  User’s Guide for DS1000E, DS1000D Series 
2-99 
Manual Mode 
Figure 2- 121 Table 2- 79 The Manual Mode menu   
Menu 
Settings 
Comments 
Mode 
Manual 
Adjust  the  cursor  to  measure  X/Y 
parameters manually 
Type 
X 
Y 
Shown as vertical line to measure the 
horizontal parameters   
Shown  as  horizontal  line  to  measure 
the vertical parameters   
Source 
CH1 
CH2 
MATH/ 
FFT 
LA 
Select the measurement signal source. 
(LA is only for DS1000D series) 
In this mode, the oscilloscope measures the Y or X coordinate values of the cursors, 
and the increments between the two cursors. 
To do manual Cursor Measurements, please do as the following steps: 
1.  Select the Manual mode for cursor measurement by pressing soft button as: 
Cursor→Mode→Manual. 
2.  Select channel Source for measurements by pressing soft button as: 
Cursor→Source→CH1, CH2, MATH (FFT) or LA (DS1000D series) 
NOTE: While measuring the channel of MATH, the results are valued with “d” 
(division) as units. 
3.  Select the cursors type by pressing soft button as Cursor→Type→X or Y. 
4.  Move the cursors to adjust the increment between the cursors: (Details in the 
following Table) 