shall be liable for the damage to instrument resulting from shipment, without free maintenance or replacement by RIGOL. 2.
RIGOL Panels and User Interface Being familiar with the operation panels is the first priority when you get a new DS1000B series digital oscilloscope. This chapter will bring you a brief introduction and description of the panels and functions of oscilloscope, which can help users to know well about the instrument in a short time. Front Panel Figure 1-1, the front Panel; the knobs are used most often and are similar to the knobs on other oscilloscopes. There are five grey buttons defined as No.1 to No.
RIGOL Front Panel Interfaces: USB Host: Connect DS1000B with external USB devices when the oscilloscope is regarded as “Host Device”, such as connecting a USB flash device. Analog Signal Channels: Four channels are provided: CH1, CH2, CH3 and CH4. EXT TRIG Input: It is applied to input external trigger signal. NOTE: There are two USB Host ports respectively on the front and rear panel.
RIGOL Notation definitions in this Manual: Throughout this guide, notation symbols of buttons and knobs are the same of those on front-panel. A box around the name of the key denotes MENU function buttons on front-panel, such as Measure. ( denotes the five SCALE knobs. denotes the LEVEL knob. The name with a drop shadow denotes the menu operating key, such as Waveform soft key in Storage menu. Arrow “→” is used to divide operation steps.
RIGOL User Interface Run Status Timebase U-Disk Location of Waveform Connected Window in Memory Trigger Point in Memory Trigger Point in Waveform Window Trigger Symbol CH1 Menu CH2 CH3 Waveform Display Window CH4 CH1 Volts/div CH2 Volts/div CH3 Volts/div CH4 Volts/div Figure 1-3 User Interface Overview 1-6 User’s Guide for DS1000B series
RIGOL Functional Check Perform this quick functional check to verify that the instrument is operating correctly. 1. Turn on the instrument Use the power cord designed for the oscilloscope only to connect the instrument and AC power which delivers 100 to 240 VACRMS, 45Hz to 440Hz. Press the power switch to turn on the instruments, waiting until the display shows the waveform window. 2. Recall factory setting Press Storage → Storage → Factory → Recall, to recall factory setting.
RIGOL Figure 1-5 Probe connection Figure 1-6 Probe adjustment Attach the probe tip to the Probe compensator connector and the reference lead to the ground pin. Press Auto, within a few seconds, a square wave will display (approximately 1kHz, 3Vpp). Push the OFF button or push the CH1 button again to turn off Channel 1. Push the CH2, CH3, CH4 buttons to turn on other channels, repeat step 2&3. NOTE: Signal outputed from the Probe compensator is ONLY used for probe compensation, not for calibration.
RIGOL To Compensate the Probes Perform this adjustment to match the characteristics of the probe and the channel input. This should be performed whenever attaching a probe to any input channel at the first time. 1. From CH1 menu, set the Probe attenuation to 10X (press CH1→Probe→10X). Set the switch to 10X on the probe and connect it to CH1 of the oscilloscope. When using the probe hook-tip, inserting the tip into the hook-tip firmly to ensure a tight connection. 2.
RIGOL To Display a Signal Automatically The oscilloscope has an automated feature to display the input signal in best-fit status. The input signal should be 50 Hz or higher and a duty cycle is greater than 1%. Using the automatic setting: 1. Connect a signal to the oscilloscope (such as channel 1) as described above. 2. Press Auto. The oscilloscope may change the current settings to display the signal; adjusts the vertical and horizontal scaling, the trigger coupling, type, position, level, and mode.
RIGOL To Understand the Vertical System As shown in Figure 1-8, there are several buttons and knobs in the VERTICAL control area, which are used for setting Channels, MATH and REF functions as well as adjusting vertical position and Volts/div of waveforms. In addition, different colors are designed for different channels, corresponding to different waveforms, so as to observe clearly. Figure 1-8 Vertical control 1.
RIGOL Measurement hints If the channel is DC coupled, measuring the DC components of the signal by simply noting its distance from the ground symbol. If the channel is AC coupled, the DC component of the signal is blocked, allow you to use greater sensitivity to display the AC component of the signal.
RIGOL To Understand the Horizontal System Figure1-9 shows the HORIZONTAL controls: MENU button, and knobs of horizontal system. Following the exercise to familiarize with the buttons, knobs, and status bar. Figure 1-9 Horizontal control 1. knob. Set horizontal “s/div” by knob changes the sweep speed in a 1-2-5 step The horizontal sequence, and displays the value at the upper-left of the display. The time base ranges of the DS1000B series is from 2 ns/div* to 50 s/div.
RIGOL 3. The horizontal knob moves displayed signal horizontally on the waveform window. The horizontal knob adjusts the trigger offset of signal, when turning the knob; you can note that the waveform moves horizontally in conjunction with the knob. Horizontal offset back to 0 shortcut key Rotate the horizontal knob to adjust the horizontal position of the signal in the waveform window and press it down to restore the trigger offset or (delayed sweep offset) to the zero point.
RIGOL To Understand the Trigger System Figure 1-10 shows the TRIGGER control: MODE, MENU, FORCE, 50% and a knob. Following the exercises to familiarize with the buttons, trigger level knob and status bar. Figure 1-10 Trigger control 1. Three kinds of trigger modes can be switched among Auto, Normal and Single with pressing MODE button. 2. Use the knob to change the settings of trigger level.
RIGOL 3. Press MENU button in the Trigger control area. A soft button menu appears on the display showing the trigger setting choices as shown in Figure 1-11. Figure 1-11 Press NO.1 menu button to choose Edge. Press NO.2 menu button to choose Source as CH1. Press NO.3 menu button to choose Slope as Press NO.4 menu button to choose Sweep as Auto. Press NO.5 menu button to enter secondary menu of Set Up. . c.
RIGOL To Understand the Quick Function The following figure shows that there are two buttons in the QUICK function area. The exercise below will gradually conduct you to familiarize the settings of the buttons. Figure 1-12 The quick function menu 1. Measure There are two Measure buttons on the front panel, belonging to Function Menu area and Quick Function area respectively.
RIGOL Chapter 2 Operating Your Oscilloscope By now, a user should understand the VERTICAL, HORIZONTAL and TRIGGER control systems and know how to determine the system setup from the status bar of a DS1000B series digital oscilloscope. This chapter will go through all groups of front-panel buttons, knobs and menus; and further the knowledge of the operation by hints in this guide.
RIGOL To Set up the Vertical System Four analog channels are provided by DS1000B: CH1, CH2, CH3 and CH4. Press the corresponding button on the front panel to turn the channels on/off. The backlight indicates the channel is currently active. Press the button again to turn the channel off. When channel is currently selected, press OFF will turn it off and the backlight goes off.
RIGOL Knobs Explanation Use the vertical controls to display signal waveforms by adjusting the vertical knob, and the knob, and setting the input parameters. 1. Using vertical knob. The vertical control changes the vertical position of signal waveforms in all channels (excluding MATH and REF). The resolution changes according to the vertical level set. Pressing this knob will clear the channel’s vertical offset to zero. 2. Using vertical knob.
RIGOL Channel Settings Each channel of DS1000B series digital oscilloscope has an independent operation menu and it will pop-up after pressing any button among CH1, CH2, CH3, CH4. The menu to set CH1 is shown in the table below. Press CH1 to enter the following menus. Figure 2-1 Table 2-2 Channel menu (Page 1/2) Menu Coupling Settings Comments AC Block DC component of the input Signal. Pass both AC and DC components of the input signal Disconnect the input signal. DC GND ON BW Limit OFF Probe 0.
