Chapter 4 To Set the Sample System RIGOL
DS1000Z User’s Guide
4-7
current sample rate.
When two channels are enabled, the memory depths available include Auto, 6kPoints,
60kPoints, 600kPoints, 6MPoints and 12MPoints (optional). In “Auto” mode, the
oscilloscope selects the memory depth automatically according to the current sample
rate.
When four channels are enabled, the memory depths available include Auto,
3kPoints, 30kPoints, 300kPoints, 3MPoints and 6MPoints (optional). In “Auto” mode,
the oscilloscope selects the memory depth automatically according to the current
sample rate.
Antialiasing
At slower sweep speed, the sample rate is reduced and a dedicated display algorithm
is used to minimize the possibility of aliasing.
Press Acquire Anti_Aliasing to enable or disable the antialiasing function. By
default, antialiasing is disabled. The displayed waveforms will be more susceptible to
aliasing when this function is disabled.