User`s guide

RIGOL Chapter 16 Troubleshooting
DS1000Z Users Guide
16-2
signal.
(4) Change the trigger holdoff setting.
5. No display after pressing RUN/STOP:
Check whether the MODE at the front panel trigger control area (TRIGGER) is
on “Normal” or “Single” and whether the trigger level exceeds the waveform
range. If yes, set the trigger level to the middle or set the MODE to “Auto”.
Note: Using AUTO could automatically finish the above setting.
6. The waveform displayed is ladder-like:
(1) The horizontal time base might be too low. Increase the horizontal time
base to increase the horizontal resolution and improve the display.
(2) If Display Type is “Vectors”, the lines between the sample points may
cause ladder-like display. Set Type toDotsto solve the problem.
7. Fail to connect PC through USB:
(1) Check whether the USB data cable is correctly connected to the oscilloscope
and PC.
(2) Check whether the USB data cable is in good condition and if needed,
restart the oscilloscope.
8. The USB storage device cannot be recognized:
(1) Check whether the USB storage device can work normally.
(2) Make sure that the USB storage device being used is flash type. This
oscilloscope does not support hardware type USB storage device.
(3) Make sure whether the capacity of the USB storage device is too large. It is
recommended that the capacity of the USB storage device being used with
this oscilloscope is no larger than 8 GBytes.
(4) Restart the instrument and then insert the USB storage device to check it.
(5) If the USB storage device still can not be used normally, please contact
RIGOL.