RIGOL Figure 2-2 Table 2-3 Channel menu (Page 2/2) Menu Settings Comments 2/2 Back to the previous menu page (The followings are the same, no more explanation). Volts/Div Coarse Fine Define a 1-2-5 sequence. To change the Volts/Dive settings in small steps between the coarse settings. Invert ON OFF Turn on the invert function. Restore to original display of the waveform. V/ A/ W/ U Set “V”, “A”, “W” or “U” as the unit of vertical channel.
RIGOL 1. To set up channel coupling To use Channel 1 as an example, input a square wave signal with DC shift. Press CH1→Coupling→DC to set “DC” coupling. Both the AC component and the DC component of the input signal will pass. DC Coupling Setup DC Coupling Status Symbol Figure 2-3 DC coupling settings Press CH1→Coupling→AC, to set “AC” coupling. It will pass AC component of the input signal with frequency higher than 5 Hz and restraint DC component of the input signal.
RIGOL Press CH1→Coupling→GND, to set “GND” coupling, it disconnects the input signal.
RIGOL 2. To set up the channel bandwidth limit Take Channel 1 for an example, input a signal containing high frequency component. Press CH1→BW Limit→OFF, to set up bandwidth limit to “OFF” status. The oscilloscope is set to full bandwidth and passing the high frequency component in the signal. Turn off the BW Limit Figure 2-6 Turn off the BW limit Press CH1→BW Limit→ON, to set up bandwidth limit to “ON” status. It will restraint the frequency component higher than 20MHz.
RIGOL 3. To set up Probe Attenuation The oscilloscope allows selecting the attenuation factor for the probe. The attenuation factor changes the vertical scaling of the oscilloscope so that the measurement results reflect the actual voltage levels at the probe tip. Figure 2-8 shows an example for using a 1000:1 probe and its attenuation factor.
RIGOL 4. Digital Filter Press CH1→ Digital Filter to display the digital filter menu. Turn Off Digital Filter Figure 2-9 Turn off digital filter Turn On Digital Filter Figure 2-10 Turn on digital filter Turn the digital filter on or off as shown in the following figure. Turn ( ) knob to adjust the upper and lower limit of frequency after turning on the digital filter.
RIGOL Figure 2-11 Table 2-5 The Filter menu Menu Settings Comments Digital Filter ON OFF Turn on the digital filter. Turn off the digital filter. Filter Type Setup Setup Setup Setup as as as as Upper Limit[1] Turn ( ) knob to set upper limit high limit. Lower Limit[1] Turn ( LPF (Low Pass Filter). HPF (High Pass Filter). BPF (Band Pass Filter). BRF (Band Reject Filter). ) knob to set lower limit.
RIGOL 5. To change the Volts/Div settings The Volts/Div control has Coarse or Fine configuration. The Vertical Sensitivity is 2mv/div - 10V/div. Coarse: It is the default setting of Volts/Div in a 1-2-5-step sequence from 2mV/div, 5mV/div, 10mV/div, to 10 V/div. Fine: This setting changes the vertical scale to small steps between the coarse settings. It will be helpful to adjust the waveform in smooth steps.
RIGOL 6. To invert a waveform Invert turns the displayed waveform 180 degrees with respect to the ground level. When the oscilloscope is triggered on the inverted signal, the trigger is also inverted. Figure 2-13 and Figure 2-14 show the changes before and after the inversion respectively.
RIGOL Math Function The mathematic functions include “add”, “subtract”, “Multiply” and “FFT” for CH1, CH2, CH3 and CH4. The mathematic result can be measured by the grid and the cursor. Press MATH button to enter “Math” menu, meanwhile, sign of math scale will appear under the screen. See figure below, waveform of CH1 is marked yellow, waveform of CH2 is marked blue, and waveform of CH1 adding CH2 is marked purple.
RIGOL Figure 2-17 Table 2-7 The Math menu (Page 2/2) Menu Reset User’s Guide for DS1000B Series Settings Comments The multifunctional knob ( ) adjusts the vertical position of the Math waveform. The multifunctional knob ( ) adjusts the vertical amplitude of the Math waveform. Reset the math waveform to vertical mid-point.
RIGOL 1. Using FFT function The FFT (Fast Fourier Transform) process converts a time-domain signal into its frequency components mathematically. FFT waveforms are useful in the following applications: Measuring harmonic content and distortion in systems Characterizing noise in DC power supplies Analyzing vibration Press MATH→Operate→FFT to enter FFT setting menu. Figure 2-18 Figure 2-19 Table 2-8 The FFT menu (Page 1/2) Menu Settings Comments Operate A+B A-B A×B FFT Add source A to source B.
RIGOL Key points for FFT 1. 2. 3. Signals that have a DC component or offset can cause incorrect FFT waveform component magnitude values. To minimize the DC component, choose AC Coupling on the source signal. To reduce random noise and aliases components in repetitive or single-shot events, set the oscilloscope acquisition mode to Average. To display FFT waveforms with a large dynamic range, use the dBVRMS scale. The dBVRMS scale displays component magnitudes using a log scale.
RIGOL 2. Selecting an FFT Window DS1000B series oscilloscope provides four FFT windows. Each window is a trade-off between frequency resolution and amplitude accuracy. It depends on the desired measurement and the source signals characteristics to determine the window to use. Use the following guidelines to select the best window. DS1000B provides four FFT functions with different features, which are needed to be selected according to measuring waveform.
RIGOL Term Explanations FFT Resolution: The FFT resolution is the quotient between sampling rate and the number of FFT points. With a fixed FFT points, the lower sampling rate, the better the resolution. Nyquist Frequency: The highest frequency that any real-time digitizing oscilloscope can acquire without aliasing. It’s normally half of the sample rate. This frequency is called the Nyquist frequency. Frequency above the Nyquist frequency will be under sampled, causing a situation known as aliasing.
RIGOL REF Function Reference Waveforms are saved waveforms to be selected for display. The reference function will be available after saving the selected waveform to non-volatile memory. During the actual measurement process, you can use DS1000B series digital oscilloscope to observe the waveforms of correlative components. It will help you to determine the cause of malfunction when you compare the measured waveform with the reference waveform.
RIGOL Figure 2-21 Table 2-12 REF menu when using internal/external memory (Page 2/3) Menu Settings Comments Save REF waveform to outer memory location. Save The multifunctional knob ( the vertical position of REF ) adjusts The multifunctional knob ( ) adjusts the vertical amplitude of REF Figure 2-22 Table 2-13 REF menu when using internal memory (Page 3/3) Menu Figure 2-23 settings comments Reset Reset REF waveform. Imp./Exp. Go to import menu (see table 2-19).
RIGOL 1. Import and Export Switch to external memory and select Internal location. Then, press REF →Imp. /Exp. and enter the following interface. Figure 2-24 Import and export interface Figure 2-25 2-22 Table 2-15 Imp. /Exp. Menu Menu Settings Comments Explorer Path Directory File Switch to Path, directory or file. Export Export the REF file from internal memory to external memory (see table 2-16). Import Import the REF file to internal memory. Delete File Delete file.
