User`s guide
RIGOL 
© 2008 RIGOL Technologies, Inc.                                                                                                 
  User’s Guide for DS1000E, DS1000D Series 
5-3 
Accuracy 
Delta Time 
Measurement 
Accuracy   
(Full Bandwidth) 
Single-shot: ± (1 sample interval  + 50ppm ×  reading +  0.6 
ns)   
>16 averages: ± (1sample interval + 50ppm ×  reading + 0.4 
ns) 
Vertical 
A/D converter 
8-bit resolution, each channel samples simultaneously
[2]
Volts/div Range 
2mV/div~10V/div at input BNC 
Maximum Input 
Analog channel maximum input voltage 
CAT I    300Vrms, 1000Vpk;instantaneous voltage 1000Vpk 
CAT II 100Vrms,1000Vpk 
RP2200 10:1:CAT II 300Vrms 
RP3200 10:1:CAT II 300Vrms 
RP3300 10:1:CAT II 300Vrms 
Offset Range   
± 40V(200mV-10V),    ± 2V(2mV-100mV) 
Analog Bandwidth 
100MHz (DS1102D,DS1102E) 
50MHz (DS1052D, DS1052E) 
Single-shot 
Bandwidth 
80MHz (DS1102D, DS1102E) 
50MHz (DS1052D, DS1052E) 
Selectable Analog 
Bandwidth Limit 
(typical) 
20MHz     
Lower  Frequency 
Limit (AC –3dB) 
≤5Hz (at input BNC) 
Rise Time at BNC, 
typical 
<3.5ns, <7ns,   
On (100M, 50M) respectively   
DC Gain Accuracy 
2mV/div-5mV/div: 
± 4% (Sample or Average acquisition mode) 
10mV/div-10V/div: 
± 3% (Sample or Average acquisition mode) 
DC Measurement 
Accuracy,  Average 
Acquisition Mode 
Average of ≥16 Waveforms with vertical position at zero: 
± (DC Gain Accuracy× reading+0.1div+1mV)   
Average of ≥16 Waveforms with vertical position not at zero: 
±[DC  Gain  Accuracy× (reading+vertical  position)+(1%  of 
vertical position) + 0.2div] 
Add 2mV for settings from 2mV/div to 200 mV/div 