Chapter 6 To Make Measurements RIGOL
DS1000Z Users Guide
5-23
Setup Type:
Press SetupType to select the desired setup type.
Setup: set the time that the data stays stable and constant before the clock
edge appears. Press Setup to set the setup time and the range is from 8 ns to 1
s.
Hold: set the time that the data stays stable and constant after the clock edge
appears. Press Hold to set the hold time and the range is from 8 ns to 1 s.
SetupHold: set the time that the data stays stable and constant before and
after the clock edge appears. Press Setup and Hold to set the setup time and
hold time respectively and the range is from 8 ns to 1 s.
Trigger Mode:
Press Sweep to select the Trigger Modeunder this trigger type as auto, normal
or single. The corresponding status light of the current trigger mode turns on.
Trigger Setting:
Press Setting to set the trigger parameters (trigger holdoff and noise rejection)
under this trigger type.
Trigger Level:
Press DataSrc and use TRIGGER LEVEL to modify the trigger level of the data
source. Press ClkSrc and use TRIGGER LEVEL to modify the trigger level of the
clock source. For details, please refer to the description of
Trigger Levelon page
5-9.