RIGOL 2. Export Switch to external memory and select Internal location. Then, press REF→Imp. /Exp.→ Export and go to the following interface. Figure 2-26 Export interface Figure 2-27 Table 2-16 The Export menu Menu Settings Comments Move the input focus point of files’ name up and down. Move the focus point to next location. To delete chosen letter. Save User’s Guide for DS1000B Series Execute the operation.
RIGOL 3. Save to External Memory Switch to external memory and select External location. Then, press REF→Save and enter the following interface. Figure 2-28 Save interface Figure 2-29 2-24 Table 2-17 The Save menu Menu Settings Comments Explorer Path Directories Files Switch among Path, Directories and Files. New File (Folder) Set up new file in Path and File. Set up new folder in directory. Delete File (Folder) Delete chosen file (Folder).
RIGOL 4. New File (or New Folder) Switch to external memory and select External location. Then, press REF→Save→New File (or New Folder) and go to the following interface. File Name Input Pane Switch Chinese and English Switch Capital On or Off Figure 2-30 File name in English inputting interface Figure 2-31 Table 2-18 The New File menu Menu Settings Comments Move the focus point of file name up and down. Move the focus point to the next location. To delete chosen letter.
RIGOL 5. Import Switch to external memory and select External location. Then, press REF→Import and enter the following interface. Figure 2-32 Import interface Figure 2-33 Table 2-19 The Import menu Menu Settings Comments Explorer Path Directories Files Switch among Path, Directories and Files. Import 2-26 Import the REF file into internal memory.
RIGOL 6. Display a Reference Waveform Figure 2-34 Reference waveform display Operation introductions: 1. 2. Press REF button to show the reference waveform menu. Press soft button 3 to select the reference channel: CH1, CH2, CH3, CH4 or MATH. 3. Turn vertical and vertical to adjust the REF waveform to a suitable position. Press soft button No.4 to select the save location of REF waveform. Enter to REF menu (Page2/3) Press soft button No.2 to save the waveform as REF. 4. 5.
RIGOL To Set up the Horizontal System Press MENU button to enter the horizontal setting system. Figure 2-35 Table 2-20 The Horizontal menu Menu Settings Comments Delayed ON OFF Turn on the Delayed Scan mode. Turn off the Delayed Scan mode. ScreenCenter TimeExpand Trigger User Y-T Time Base X-Y Roll[1] Trig-Offset Reset The waveform expands or compresses horizontally around the center of the screen when the horizontal time base is changed.
RIGOL Term Explanations Slow Scan Mode: This mode is available when the horizontal time base is set to 50ms or slower. In this mode, the oscilloscope acquires sufficient data for the left part to the trigger point, then wait for trigger, when trigger occurs, it continue to draw the rest part from the trigger point to the end of the right side. When choosing this mode to view low frequency signals, it is recommended that the channel coupling be set as DC.
RIGOL Knobs Explanation The horizontal controls change the horizontal scale and position of waveforms. The horizontal center of the screen is the time reference for waveforms. Changing the horizontal scale causes the waveform to be expanded or compressed with respect to the screen center. Horizontal position changes the displayed waveform position, relative to the trigger point. 1. The horizontal knob adjusts the horizontal position of all channels (including Math) waveforms.
RIGOL Delayed Scan The Delayed Scan is a magnified portion of the main waveform window. Use Delayed Scan to locate and horizontally expand part of the main waveform window for a more detailed (higher horizontal resolution) analysis of signal. The Delayed Scan time base setting cannot be set slower than the Main time base setting. Press MENU→Delayed→ON, or press horizontal scan function. See the waveform below.
RIGOL X-Y Format This format is useful for studying phase relationships between CH1 and ch2 (CH3 and CH4 can’t be used).CH1 in the horizontal axis(X) and CH2 in the vertical axis(Y), the oscilloscope uses a none-trigger acquisition mode, data is displayed as dots. Press MENU→Time Base→X-Y to enter X-Y mode. Figure 2-37 X-Y display format The oscilloscope can acquire waveforms with any sampling rate in Y-T mode (up to 2GSa/s for single channel or half channel [1]).
RIGOL To Set up the Trigger System The trigger determines when the oscilloscope starts to acquire data and display a waveform. When a trigger is set up properly, it can convert unstable displays or blank screens into meaningful waveforms. When the oscilloscope starts to acquire a waveform, it collects enough data so that it can draw the waveform to the left of the trigger point. The oscilloscope continues to acquire data while waiting for the trigger condition to occur.
RIGOL Edge Trigger An edge trigger determines whether the oscilloscope finds the trigger point on the rising or the falling edge of a signal. Select Edge trigger Mode to trigger on Rising edge, falling edge or rising & falling edge. Press MENU→Mode→Edge to enter the following menu. Figure 2-39 Table 2-21 The Trigger menu Menu Settings Comments Source CH1 CH2 CH3 CH4 EXT EXT/5 Select CH1 as trigger source. Select CH2 as trigger source. Select CH3 as trigger source. Select CH4 as trigger source.
RIGOL Pulse Width Trigger Pulse trigger occurs according to the width of pulse. The abnormal signals can be detected through setting up the pulse width condition. Press MENU→Mode→Pulse to enter the following menu. Figure 2-40 Table 2-22 The Trigger menu (Page 1/2) Menu Source Settings Comments CH1 CH2 CH3 CH4 EXT EXT/5 Select CH1 as trigger source. Select CH2 as trigger source.
RIGOL Video Trigger Choose video trigger to trigger on fields or lines of NTSC, PAL, or SECAM standard video signals. Trigger coupling preset to DC. Press MENU→Mode→Video to enter the following menu. Figure 2-42 Table 2-24 The Video Trigger menu (Page 1/2) Menu Settings Comments Source CH1 CH2 CH3 CH4 EXT EXT/5 Selects CH1 as trigger source. Selects CH2 as trigger source. Selects CH3 as trigger source. Selects CH4 as trigger source. Select EXT TRIG as trigger source.
RIGOL Figure 2-43 Table 2-25 The Video Trigger menu (Line Num, Page 2/2) Menu Line Num Standard Settings < Line sync > PAL/SECM NTSC Auto Sweep Normal Single Set Up Comments The range is from No.1 to No.525 when NTSC video standard is used and from No.1 to No.625 when PAL/SECAM video standard is used. Select Video standard. Acquire waveform even no trigger occurred. Acquire waveform when trigger occurred. When trigger occurs, acquire one waveform and then stop.
RIGOL Select “Line Synchronization”: Figure 2-45 Video Trigger: Line Synchronization 2-38 Select “Filed Synchronization”: Figure 2-46 Video Trigger: Field Synchronization User’s Guide for DS1000B series
RIGOL Pattern Trigger Pattern trigger recognises the trigger condition through inspecting the code preset which means the logic values or the logic combination.Every channel has its own logic value,such as,logic high value,logic low value,and the neglect value. Press MENU→Mode→Pattern to enter the following menu.
RIGOL Figure 2-48 Pattern trigger: set code as falling edge 2-40 User’s Guide for DS1000B series
RIGOL Alternative Trigger When alternative trigger is on, the trigger sources come from two vertical channels. This mode can be used to observe two non-related signals. You can choose two different trigger modes for the two vertical channels. The options are as follows: Edge, Pulse and video. The info of the trigger level of the two channels will be displayed on the upper-right of the screen. Press MENU→Mode→Alternative to enter the following menu.
