User`s guide
Chapter 1 Quick Start RIGOL
DS1000Z User’s Guide
1-13
4. Trigger Out/Pass/Fail
Trigger Out
The oscilloscope outputs a signal that can reflect the current capture rate of
the oscilloscope at each trigger via this connector. You can connect the
signal to a waveform display instrument to measure the frequency of the
signal. The measurement result is equal to the current capture rate.
Pass/Fail
In the pass/fail test, this connector outputs a high level when failed
waveforms are detected by the oscilloscope and outputs low level when
passed waveforms are detected by the oscilloscope.
5. Source Output
The oscilloscope provides two built-in source channel output terminals. When
the output of Source1 or Source2 is enabled, the [Source1] or [Source2]
connectors at the rear panel outputs the current signal.
6. Lock Hole
You can lock the instrument to a fixed location using the security lock (please
buy it yourself) via the lock hole.
7. Fuse
If a new fuse is required, please use the specified fuse (250V, T2A) and follow
the steps below.
a) Turn off the instrument and remove the power cord.
b) Insert a straight screwdriver into the slot at the power socket and prize out
the fuse seat gently.
c) Take out the fuse and replace it with specified fuse, and then install the fuse
seat to the original position.
8. AC Power Socket
AC power input terminal. The power requirements of this oscilloscope are
100-240 V, 45-440 Hz. Use the power cord provided with the accessories to
connect the instrument to AC power. Then, you can press the power key at the
front panel to start the instrument.