RIGOL Quick Guide MSO1000Z/DS1000Z Series Digital Oscilloscope Jun. 2015 RIGOL Technologies, Inc.
RIGOL Guaranty and Declaration Copyright © 2013 RIGOL Technologies, Inc. All Rights Reserved. Trademark Information RIGOL is a registered trademark of RIGOL Technologies, Inc. Publication Number QGA19105-1110 Notices z z z z z z RIGOL products are covered by P.R.C. and foreign patents, issued and pending. RIGOL reserves the right to modify or change parts of or all the specifications and pricing policies at company’s sole decision.
RIGOL Safety Requirement General Safety Summary Please review the following safety precautions carefully before putting the instrument into operation so as to avoid any personal injury or damage to the instrument and any product connected to it. To prevent potential hazards, please use the instrument only specified by this manual. Use Proper Power Cord. Only the power cord designed for the instrument and authorized for use within the local country could be used. Ground the Instrument.
RIGOL Do Not Operate in Wet Conditions. In order to avoid short circuiting to the interior of the device or electric shock, please do not operate the instrument in a humid environment. Do Not Operate in an Explosive Atmosphere. In order to avoid damage to the device or personal injuries, it is important to operate the device away from an explosive atmosphere. Keep Product Surfaces Clean and Dry.
RIGOL Safety Terms and Symbols Terms Used in this Manual. These terms may appear in this manual: WARNING Warning statements indicate conditions or practices that could result in injury or loss of life. CAUTION Caution statements indicate conditions or practices that could result in damage to this product or other property. Terms Used on the Product.
RIGOL Allgemeine Sicherheits Informationen Überprüfen Sie diefolgenden Sicherheitshinweise sorgfältigumPersonenschädenoderSchäden am Gerätundan damit verbundenen weiteren Gerätenzu vermeiden. Zur Vermeidung vonGefahren, nutzen Sie bitte das Gerät nur so, wiein diesem Handbuchangegeben. Um Feuer oder Verletzungen zu vermeiden, verwenden Sie ein ordnungsgemäßes Netzkabel. Verwenden Sie für dieses Gerät nur das für ihr Land zugelassene und genehmigte Netzkabel. Erden des Gerätes.
RIGOL Nicht in feuchter Umgebung betreiben. Zur Vermeidung von Kurzschluß im Geräteinneren und Stromschlag betreiben Sie das Gerät bitte niemals in feuchter Umgebung. Nicht in explosiver Atmosphäre betreiben. Zur Vermeidung von Personen- und Sachschäden ist es unumgänglich, das Gerät ausschließlich fernab jedweder explosiven Atmosphäre zu betreiben. Geräteoberflächen sauber und trocken halten.
RIGOL Sicherheits Begriffe und Symbole Begriffe in diesem Guide. Diese Begriffe können in diesem Handbuch auftauchen: WARNING Die Kennzeichnung WARNING beschreibt Gefahrenquellen die leibliche Schäden oder den Tod von Personen zur Folge haben können. CAUTION Die Kennzeichnung Caution (Vorsicht) beschreibt Gefahrenquellen die Schäden am Gerät hervorrufen können. Begriffe auf dem Produkt.
RIGOL Measurement Category Measurement Category MSO1000Z/DS1000Z series digital oscilloscopes can make measurements in Measurement Category I. WARNING This oscilloscope can only be used for measurements within its specified measurement categories. Measurement Category Definitions Measurement category I is for measurements performed on circuits not directly connected to MAINS. Examples are measurements on circuits not derived from MAINS, and specially protected (internal) MAINS derived circuits.
RIGOL Working Environment Temperature Operating: 0ć to +50ć Non-operating: -40ć to +70ć Humidity 0ć to +30ć: % relative humidity +30ć to +40ć: % relative humidity +40ć to +50ć: % relative humility WARNING To avoid short circuit inside the instrument or electric shock, please do not operate in humid environment. Altitude Operating: below 3 km Non-operating: below 15 km Installation (Overvoltage) Category This product is powered by mains conforming to installation (overvoltage) category II.
RIGOL Safety Class Class 1 – Grounded Product General Care and Cleaning General Care Do not store or leave the instrument where it may be exposed to direct sunlight for long periods of time. Cleaning Clean the instrument regularly according to its operating conditions. To clean the exterior surface, perform the following steps: 1. Disconnect the instrument from all power sources. 2. Clean the loose dust on the outside of the instrument with a lint-free cloth (with a mild detergent or water).
RIGOL Document Overview This manual is used to guide users to quickly get familiar with the front panel, rear panel, user interface and basic operation method of MSO1000Z/DS1000Z series digital oscilloscope. You can download the newest version of the manual from RIGOL official website (www.rigol.com). Format Conventions in this Manual: 1. Key The front panel keys are denoted by the format of "Key Name (Bold) + Text Box". For example, Utility denotes the "Utility" key. 2.