RIGOL Figure 2-51 Table 2-30 The Alternative trigger menu (Page 1/2 Trigger Type: Pulse) Menu Settings Alt Channel CH1-CH2 CH1-CH3 CH1-CH4 CH2-CH3 CH2-CH4 CH3-CH4 Comments To specify the current alternative trigger channels as any two channels selected form CH1, CH2, CH3, CH4. select CH1, CH2, CH3, CH4 To select vertical channel among CH1, CH2, CH3, CH4 in alternative trigger mode so as to configure further settings. type Edge trigger Pulse trigger Video trigger Select trigger channel.
RIGOL Figure 2-53 Table 2-32 The Alternative trigger menu (Page 1/2 Trigger Type: Video) Menu Settings Comments Alt Channel CH1-CH2 CH1-CH3 CH1-CH4 CH2-CH3 CH2-CH4 CH3-CH4 To specify the current alternative trigger channels as any two channels selected form CH1, CH2, CH3, CH4. select CH1, CH2, CH3, CH4 To select vertical channel among CH1, CH2, CH3, CH4 in alternative trigger mode so as to configure further settings.
RIGOL Trigger Settings Set up different trigger settings according to different trigger modes. In Edge, Pulse and Pattern trigger modes. You need to set up coupling, sensitivity and trigger holdoff; in video trigger mode, to set up sensitivity and holdoff; in alternate trigger, different trigger types have different setting options.
RIGOL Figure 2-56 Table 2-35 Page 2/2 Settings for Edge, Pulse and Pattern Triggers Menu Holdoff Figure 2-57 Settings Reset comments Reset Holdoff time to 500ns. Table 2-36 Settings for Video Trigger Menu Sensitivity Holdoff Holdoff User’s Guide for DS1000B Series Settings Comments Set trigger sensitivity between 0.1div to 1.00div. Reset Set time slot before another trigger event between 100ns to 1.5s. Reset Holdoff time to 500ns.
RIGOL Trigger Holdoff You can use trigger Holdoff to stabilize a complex waveform, such as a pulse sequence. Holdoff time is the oscilloscope’s waiting period before starting a new trigger. During Holdoff, oscilloscope will not trigger until Holdoff ends. For instance: To trigger on the first pulse on a group of them, users can set the holdoff time to Pulse cluster width. Trigger Position Trigger Holdoff Time of Holdoff Figure 2-58 Trigger holdoff Operation Explanations: 1.
RIGOL Trigger Terms Explanation 1. Trigger Source: Trigger occurs from several sources: Input channels (CH1, CH2, CH3, CH4), AC Line, Ext, Ext/5. CH1, CH2, CH3, CH4: It is the most commonly used trigger source. The channel works when it is selected as a trigger source whatever displayed. Ext Trig: The instrument can be triggered from an external source while acquiring data from CH1, CH2, CH3, and CH4.
RIGOL display becomes stable on the screen. Any factor results in the instability of waveforms can be detected by Auto Trigger, such as the output of Power supply. NOTE: When horizontal control is set under 50 ms/div, Auto mode allows the oscilloscope not to capture trigger signal. Normal: The Normal mode allows the oscilloscope to acquire a waveform only when it is triggered. If no trigger occurs, the oscilloscope keeps waiting, and the previous waveform will remain on the display.
RIGOL 5. Adjustable trigger sensitivity To avoid the influence of noise from the physical world, and get the stable trigger, the trigger circuit has adopted Stickiness. In DS1000B series, the stickiness is adjustable from 0.1div-1.0div, which means when it sets to 1.0div; the trigger circuit will not affect any signal with peak-peak amplitude less than 1.0div, so as to avoid the influence of the noise.
RIGOL To Set up the Sampling System Figure 2-59 shows, the Acquire button at the MENU of the front panel. Acquire Setup Button Figure 2-59 The Acquire setup button Press Acquire button, the interface menu shows as follows: Figure 2-60 Table 2-37 The Acquire menu Menu Settings Comments Acquisition Normal Average Peak Detect Normal Acquisition mode. Average Acquisition mode. Peak Detect Acquisition mode. Averages[1] 2 to 256 Set average times by N-th power of 2 from 2 to 256.
RIGOL The waveform displayed on the screen will change in conjunction with the setting of Acquire menu. To reduce the displayed random noise, select the Average Acquisition. This mode would make the screen refresh slower. To Avoid signal aliasing, select Peak Detect Acquisition. Select Real-time acquisition to observe the single-shot or pulse signals. Select Equ-Time to observe high frequency repetitive signals.
RIGOL Figure 2-63 Adopt Normal acquisition Figure 2-64 Adopt peak detect acquisition, signal contains envelope Key points When the scope is acquiring waveforms, the Stop Acquisition: waveforms is in a live status; when acquisition is stopped, frozen waveform will be displayed, the position and scale can still be adjusted by vertical control and horizontal control.
RIGOL Key Points: Real-time Sampling: DS1000B has Real-time sampling rate up to 2GSa/s. At the time base 20ns or faster, the oscilloscopes use the sine(x)/x interpolation to expand the horizontal time base. Equivalent sampling: Known as Repetitive sampling to get up to 20ps of horizontal resolution (equivalent 50Gsa/s). This mode is good for observing repetitive signals, and it is not recommended for single-shot or pulse. Normal Acquisition: Oscilloscope acquires signal by equal time interval.
RIGOL To Set up the Display System The menu button for the display system on the front panel is shown in the following figure. Display Setup Button Figure 2-65 The display setup button Press the Display button, the interface menu for settings shows as follows. Figure 2-66 Table 2-38 The Display menu (Page 1/3) Menu Setting Comments Type Vectors Dots Display waveforms as vectors. Display waveforms as dots. Clear all existing waveforms from screen.
RIGOL Figure 2-67 Table 2-39 The Display menu (Page 2/3) Menu Settings Comments Grading ON OFF The waveform is displayed in multiple levels of pixel intensity. Display grids and coordinates on the screen. Turn off the grids. Turn off the grids and coordinates. Grid Menu Display Figure 2-68 1s 2s 5s 10s 20s Infinite Set the time before menu fades away. The menu will be hidden after the set time of last button pressing.
RIGOL Term Explanations Refresh rate: It is an important performance of digital oscilloscopes. It means the number of display refreshing per second and it will affect the ability to observe signal. Adjusting waveform intensity Default setup of multi-function knob ( 2-56 ) is adjusting waveform intensity.
RIGOL To Store and Recall Figure 2-69 shows the menu button for the storage system on the front panel. Storage Setup Button Figure 2-69 The storage setup button Press the Storage button, the interface menu for settings shows as follows. Figure 2-70 Table 2-41 The Storage menu (Waveform and Setups storage) Menu Settings Waveform Storage [1] Comments [2] Setups 8-Bitmap 24-Bitmap PNG CSV Factory Create, Save, Recall, Delete waveform files. Create, Save, Recall, Delete setups files.
RIGOL Figure 2-71 Table 2-42 The Storage menu (Bitmap storage) Menu Storage Para Save Figure 2-72 8-Bitmap 24-Bitmap ON OFF Comments Create or delete 8bits Bitmap files. Create or delete 24bits Bitmap files. Save the current oscilloscope settings in different format with the same file name. External Go to menu for external memory operation (see Table 2-47). Disk Mgr. Go to disk manage menu (see Table 2-48).
RIGOL Figure 2-73 Table 2-44 The Storage menu (CSV storage) Menu Settings Comments Storage CSV Create or delete CSV files. Displayed Save currently displayed waveform data to CSV file. Save the whole waveform data in memory to CSV file. Save the current oscilloscope settings in different format with the same file name. Go to menu for external memory operation (see Table 2-47). Go to disk manage menu (see Table 2-48). Data Depth Para Save Maximum ON OFF External Disk Mana.
