RIGOL Service Guide DS1000E, DS1000D Series Digital Oscilloscopes DS1102E, DS1052E, DS1102D, DS1052D May 2009 RIGOL Technologies, Inc.
RIGOL Copyright © 2009 RIGOL Technologies, Inc. All Rights Reserved. Trademark Information RIGOL is registered trademark of RIGOL Technologies, Inc. Notices RIGOL products are protected by patent law in and outside of P.R.C.. RIGOL Technologies, Inc. reserves the right to modify or change parts of or all the specifications and pricing policies at company’s sole decision. Information in this publication replaces all previously corresponding material.
RIGOL Safety Notices Review the following safety precautions carefully before operating the instrument to avoid any personal injuries or damages to the instrument and any product connected to it. To prevent potential hazards, do use the instrument specified by this user’s guide only. The instrument should be serviced by qualified personnel only. To avoid instrument damage or personnel injury caused by misoperation, the instrument should be serviced by qualified personnel only. Use Proper Power Cord.
RIGOL Keep Well Ventilation. Inadequately ventilated will cause the temperature rises or damages to the device. Please keep well ventilation and inspect the intake and fan regularly. Do not Operate in Wet/Damp Conditions. In order to avoid short circuit to the interior of the device or electric shock, please do not operate in a humid environment. Do not Operate in an Explosive atmosphere. In order to avoid damages to the device or personal injury, please operate far away from an explosive atmosphere.
RIGOL Safety Terms and Symbols Terms in this Guide. These terms may appear in this manual: ! WARNING Warning statements indicate the conditions or practices that could result in injury or loss of life. ! CAUTION Caution statements indicate the conditions or practices that could result in damage to this product or other property. Terms on the Product. These terms may appear on the product: DANGER indicates an injury or hazard that may immediately happen.
RIGOL Document Overview Chapter 1 Specifications List the specifications and general specifications of DS1000E, DS1000D series. Chapter 2 Prepare for Use Introduce the preparatory work should be done before using the oscilloscope. Chapter 3 Performance Test Introduce how to execute the performance test to understand current performance status of the oscilloscope. Chapter 4 Calibration Introduce how to calibrate the oscilloscope.
RIGOL Contents Safety Notices ............................................................................................. II Safety Terms and Symbols ........................................................................... IV Document Overview ..................................................................................... V Contents .................................................................................................... VI Figures......................................................
RIGOL Warranty .................................................................................................. 8-2 Care and Cleaning ..................................................................................... 8-3 Contact Us ................................................................................................ 8-4 Index ...........................................................................................................
RIGOL Figures Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure VIII 2-1 Connect the power cord ................................................................... 2-3 2-2 Power-on inspection ........................................................................ 2-3 2-3 Connect probe to oscilloscope ...........
RIGOL Tables Table Table Table Table 2-1 3-1 3-2 7-1 The auto items .................................................................................. 2-9 Pass/Fail setting items........................................................................ 3-8 Testing specifications and the ranges ................................................ 3-11 Replaceable parts list .........................................................................
Chapter 1 Specifications RIGOL Chapter 1 Specifications All specifications apply to DS1000E, DS1000D series Oscilloscopes unless noted otherwise. To meet these specifications, two conditions must first be met: The instrument must have been operating continuously for thirty minutes within the specified operating temperature. Do perform the “Self Cal” operation, accessible through the Utility menu, if the operating temperature changes by more than 5°C.
Chapter 1 Specifications RIGOL Specifications Acquisition Sampling Modes Sampling Rate Averages Real-Time Equivalent DS1102X DS1052X 1GSa/s, 200MSa/s 25GSa/s 10GSa/s N time acquisitions, all channels simultaneously, N is selectable from 2, 4, 8, 16, 32, 64, 128 and 256.
Chapter 1 Specifications A/D converter Volts/div Range Maximum Input Offset Range Analog Bandwidth Single-shot Bandwidth Selectable Analog Bandwidth Limit (typical) Lower Frequency Limit (AC –3dB) Rise Time at BNC, typical Dynamic range DC Gain Accuracy DC Measurement Accuracy, Average Acquisition Mode Delta Volts Measurement Accuracy (Average Acquisition Mode) Trigger Trigger Sensitivity Trigger Level Range RIGOL 8-bit resolution, each channel samples simultaneously[2] 2mV/div~10V/div at input BNC An
Chapter 1 Specifications RIGOL Trigger Offset Normal mode: pre-trigger (storage depth/ 2*sampling rate), delayed trigger 1s Trigger Holdoff range Slow Scan mode: pre-trigger 6div, delayed trigger 6div 500ns~1.
