To Output Basic Waveform
The built-in signal source of DS1000Z can output various kinds of basic waveforms,
including sine, square, ramp, pulse, DC and noise.
To Output Sine Waveform
Press Src1Conf to enter the waveform setting interface. Press Wave to select
“Sine”, at this point, you can set the sine waveform parameters.
1. Output
Press Output to turn the signal output on or off. This function is the same with
that of Source Output. When “ON” is selected, the instrument outputs the
signal currently set from the [Source1] connector at the rear panel.
2. To set the frequency
Press Frequency to set the frequency of the current signal. For the setting
methods, refer to “
Parameter Setting Methods”.
The frequency range of different waveform differs.
Sine: 100 mHz to 25 MHz
Square: 100 mHz to 15 MHz
Ramp: 100 mHz to 100 kHz
Pulse: 100 mHz to 1 MHz
DC and Noise: do not have the frequency parameter
3. To set the amplitude
Press Amplitude to set the amplitude of the current signal. For the setting
methods, refer to “
Parameter Setting Methods”. When the impedance is set
to HighZ, the range available is from (-2.5 V+current amplitude/2) to (2.5
V-current amplitude/2) and when the impedance is set to 50 Ω, the range
available is (-1.25 V+current amplitude/2) to (1.25 V-current amplitude/2).
4. To set the DC offset voltage
Press Offset to set the DC offset of the current signal. For the setting methods,
refer to “
Parameter Setting Methods”. When the impedance is set to HighZ,