RIGOL Internal Memory Press Storage→Internal to go to the following menu. Figure 2-75 Table 2-46 The Memory menu Menu Settings Comments Internal Int_00 (S) Int_01 (S) . . Int_09 (N) Set up the location of files in internal memory among Int_00 to Int_09. Thereinto, S in brackets indicates the location has storage file, N indicates has not. Recall Recall waveform files and setup files from the internal memory location. Save Save waveform files and setup files to the internal memory location.
RIGOL External Memory Press Storage→ External to go to the following interface. Path Back to Previous Folder File Current Folder File Info. Figure 2-76 External memory interface Figure 2-77 Table 2-47 The Memory menu Menu Settings Comments Explorer Path Directories Files Switch among Path, Directory and File. New File (Folder) To create new file or folder.(see table 2-18) Delete File (Folder) Delete file (Folder). Recall Recall waveform and setup from USB flash device.
RIGOL U-disk Selection Figure 2-78 U-disk selection DS1000B series have two USB Host ports on the front and rear panel. When both ports were inserted USB flash drive, the screen will appear a U-disk selection interface as Figure 2-78. Turn the multifunction knob to choose the driver. The front one marked “F:” and the rear one “G:”.
RIGOL Disk Management Press Storage→Disk Mgr.. Then go to the following menu. Figure 2-79 Figure 2-80 Table 2-48 The Disk Management menu (Page 1/2) Menu Settings Comments Explorer Path Directories Files Switch among Path, Directory and File. New folder To create new folder (same as new files, see table 2-18). Delete File Delete selected file. Recall Recall waveform, setup, recorded waveform, Pass/Fail file.
RIGOL 1. Disk Information Press Storage →Disk Mgr. →Disk info to go to the disk information interface. Figure 2-81 Disk information interface 2. Rename Press Storage→Disk Mgr.→Rename to go to the following interface.
RIGOL Figure 2-83 Table 2-50 The Rename menu Menu Settings Comments To move the input focus point of file name up and down. To move the focus point to the next location. To delete chosen letter. OK Rename the file. Key point Factory: The oscilloscope has default settings and can be recalled at anytime by user. Memory location: Specify the memory location to save/recall the waveforms and setups. Recall: Recall saved waveforms, setups and default settings. Save: Save waveforms and setups.
RIGOL To Set up the Utility Figure 2-69 shows the menu button for the utility system on the front panel. System Function Setup Button Figure 2-84 The Utility setup button Press the Utility button to show the menu of the settings in the Utility system. Figure 2-85 Table 2-51 The Utility menu (Page 1/3) Menu Setting I/O setting Sound Counter Comments Setup I/O configuration. (ON) (OFF) OFF ON Turn beeper sound on/off. Turn off Frequency Counter. Turn on Frequency Counter.
RIGOL Figure 2-86 Table 2-52 The Utility menu (Page 2/3) Menu Figure 2-87 Settings Comments Pass/Fail Setup Pass/Fail test. Record Setup Waveform Recorder. Print set Setup printing. Table 2-53 The Utility menu (Page 3/3) Menu Settings Comments Self-Cal Execute Self-calibration. System Info Show the following information: Serial number, software version, installed module. Preference Go to preference menu. Date/Time Set the real date and time for system.
RIGOL The I/O Setup Press Utility→IO Setting to go to the following menu. Figure 2-88 Table 2-54 The I/O Setup menu Menu Display Set network function, the local area network interface corresponds with the LXI standard. LAN Set USB Device GPIB# [1] Comments Auto Detect Computer PictBridge Set USB device function, connect USB interface to the needed device. 0~30 Set the GPIB address.
RIGOL 1. LAN Setting Access the oscilloscope into Network, and press LAN Set to enter the setting interface, see figure below: Recall the LAN Initialize setting Move the focus up and down Move the focus left and right User-defined area Figure 2-89 LAN Setting interface The items under the black background in the setting area are available for current setting. Enter the desired parameter type or parameters for those items with the knob press down or rotating.
RIGOL 2. Remote Control (1) By Web Page When the network configuration is completed, open the IE and enter the IP address of oscilloscope, it will enter the interface with relative information below. Figure 2-90 Remote control interface Click “Web Control” and set the corresponding parameters listed in the window below using the mouse.
RIGOL (2) By Commands DS1000B is a USB TMC (USB Test & Measurement Class) instrument, which could be controlled by commands on the basis of NI-VISA (NI-National Instrument, VISA-Virtual Instrument Software Architecture) or Agilent-VISA (for the details about commands, refer to the relative Programming Guide). USB or network is recommended for the communication between the instrument and a PC to integrate it into your testing system. The library related to VISA should be installed before using it.
RIGOL Figure 2-93 VISA TCP/IP Resource interface Click “Next” and select “Auto-detect of LAN Instrument”.
RIGOL The software searching the available network device automatically. Figure 2-95 Search the network device interface Click “Finish” and you will see the oscilloscope that has been accessed to the network from the subdirectory “VISA TCP/IP Resource” under the main Explorer interface. Figure 2-96 Subdirectory of VISA TCP/IP Resource Choose a relative resource and open the communication testing panel to send commands and read the data.
RIGOL Figure 2-97 Communication testing dialog Make a connection by USB Connect the oscilloscope to a PC by using an USB data cable. Then, turn on the oscilloscope, at the moment, the computer will pop up a Hardware Installation wizard and please install the “USB Test and Measurement Device” according to the wizards. After installation, open Measurement & Automation Explorer and you will see the information about USB device that has been connected in the subdirectory.
RIGOL Figure 2-98 Measurement & Automation Explorer interface Select the responding resource and open the VISA Test Panel to send commands and read data.
RIGOL Language The DS1000B series oscilloscopes have multi-language user menu, choose as your desire. Press Utility → Language to select the language. Press button No.4 until the desired language appears.
RIGOL Pass/Fail The Pass/Fail function monitors changes of signals by comparing the input signal which is within the pre-defined mask. The testing results not only can be displayed on screen, but also can be alarmed by opening the system sound settings. Press Utility→Pass/Fail to go to the following menu. Figure 2-101 Table 2-55 The Pass/Fail menu (Page 1/2) Menu Setting Comments Enable Test ON OFF Turn on Pass/Fail test. Turn off Pass/Fail test.
RIGOL 1. Mask setting Press Utility→Pass/Fail→Mask Setting to go to the following menu. Figure 2-103 Table 2-57 The Mask setting menu (Page 1/2) Menu X Mask Y Mask Settings Comments Set horizontal clearance to the waveform (0.04div-4.00div). Set vertical clearance to the waveform (0.04div-4.00div). Create Mask Location Figure 2-104 Internal External Set the memory location of the mask files.
RIGOL Figure 2-105 Table 2-59 The Mask setting menu (External, Page 2/2) Menu 2. Settings Comments Save Go to save menu (same as REF save menu, see table 2-16). Recall Go to recall menu (see Table 2-63). Import Go to import menu. (Same as REF import menu, see table 2-18). Recall Press Utility→Pass/Fail→Mask Setting→Recall to recall the setting file from internal/external memory.