Chapter 1 Specifications RIGOL General Specifications Display Display Type 5.7 in.
Chapter 2 Prepare for Use RIGOL Chapter 2 Prepare for Use This chapter contains the following topics: General Inspection Power-On Inspection Connect the Probe Probe Compensation Digital Probe (Only for DS1000D Series) Display a Waveform Automatically Service Guide for DS1000E, DS1000D Series 2-1
Chapter 2 Prepare for Use RIGOL General Inspection When you get a new DS1000E, DS1000D series oscilloscope, please inspect the instrument according to the following steps: 1. Inspect the shipping container for damage. Keep a damaged shipping container and cushioning material until the contents of the shipment have been checked for completeness and the instrument has been checked mechanically and electrically. 2. Inspect the instrument.
Chapter 2 Prepare for Use RIGOL Power-On Inspection Generally, normal operating voltage and frequency for DS1000E, DS1000D series digital oscilloscope are 100-240VRMS and 45-440Hz respectively. Please connect one terminal of the power cord to the socket in left side of the oscilloscope and the other to the AC power source. Figure 2-1 Connect the power cord Press the power button on top of the oscilloscope, some keys on the front panel will light for about 2 seconds until the normal display appears.
Chapter 2 Prepare for Use RIGOL Connect the Probe DS1000E series is a type of digital oscilloscope with dual channel inputs and one external trigger input. While, DS1000D series is a type with dual channel input, one external trigger input and sixteen channel digital inputs. Please take the following steps to connect the probe: 1. Attach BNC connector of the probe to the channel input or external trigger interfaces, insert it vertically until it latches into places.
Chapter 2 Prepare for Use RIGOL Probe Compensation You are suggested to compensate probe before first using so as to match its characteristics with oscilloscope. Measurement error may be generated without compensation or because of deviation compensation. To compensate the probe, follow the steps below. 1. Set both the attenuation switch of the probe and the probe scale in the menu to 10X. Probe Attenuation Quotient Figure 2-5 Set the probe attenuation quotient and switch 2. Connect the probe to CH1.
Chapter 2 Prepare for Use RIGOL correct in accordance with the shape of waveform. Over compensation Correct Compensation Under Compensation Figure 2-7 Waveform compensation 4. If necessary, use a non-metallic screwdriver to adjust the variable capacitor of the probe until a “Correct Compensation” wave displayed on the screen. 5. Continue in the same way to check CH2.
Chapter 2 Prepare for Use RIGOL Digital Probe (Only for DS1000D Series) DS1000D series provide sixteen digital channels. Take the steps below to connect instrument and test points by digital probes: 1. Switch off power supply of the device under test if necessary to avoid short circuit. Since no voltage is applied to the leads at this step, you may keep the oscilloscope on. 2. Connect one terminal of the flat cable FC1868 to the Logic Analyzer Input; connect the other end to Logic Head LH1116.
Chapter 2 Prepare for Use RIGOL Figure 2-10 Attach the clip to testing point 5. Connect the test clip to the ground.
Chapter 2 Prepare for Use RIGOL Display a Waveform Automatically DS1000E, DS1000D series digital oscilloscopes have an automatic feature to display the input signal best-fit. The oscilloscope enables to automatically set up VERTICAL, HORIZONTAL and TRIGGER parameters for the input signal under display. The input signal should be 50Hz or higher of frequency and the duty cycle of which should greater than 1%. 1. Operation steps Connect a signal to input channel.
Chapter 3 Performance Test RIGOL Chapter 3 Performance Test This chapter contains the following topics: Interfaces Test USB Host Interface Test USB Device Interface Test RS-232 Interface Test P/F Interface Test Logic Analyzer Interface Test (Only for DS1000D Series) Specifications Test Service Guide for DS1000E, DS1000D Series 3-1
Chapter 3 Performance Test RIGOL Interfaces Test USB Host Interface Test Purpose: Test if the USB Host interface works normally through U disc. Tools: A set of DS1000E, DS1000D series digital oscilloscope An U disk Steps: 1. Insert the U disk into the USB Host interface on the front panel of the oscilloscope. Figure 3-1 Connect the U disk 2.