RIGOL Contents Guaranty and Declaration ................................................................................................I Safety Requirement ....................................................................................................... II General Safety Summary .............................................................................................. II Safety Terms and Symbols...........................................................................................
RIGOL Quick Start General Inspection 1. Inspect the shipping container for damage. Keep the damaged shipping container or cushioning material until the contents of the shipment have been checked for completeness and the instrument has passed both electrical and mechanical tests. The consigner or carrier shall be liable for the damage to instrument resulting from shipment. RIGOL would not be responsible for free maintenance/rework or replacement of the unit. 2. Inspect the instrument.
RIGOL Appearance and Dimensions 2 Figure 1 Front View Unit: mm Figure 2 Top View Unit: mm MSO1000Z/DS1000Z Quick Guide
RIGOL To Prepare for Operation To Adjust the Supporting Legs Adjust the supporting legs properly to use them as stands to tilt the oscilloscope upwards for stable placement of the instrument as well as easier operation and observation of the instrument. Figure 3 To Adjust the Supporting Legs To Connect to AC Power Supply This oscilloscope can accept 100-240 V, 45-440 Hz AC power supply.
RIGOL Power-on Inspection When the oscilloscope is energized, press the power key at the lower-left corner of the front panel to turn on the oscilloscope. During the start-up process, the oscilloscope performs a series of self-test items. After the self-test is finished, the welcome screen is displayed. The instrument is installed with the trial versions of the options before leaving factory and the remaining time is about 2000 minutes.
RIGOL Connect the Logic Probe: 1. Connect the single-wire terminal of the logic probe to the digital channel input terminal at the front panel of the oscilloscope in the correct direction. 2. Connect the other terminal of the logic probe to the signal under test. MSO1000Z series is provided with the RPL1116 logic probe (standard) which provides two connecting methods to connect to the signal under test to realize convenient and flexible detection.
RIGOL Function Inspection 1. Press Storage Æ Default to restore the oscilloscope to its default configuration. 2. Connect the earth alligator clip of the probe to the "Ground Terminal" as shown in the figure below. 3. Use the probe to connect the input terminal of CH1 and the "Compensation Signal Output Terminal" of the oscilloscope. Compensation Signal Output Terminal Ground Terminal Figure 7 To Use the Compensation Signal 4. Press AUTO. 5. Observe the waveform on the display.
RIGOL Probe Compensation When the probes are used for the first time, you should compensate the probes to match the input channels of the oscilloscope. Non-compensated or poorly compensated probes may cause measurement inaccuracy or error. The probe compensation procedures are as follows: 1. Perform steps 1, 2, 3 and 4 of "Function Inspection". 2. Check the displayed waveforms and compare them with the following figures.
RIGOL Front Panel Overview 1 3 2 11 12 13 4 14 5 15 6 16 17 7 8 18 9 19 10 20 Figure 10 Front Panel Overview Table 1 Front Panel Description No. Description 1 Measurement Menu Softkeys 2 LCD 3 Function Menu Softkeys 4 Multifunction Knob Common Operation Keys 5 6 CLEAR 7 AUTO 8 RUN/STOP 9 SINGLE 10 Help/Print No.
RIGOL Rear Panel Overview 1 4 5 2 6 3 7 8 Figure 11 Rear Panel Overview 1ˊ Handle Pull up the handle vertically for easy carrying of the instrument. When you do not need the handle, press it down. 2ˊ LAN Connect the instrument to the network via this interface for remote control. This oscilloscope conforms to the LXI CORE 2011 DEVICE class instrument standards and can quickly build test system with other instruments.
RIGOL z Pass/Fail: The instrument will output a negative pulse via this connector when a failed waveform is detected during the pass/fail test. The instrument continuously outputs a low level via this connector when no failed waveform is detected. 5ˊ Source Output The output terminals of the built-in dual-channel source of the oscilloscope. When Source1 or Source2 is enabled, the signal currently set can be output through the [Source1] or [Source2] connector at the rear panel.
RIGOL Front Panel Function Overview VERTICAL CH1, CH2, CH3, CH4: analog channel setting keys. The 4 channels are marked by different colors which are also used to mark both the corresponding waveforms on the screen and the channel input connectors. Press any key to open the corresponding channel menu and press again to turn off the channel.
RIGOL Logic Analyzer Press this key to open the logic analyzer control menu. You can turn on or off any channel or channel group, modify the display size of the digital channel, modify the logic threshold of the digital channel as well as group the 16 digital channels. You can also set a label for each digital channel. Note: ȸ This function is only applicable to MSO1000Z and DS1000Z Plus with the MSO upgrade option.