RIGOL Waveform Recorder Waveform recorder records input waveforms from CH1, CH2, CH3 and CH4 with a maximum record length of 800 frames. This performance can also be activated by the Pass/Fail test, which makes this function especially useful to capture abnormal signals in long term without keeping an eye watching it. 1. 2. OFF: It will close all record functions. Record: To record waveform by selected time interval till the last frame. Press Utility→Record→Mode→Record to go to the following menu.
RIGOL Figure 2-108 Table 2-62 The Record menu (Page 2/2) Menu Interval 3. Settings Comments <1.00ms-1000s> Set time interval between record frames. Play Back: Play back the recorded waveforms. Figure 2-109 Table 2-63 The Play Back menu (Page 1/2) Menu Settings (Run) Operate (Stop) Figure 2-110 Comments Play stopped, press to Start playback. Press to stop playing. Play mode Set repeat play mode. Set single time play mode. Interval Set up interval value between frames. <1.
RIGOL Run/Stop Button The Run/Stop button can also replay or continue the waveform display. The Current Frame will go forward one frame by pressing Single button in Play back mode. 4. Storage: Store recorded waveforms in non-volatile memory according to the setup frames. Figure 2-111 Table 2-65 The Storage menu (Page 1/2) Menu Start Frame End Frame Location Figure 2-112 <1-800> <1-800> Internal external Comments Set first frame to be saved. Set last frame to be saved. Set up Store location.
RIGOL Figure 2-113 Table 2-67 The Storage menu (External location Page 2/2) Menu Settings Comments Save Go to save menu (same as REF save menu. see table 2-160). recall Go to recall menu (see Table 2-63). Import Go to import menu. (Same as REF import menu, see table 2-18).
RIGOL Print Setting DS1000B series oscilloscopes support two print modes: “normal” and “PictBridge”. Please execute print operation according to the following steps: 1. Connect printer: If use PictBridge printer, please select “PictBridge” mode, and then connect the oscilloscopes regarded as “Device equipment” and printer via USB Device port on the rear panel. 40 US C Digital Oscilloscope 100-127V, 45-440Hz 100-240V, 45-65Hz LINE60WMAX Fuse: 250V, T2A RIGOL Technologies, Inc.
RIGOL 2. To set up print parameters: Press Utility→Print set to into a setting menu. Please setup the printer type and the print mode as the case may be, or else the instrument will refuse printing if they can’t match well. Press PrintMod button, choose the print mode as Normal, and navigate into the following menu. Figure 2-116 Table 2-68 Print Set menu (Normal mode) Menu Settings Comments PrintMod Normal PictBridge Set the print mode as normal. Set the print mode as PictBridge.
RIGOL Figure 2-118 Table 2-70 Print setup menu (PictBridge mode, Page 2/4) Menu Paper size File type copies Figure 2-119 Settings Default, A2, A3, A4, A5, A6, B5 Default Exif/Jpeg Bmp 1~999 Print Quality Data print Settings Default Normal Draft fine Default ON OFF ON inverted 2-86 Select the paper size for printing. To specify the image type for printing as Exif/Jpeg or Bmp format. To set the print copies from 1to 999.
RIGOL 3. Print: Make sure the Printer has been connected and the print settings are all ready before printing. Then, press Utility → Print or press operation, If you want to go on printing, please use the User’s Guide for DS1000B Series to execute print button.
RIGOL Self-Calibration The Self-Calibration adjusts the internal circuitry to get the best accuracy. Use this function to calibrate the vertical and horizontal systems. For maximum accuracy at any time, run this calibration if the ambient temperature changes by 5℃ or more. Before running this procedure, make sure that disconnect any probes or cables from all channel inputs, otherwise failure or damage to the oscilloscope may occur. Then, press the Utility → Self-Cal.
RIGOL System Information Press Utility→System Info to view the system information, including Model, Serial Number, Software version and installed module.
RIGOL Preference Press Utility→Preference to go to the following menu Figure 2-123 Table 2-73 The Preference menu Menu Display Comments Screen saver 1 min … 5 hour OFF Set up screen saver timer. Expand Ground Center Set up waveform vertical expand reference. Stickykey Turn the sticky status of CH1, CH2, CH3, CH4, MATH, REF, Trig.Lev. and Trig.Pos. on or off. Default Load Use the last setup in the next power on. Use the default setup of the system in the next power on.
RIGOL Date and Time DS1000B Series provide build-in system real-time date and time. Users can set the date and time as shown in the following figure.
RIGOL To Measure Automatically The Measure button in the menu area activates the automatic measurement function. The instruction below shows the powerful measurement function of DS1000B series oscilloscopes.
RIGOL Figure 2-128 Table 2-76 The Measure menu (Page 2/2) Menu Settings Display All OFF ON QuickMea Delay/Phase Setup User’s Guide for DS1000B Series Comments Turn off all measurements result. Turn on all measurements result. To specify three quick measure items which you can choose among the above 22 measurements. To configure the delay and phase channels, in order to measure delay and phase value relative to time measurement between any two vertical channels form CH1 to CH4.
RIGOL Quick Measurement Setup Press Measure→QuickMea button, go to the quick measure menu。 Figure 2-129 Table 2-77 Quick measure menu Menu Item1 Item2 Item3 2-94 Settings Comments To specify Item1, Item2 and Item3 respectively, each item can be chosen from 22 kinds of measure items.
RIGOL Delay/Phase Setup Press Measure→Delay/Phase Setup button, and go to the delay/phase setup menu. Figure 2-130 Table 2-78 The delay/phase setup menu.
RIGOL Voltage Measurements The DS1000B series oscilloscopes provide automatic voltage measurements including Vpp, Vmax, Vmin, Vavg, Vamp, Vrms, Vtop, Vbase, Overshoot and Preshoot. Figure 2-131 below shows a pulse with some of the voltage measurement points. Overshoot Vmax Vtop Vpp Vamp Vbase Vmin Preshoot Figure 2-131 Voltage parameters Vpp: Peak-to-Peak Voltage. Vmax: The maximum amplitude. The most positive peak voltage measured over the entire waveform. Vmin: The minimum amplitude.
RIGOL Press Measure→Voltage button, and go to the following menu. Figure 2-132 Table 2-79 The Voltage Measurement menu (Page 1/3) Menu Figure 2-133 Comments Vmax Measure maximum waveform. voltage of a Vmin Measure minimum waveform. voltage of a Vpp Measure Peak-to-Peak Voltage. Vtop Measure a flat top voltage of a square waveform. Table 2-80 The Voltage Measurement menu (Page 2/3) Menu Figure 2-134 Settings Settings Comments Vbase Measure a flat base voltage of a square waveform.
RIGOL Time Measurements The DS1000B series oscilloscopes provide 12 kinds of timing parameters auto-measurements: Frequency, Period, Rise Time, Fall Time, +Width, -Width, Delay A→B , Delay A→B , +Duty and –Duty, Phase A→B ,Phase A→B . Figure 2-135 shows a pulse with some of the time measurement points. Fall Time Rise Time 90% 50% 10% +Width -Width Figure 2-135 Time parameters Rise Time: Time that the rising edge of the first pulse in the waveform takes to rise from 10% to 90% of its amplitude.
RIGOL Press Measure→Time button, and go to the following menu Figure 2-136 Table 2-82 The Time Measurement menu (Page 1/3) Menu Figure 2-137 Comments Period Measure Period of a waveform. Freq Measure Frequency of a waveform. Rise time Measure Rise Time of a rising edge. Fall time Measure Fall Time of a falling edge. Table 2-83 The Time Measurement menu (Page 2/3) Menu Figure 2-138 Settings Settings Comments +Width Measure +Pulse Width of a pulse wave.