Chapter 3 Performance Test RIGOL 3. If an update program about the oscilloscope available in the U disc is detected, a prompt will appeared to ask you whether to update or not; if select OK, the corresponding update process will be shown on the screen. 4. Press Storage→External to copy or delete the information in U disc. 5. Remove the U disc and a corresponding prompt “USB device removed” will appear.
Chapter 3 Performance Test RIGOL USB Device Interface Test Purpose: Test if the USB Device interface works normally through Ultrascope for DS1000E, DS1000D series. Tools: A set of DS1000E, DS1000D series digital oscilloscope A PC with USB interface A standard USB cable (Type AB) Ultrascope for DS1000E, DS1000D series Steps: 1. Install the Ultrascope for DS1000E, DS1000D series software on the PC. 2.
Chapter 3 Performance Test 4. RIGOL Click Tools → Connect to Oscilloscope; If successful, the corresponding indicator at the upper right corner of the software will be changed from red to blue; if failed, a prompting message will be appeared as below. Figure 3-5 Fail to connection Hint For the newest version of this software please go to www.rigolna.com download.
Chapter 3 Performance Test RIGOL RS-232 Interface Test Purpose: Test if RS-232 interface works normally through Ultrascope for DS1000E, DS1000D series. Tools: A set of DS1000E, DS1000D series digital oscilloscope A PC with RS-232 interface A standard RS-232 cable Ultrascope for DS1000E, DS1000D series Steps: 1. Install the Ultrascope for DS1000E, DS1000D series software on the PC. 2. Connect the oscilloscope with PC using an RS-232 cable. Figure 3-6 RS-232 interface connection 3.
Chapter 3 Performance Test 4. RIGOL Click Communication Setting and setup the communication interface under using and the baud rate. Figure 3-8 RS-232 interface setting 5. Press Utility→I/O Setting and set the baud rate as the same value in Ultrascope. 6. Click Tools → Connect to Oscilloscope; If successful, the corresponding reminder light at the upper right corner of the software will be changed from red to blue; otherwise, a prompting message will be appeared as below.
Chapter 3 Performance Test RIGOL P/F Interface Test Purpose: Test if P/F interface work normally through RS5101 module. Tools: A set of DS1000E, DS1000D series digital oscilloscope A BNC cable A RD5101 module Steps: 1. Power on the oscilloscope and RD5101 respectively. 2. Connect P/F interface on DS1000E, DS1000D with [Pass/Fail out] interface on RD5101 by BNC cable. Figure 3-10 P/F interface connection 3. Turn on the oscilloscope, press Storage → Storage → Factory factory settings.
Chapter 3 Performance Test RIGOL Logic Analyzer Interface Test (Only for DS1000D Series) Purpose: Test if the Logic Analyzer (hereinafter referred to as LA) interface works normally through the logic signal output module (DG-POD-A) of DG3000. Tools: A set of DS1000D series digital oscilloscope A set of DG3000 Function/ Arbitrary waveform Generator An active logic head Two logic cables 17 logic testing leads A DG-POD-A module Steps: 1.
Chapter 3 Performance Test RIGOL Figure 3-12 Setting interface 7. Press LA button on the oscilloscope, select D7-D0 and D15-D8 to open all the logic channels. 8. Select EDGE mode under the trigger menu and select any one source among Do to D15, if obtain the stable interface as follows, the LA module will be proved works normally.
Chapter 3 Performance Test RIGOL Specifications Test Purpose: Test if all the specifications are in the specified range. Requirement: Verify whether the oscilloscope is qualified through FLUKE calibration instrument and relative softwares.
Chapter 3 Performance Test RIGOL 1MΩ±2%, in parallel with 15pF± 3pF 1MΩ±2%, in parallel with 15pF± 3pF Voltage accuracy DC linearity DC offset Relative error between the average and the actual: CH1, CH2 2mV, 5mV scale >5mV scale DC voltage error array: CH1, CH2 200mV scale 5V scale Relative error of the voltage: CH1, CH2, 5V scale ± 4% ± 3% -2.60~2.60(mV) -2.60~2.60(mV) 0~6% Horizontal time Accuracy delta-t error -5.00e-9~5.