RIGOL HORIZONTAL HORIZONTAL POSITION: modify the horizontal position. The trigger point would move left or right relative to the center of the screen when you turn the knob. During the modification, waveforms of all the channels would move left or right and the horizontal position message (e.g. ) at the upper-right corner of the screen would change accordingly. Press down this knob to quickly reset the horizontal position (or the delayed sweep position).
RIGOL CLEAR Press this key to clear all the waveforms on the screen. If the oscilloscope is in "RUN" state, new waveforms will still be displayed. AUTO Press this key to enable the waveform auto setting function. The oscilloscope will automatically adjust the vertical scale, horizontal time base and trigger mode according to the input signal to realize optimum waveform display.
RIGOL Multifunction Knob Adjust waveform brightness: In non-menu-operation mode, turn this knob to adjust the brightness of waveform display. The adjustable range is from 0% to 100%. Turn clockwise to increase the brightness and counterclockwise to reduce. Press down this knob to reset the brightness to 60%. You can also press Display Æ Intensity and use the knob to adjust the waveform brightness. Multifunctional: In menu operation, the backlight of the knob goes on.
RIGOL Display: press this key to enter display setting menu to set the display type, persistence time, wave intensity, grid type and grid brightness. Utility: press this key to enter the system function setting menu to set the system-related functions or parameters, such as the I/O, sound and language. Besides, some advanced functions (such as the pass/fail test, waveform record, etc.) are also supported. Print Press this key to print the screen or save the screen to a USB storage device.
RIGOL User Interface MSO1000Z/DS1000Z provides 7.0 inch WVGA (800*480) TFT LCD. 1 2 12 3 4 13 14 5 6 15 7 16 8 17 18 9 19 10 20 11 21 22 Figure 12 User Interface 1. Auto Measurement Items Provide 16 horizontal (HORIZONTAL) and 17 vertical (VERTICAL) measurement parameters. Press the softkey at the left of the screen to activate the corresponding measurement item. Press MENU continuously to switch between the horizontal and vertical parameters. 2.
RIGOL 3. Status Available states include RUN, STOP, T’D (triggered), WAIT and AUTO. 4. Horizontal Time Base z Represent the time per grid on the horizontal axis on the screen. SCALE to modify this parameter. The range z Use HORIZONTAL available is from 5 ns to 50 s. 5. Sample Rate/Memory Depth z Display the current sample rate and memory depth of the oscilloscope. z The sample rate and memory depth will change in accordance with the horizontal time base. 6.
RIGOL z When using TRIGGER LEVEL to modify the trigger level, the trigger level value will change with the up and down of . Note: In slope trigger, runt trigger and windows trigger, there are two and ). trigger level labels ( 12. CH1 Vertical Scale z Display the voltage value per grid of CH1 waveform vertically. SCALE to modify this parameter. z Use VERTICAL z The following labels will be displayed according to the current channel setting: channel coupling (e.g. ) and bandwidth limit (e.g. ). 13.
RIGOL z z Source1 Waveform. When the impedance of signal source is set to 50 ƻ, will be displayed at the bottom of the Source1 Waveform. Only applicable to digital oscilloscopes with source channels. 20. Source2 Waveform z Display the type of waveform currently set for Source2. will be displayed at the bottom of the z When the modulation is enabled, Source2 Waveform. will be displayed at z When the impedance of signal source is set to 50 ƻ, the bottom of the Source2 Waveform.
RIGOL Parameter Setting Method This oscilloscope provides two methods to set the basic parameters. Method 1: is displayed on the menu, you can rotate When directly. to set the desired value Method 2: is displayed, press down and the numeric keyboard is displayed, as When shown below. Rotate the knob to select the desired value, and then press down the knob to input the value. After all the value input, rotate the knob to select the desired unit, press down the knob to finish your input.
RIGOL To Use the Built-in Help System The help system of this oscilloscope provides instructions for all the function keys (including menu keys) at the front panel. Press Help to open the help interface and press again to close the interface. The help interface mainly consists of two parts. The left is "Help Options" and the right is "Help Display Area".
RIGOL Troubleshooting The commonly encountered failures and their solutions are listed below. When you encounter those failures, please solve them following the corresponding steps. If the problem remains still, please contact RIGOL and provide your device information (acquisition method: Utility Æ System Æ System Info). 1. The screen is still dark (no display) after power on: (1) Check whether the power switch is really on. (2) Check whether the power is correctly connected.
RIGOL 6. The display of waveform is ladder-like: (1) The horizontal time base might be too low. Increase the horizontal time base to increase the horizontal resolution and improve the display. (2) If the display type is "Vectors", the lines between the sample points may cause ladder-like display. Press Display Æ Type to set the display type to "Dots" to solve the problem. 7.