RIGOL Key Points: 1. Select the signal channel for measuring. CH1, CH2, CH3, CH4 according to the interest. Press soft buttons as follows: Measure→Source→CH1, CH2, CH3 , CH4. 2. To see all measurement values, Press No.2 button to set the Display All to ON. 18 measurement parameters will be displayed on the screen. 3. Select parameters page for measuring; press No.2 or No.3 button to select voltage or time parameters pages by pressing soft button as follows: Measure →Voltage or Time →Vmax, Vmin…… 4.
RIGOL To Measure with Cursors Figure 2-140 shows the Cursor button on the front-panel. Cursor Measurement Figure 2-140 The Cursor setup button The cursor measurement has three modes: Manual, Track and Auto Measure. 1. Manual In this mode, the screen displays two parallel cursors. Move the cursors to make custom voltage or time measurements of the signal. The values are displayed on the boxes below the menu. Before using cursors, make sure to set the Signal Source as the channel for measuring. 2.
RIGOL Manual Mode Press Cursor→Mode→Manual to display the manual mode menu. Figure 2-141 Table 2-85 The Cursors menu Menu Mode Settings Manual X Type Y Source CH1 CH2 CH3 CH4 MATH Comments Adjust the cursor to parameters manually. measure X/Y Shown as vertical line to measure the time/frequency values. Shown as horizontal line to measure the voltage values. Select the measurement signal source.
RIGOL Table 2-86 The Cursor menu Cursor Increment X Cursor A Y X Cursor B Y Operation Turn the multifunctional knob ( cursor A horizontally. Turn the multifunctional knob ( cursor A vertically. ) to move Turn the multifunctional knob ( cursor B horizontally. Turn the multifunctional knob ( cursor B vertically. ) to move ) to move ) to move NOTE: Cursor could be moved only when the curse function menu is displayed. 5.
RIGOL Track Mode Press Cursor→Mode→Track to display the manual mode menu. Figure 2-142 Table 2-87 Cursor descriptions Menu Settings Mode Track Comments Set Track mode in cursor measurement. Cursor A CH1 CH2 CH3 CH4 MATH None Set Cursor A in conjunction Set Cursor A in conjunction Set Cursor A in conjunction Set Cursor A in conjunction Set Cursor A in conjunction Turn off Cursor A i with with with with with CH1. CH2. CH3. CH4. MATH.
RIGOL Table 2-88 The Cursor usage Cursor Operation Cursor A Turn the multifunctional knob ( ) to move cursor A horizontally. Cursor B Turn the multifunctional knob ( ) to move Cursor B horizontally. NOTE: Moving cursor horizontally is not allowed when other (not tracking cursor) menu is activated. 4. To get measurement values: Position of Cursor A: Time cursor centered on the midpoint of screen; Voltage cursor centered on channel ground level.
RIGOL Auto mode Press Cursor→Mode→Auto to display the manual mode menu. Figure 2-143 Table 2-89 The Mode setting Menu Mode Settings Auto Comments Display the cursors for the current automatic measuring. (See the following figure). There will be no cursor display if no parameters are chosen in Measure menu. The oscilloscope could move cursor automatically to measure 22 parameters in Measure menu.
RIGOL To Use Run Control Buttons The RUN control buttons include Auto (auto setting), Run/Stop and Single. Auto: The Auto features automatic adjustments to produce a stable display of the input signal. Press Auto button, the following menu appears. Figure 2-145 Table 2-90 The Auto menu Menu Settings Comments Multi-Cycle Press to display multi-cycle waveform on the screen. Single Cycle Press to display single cycle waveform on the screen.
RIGOL Auto-Set functions When the Auto is pressed the oscilloscope is configured to the following defaults: Table 2-91 The Auto menu Menu Settings Display format Y-T Acquire mode Normal Vertical Position Adjusted to right position Vertical “V/div” Adjusted to right scale Volts/Div Coarse Bandwidth limit Full Signal Invert OFF Horizontal position Center Horizontal “S/div” Adjust to right position. Trigger type Edge Trigger source Find the channel with input signal automatically.
RIGOL Chapter 3 Application Examples Example 1: Taking Simple Measurements The function is used to observe an unknown signal; to display, measure frequency, and peak-to-peak amplitude. 1. (1) (2) (3) To quickly display a signal, please do the steps as follows: Set the probe and the channel attenuations to 10X. Connect signal to CH1 with the probe. Press the Auto button. The oscilloscope sets the vertical, horizontal, and trigger controls at the best status automatically.
RIGOL Example 2: View a Signal Delay Caused by a Circuit This example is to test the input and output signals of a circuit and observe the signal delay. First, set the probe and the channel attenuation to 10X and connect CH1 probe to the input, CH2 to the output of the circuit. Do these steps as follow: 1. Display the signals (CH1 and CH2): (1) Press the Auto button. (2) Adjust the vertical and the horizontal scale by turning the set appropriate ranges for display.
RIGOL Example 3: Capture a Single-Shot Signal To capture a single event, it needs to gather some pre-test knowledge of the signal in order to set up the trigger level and slope correctly. For example, if the event is derived from TTL logic, a trigger level of 2 volts should work on a rising edge. The following steps show how to use the oscilloscope to capture a single event. 1. Set the probe and the channel attenuations to 10X. 2. Set up the trigger.
RIGOL Example 4: To Reduce the Random Noise on a Signal If the signal is noisy (Figure 3-2), set up the oscilloscope to reduce the noise on the waveform and avoid its interference to the signal. Figure 3-2 Signal with noises 1. Set the probe and the channel attenuations to 10X. 2. Connect a signal to the oscilloscope and obtain a stable display. 3. Improve the trigger by setting the Coupling. (1) Press the MENU in the Trigger control area. (2) Press Set Up→Coupling and select “LF Reject”.
RIGOL 4. To reduce the noise by setting the acquisition type and adjust the waveform intensity. (1) If there is noise within the signal and the waveform looks too wide, in this case, choose average acquisition. In this mode the waveform will be thin and easy to observe and measure. To use average acquisition follows these steps. Press soft button as Acquire→Acquisition→Average Toggle the Averages soft button to select the number of averages that best eliminates the noise from the displayed waveform.
RIGOL Example 5: Making Cursor Measurements There are 22 build-in automatic measurements. They can also be conducted using cursors to make time and voltage measurements of a waveform quickly. Measure the Peak Frequency of the First Sinc Waveform To 1. 2. 3. 4. 5. measure the rising frequency at the rising edge of a signal, do these steps: Press Cursor key to see the Cursor menu. Press Mode to set Manual mode. Press Type to select X. Turn ( ) knob to place cursor A on the first peak of the wave.
RIGOL Measure the Amplitude of the First Waveform Peak of the Sinc. Please follow these steps: 1. 2. 3. 4. 5. Press Cursor key to see the Cursor menu. Press Mode to set Manual mode. Press Type to select Y. Turn ( ) knob to place cursor A on the first peak of the wave. Turn ( ) knob to place cursor B on the second peak of the wave.
RIGOL Example 6: The Application of the X-Y operation Viewing Phase Changes through a Network Theme: Connect the oscilloscope to monitor the input and output of the circuit and capture the phase changes. To view the input and output of the circuit in an X-Y display, do these steps: 1. From the probe menu set the attenuation to 10X. Set the switch to 10X on the probes. 2. Connect the CH 1 probe to the input of the network, and connect the CH 2 probe to the output. 3.
