User`s guide

RIGOL Chapter 6 To Make Measurements
DS1000Z Users Guide
6-60
Auto Measurement
DS1000Z provides auto measurements of 32 waveform parameters and the statistics
and analysis of the measurement results. Whats more, you can also use the
frequency counter to realize more precise frequency measurement.
Quick Measurement after AUTO
When the oscilloscope is correctly connected and has detected input signal, press
Auto to enable waveform auto setting function and open the following function
menu:
Figure 6-2 Quick
Measurement Menu
after AUTO
Single-period: measure the “Period” and “Frequency “of
the current signal within a single period and display the
measurement results at the bottom of the screen.
Multi-period:
measure the “Period” and “Frequency “of
the current signal within multiple periods and display the
measurement results at the bottom of the screen.
Rise Time: measure the “Rise Time” of the current signal
and display the measurement result at the bottom of the
screen.
Fall Time: measure the “Fall Time” of the current signal
and display the measurement result at the bottom of the
screen.
Note: The AUTO function requires that the frequency of the signal under test
should be no lower than 50 Hz, the duty cycle be greater than 1% and the amplitude
be at least 20 mVpp. If the parameters of the signal under test exceed these limits,
after this softkey is pressed, the quick parameter measurement items might not be
displayed in the pop-up menu.