X-Y Format 
This format is useful for studying phase relationships between two signals. 
Channel  1  in  the  horizontal  axis(X)  and  channel  2  in  the  vertical  axis(Y),  the 
oscilloscope uses a none-trigger acquisition mode, data is displayed as dots. 
Figure 2- 40 
X-Y display format 
NOTE:In Y-T format, all sample rates are available. But in X-Y format, 100 MSa/s 
is not available. In common, deceasing the sample rate can display the wavefrom 
better. 
The following modes or functions will not work in X-Y format. 
  LA Function (DS1000D series) 
  Automatic Measurements 
  Cursor Measurements 
  REF and MATH Operations 
  Delayed Scan Mode 
  Vector Display Mode 
  Horizontal    knob 