RIGOL Signal must be centered in X D A B C Figure 3-6 Ellipse method to observe the phase difference Sinθ = A/B or C/D, where θ = phase shift (in degrees) between the two signals. From the formula above, you could get: θ = ±arcsine (A/B) or ±arcsine (C/D) θ must be in the range of (0~π/2) or (3π/2~2π) if the main axis of the ellipse is at I and III quadrant, If the main axis is at II and IV quadrant, θ must be in the range of (π/2~π) or (π~3π/2).
RIGOL Example 7: Triggering on a Video Signal Test a video circuit in the DVD set. Use video trigger to obtain a stable display. Triggering on Video Fields To trigger on the video fields, please do as the following steps: 1. Press the MENU key in the Trigger control area to see the Trigger menu. 2. Press Mode to select Video. 3. Press Source to select CH1 as trigger source. 4. Press Polarity to select . 5. Press Sync as Odd Field or Even Field. 6. Adjust the stable trigger.
RIGOL Triggering on the Video Lines 1. 2. 3. 4. 5. 6. Press the MENU key in the Trigger control area to see the trigger menu. Press Mode to select Video. Press Source to select CH1 as trigger source. Press Polarity to select . Press Sync to select Line Num. Turn ( ) knob to trigger on a specified line number. 7. Adjust trigger. to set trigger level at the video sync pulse to get a stable 8. Turn the horizontal screen.
RIGOL Example 8: FFT Cursor Measurement FFT measurements include: Amplitude measurement (VRMS or dBVRMS) and Frequency measurement (Hz). Do these steps as follows: 1. Press Cursor→Press No.1 button to set cursor mode to Manual. 2. Press No.2 button associated with Type to select X or Y. 3. Press button MATH→set operate mode as FFT. 4. Press Cursor →press No.3 button to specify source as MATH. 5. Turn ( ) knob to move the cursor A and cursor B to a point of interest.
RIGOL Example 9: Pass/Fail Test In Pass/Fail Test, the oscilloscope compares the input signal to the established waveform mask. Execute the steps as following: 1. Press Utility→Pass/Fail. 2. 3. Press Enable Test and select ON. Press Mask Setting and then: Press Recall to a the saved mask or press X Mask and Y Mask to adjust the horizontal limit and vertical limit with the multifunctional knob then press Create Mask to create a new mask.
RIGOL Chapter 4 Troubleshooting 1. (1) (2) (3) (4) (5) After the oscilloscopes is powered on, the screen remains dark (no display): Check the power cord connection. Ensure the power switch is turned on. Check whether the fuse burns out or not. After the above inspection, restart the oscilloscope. If the problem still remains, please contact RIGOL for help. 2. (1) (2) (3) (4) After the signal acquisition the waveform does not appear: Check the probes connected with the signals.
RIGOL waveform on screen. 6. After the Acquisition is set to “Averages” or Display Persistence is set ON, the waveform refreshes slowly. It is normal in these settings. 7. The signal is displayed as ladder like waveform. (1) The time base setting maybe is too low. Turn the horizontal knob to increase horizontal resolution to improve the display. (2) Maybe the display Type is set to Vectors, and set it to Dots mode to improve the display. 8.
RIGOL Chapter 5 Specifications All specifications apply to the DS1000B Series Oscilloscopes unless noted otherwise. To meet these specifications, two conditions must first be met: The instrument must have been operating continuously for thirty minutes within the specified operating temperature. Must perform the Self Cal operation, accessible through the Utility menu, if the operating temperature changes by more than 5°C. NOTE: All specifications are guaranteed unless noted “typical”.
RIGOL Technical Specifications Acquisition Sampling Modes Real-Time Equivalent-time [1] 50 GSa/s[2] Sampling Rate 2 GSa/s (half channel ) 1 GSa/s (each channel) Averages N time acquisitions, all channels simultaneously, N is selectable from 2, 4, 8, 16, 32, 64, 128 and 256 Inputs Input Coupling DC, AC, GND Input Impedance 1MΩ±2.0% The input capacity is 18pF±3pF Probe Attenuation Factors 0.001X, 0.01X, 0.
RIGOL Vertical A/D converter 8-bit resolution, each channel samples simultaneously Volts/div Range 2mV/div-10V/div at input BNC Offset Range ±40V(250mV/div ~10V/div) ±2V(2mV/div~245mV/div) Analog Bandwidth 70MHz(DS1074B) 100MHz(DS1104B) 200MHz(DS1204B) Single-shot Bandwidth 70MHz(DS1074B) 100MHz(DS1104B) 200MHz(DS1204B) Selectable Analog Bandwidth Limit 20MHz (typical) Lower Frequency Limit (AC -3dB) ≤5Hz (at input BNC) Rise Time at BNC, typical <1.75ns, <3.
RIGOL Trigger Sensitivity 0.1div-1.0div (adjustable) Trigger Level Range Internal ±6 divisions from center of screen EXT ±1.2V EXT/5 ±6V Internal ±(0.
RIGOL Time value for X-axis waveform Auto Auto Measure Cursors are visible for Automatic Measurement Vpp, Vamp, Vmax, Vmin, Vtop, Vbase, Vavg, Vrms, Overshoot, Preshoot, Freq, Period, Rise Time, Fall Time, +Width, -Width, +Duty, -Duty, Delay A→B , Delay A→ B , Phase A→B ,Phase A→B NOTE: [1] Half channel indicates selecting one of the channels in CH1 and CH2, or in CH3 and CH4.
RIGOL General Specifications Display Display Type 5.7 inch.
RIGOL Chapter 6 Appendixes Appendix A: DS1000B series Accessories Standard Accessories Four Passive Probes: PVP2150 for DS1074B/DS1104B PVP2350 for DS1204B A Power Cord that fits the standard of destination country. An USB Cable A Quick Guide Optional Accessories: BNC Cable USB-GPIB Adapter 50Ω Impedance Adapter DS1000B special convenient soft bag All accessories (standard and optional) are available by contacting your local RIGOL office.
RIGOL Appendix B: Warranty RIGOL warrants that the product mainframe and product accessories will be free from defects in materials and workmanship within the warranty period. If a product proves defective within the respective period, RIGOL guarantees free replacement or repair of any defective products within a reasonable period of time. To get repair service, please contact with your nearest RIGOL sales or service office.
RIGOL Appendix C: Care and Cleaning Care Do not store or leave the instrument where it may be exposed to direct sunlight for long periods of time. Cleaning Clean the instrument regularly according to its operating conditions. 1. Disconnect the instrument from all power sources. 2. Clean the external surfaces of the instrument with a soft cloth dampened with mild detergent or water. When cleaning the LCD, take care to avoid scarifying it.
RIGOL Index 50%........................................ 1-16 AC Coupling ............................... 2-6 AC Line ................................... 2-45 Accessories ................................ 6-2 Acquire.................................... 2-48 Acquisition ................................. 5-2 Alternative Trigger .................... 2-39 Appendix ................................... 6-1 AUTO .................................... 2-106 Auto Trigger ............................
RIGOL Vertical System .......................... 2-2 Vertical Window ....................... 1-11 2 Video Signal ............................. 3-10 Video Trigger ........................... 2-35 User’s Guide for DS1000B Series .