Chapter 4 Calibration RIGOL Chapter 4 Calibration The oscilloscope can achieve the optimum state fleetly by performing the calibration program and get accurate measurement. You can execute the calibration at any time. But when the operating temperature changes up to or more than 5℃, you must perform this programme. NOTE: The oscilloscope must have been working or warm-up at least 30-minutes before running self-calibration to get best accuracy. Steps: 1.
RIGOL Chapter 4 Calibration Generally speaking, the Self-Calibration will take about 5-6 minutes, if the calibration does not pass after this time, or the progress bar stop at one of the calibration item, there may be a trouble to the instrument. To resolve the trouble, restart the instrument and perform the Self-Calibration again, if the problem still remains, contact RIGOL for help.
Chapter 5 Disassembly and Assembly RIGOL Chapter 5 Disassembly and Assembly This chapter includes the following topics: Notices Structure Chart Disassemble and Assemble the Cover Disassemble and Assemble the Upper and Rear Cover Disassemble and Assemble the Power Board and the Fan Disassemble and Assemble the Panel Disassemble and Assemble LCD and the Keyboard Disassemble and Assemble the Mainboard Service Guide for DS1000E, DS1000D Series 5-1
Chapter 5 Disassembly and Assembly RIGOL Notices Notices: Don’t disassemble the product except the work needed. Disassemble only can be done by qualified person. Cut the power before disassembling. Take ESD glove under disassembling. Use proper tools and follow the disassembly sequence. Prevent metallic parts from transfiguration and avoid being scratched when disassembling.
Chapter 5 Disassembly and Assembly RIGOL Structure Chart Power Button Handle Multi-function Knob Function Run Control Buttons Vertical Control Horizontal Control Trigger Control USB Host Interface Digital Signal Input Analog Signal Input External Trigger Input Figure 5-1 Structure chart Service Guide for DS1000E, DS1000D Series 5-3
Chapter 5 Disassembly and Assembly RIGOL Disassemble and Assemble the Cover Power Button Handle ② ② ① ① Cover Foot Mat Figure 5-2 Disassemble and assemble the cover Parts Explanations: ① ② Bolts on the bottom of the cover (M3*8 Pan head Torx recess Drilling tapping screws): 2 Bolts at the groove of the handle (M3*8 Pan head Torx recess machine screws): 2 Disassemble steps: 1. 2. 3.
Chapter 5 Disassembly and Assembly RIGOL Disassemble and Assemble the Upper and Rear Covers ④ Upper Cover Rear Cover ④ ③ ③ Panel Tank Body Figure 5-3 Disassemble and assemble the upper and rear covers Parts Explanations: ③ ④ Bolts at the interface of RS-232 (DB9 interface bolts): 2 Bolts at the upper cover (M3*6 Pan head Torx recess composite machine screws): 2 Disassemble steps: 1. 2.
Chapter 5 Disassembly and Assembly RIGOL Disassemble and Assemble the Power Board and the Fan ⑤ ⑦ Power Board ⑥ ⑦ ⑤ ⑤ Mainboard Fan ⑧ Panel ⑧ ⑨ ⑧ ⑧ Figure 5-4 Disassemble and assemble the power board and the fan Parts Explanations: ⑤ ⑥ ⑦ ⑧ Small bolts for fixing the power board (M3*6 Pan head Torx recess composite machine screws): 3 Big bolts for fixing the power board (M4*8 Pan head Torx recess machine screws): 1 Bolts (M3*8 Pan head Torx recess Drilling tapping screws) and nuts (M3 hexagon nut
Chapter 5 Disassembly and Assembly 2. 3. 4.
Chapter 5 Disassembly and Assembly RIGOL Disassemble and Assemble the Panel ⑪ ⑪ ⑪ ⑪ ⑪ ⑨ ⑨ ⑨ ⑨ ⑩ ⑩ ⑨ Tank Body Button, Knob Panel Figure 5-5 Disassemble and assemble the panel Parts Explanations: ⑨ Small knobs: 4 ⑩ Big knobs: 2 ⑪ Bolts for fixing board on the tank body (M3*8 Pan head Torx recess Drilling tapping screws): 5 Disassemble steps: 1. 2.
Chapter 5 Disassembly and Assembly Disassemble Keyboard RIGOL and Assemble LCD and the ⑫ LCD ⑬ ⑬ ⑬ ⑬ Rubber Button Keyboard Tank Body Figure 5-6 Disassemble and assemble LCD and the keyboard Parts Explanations: ⑫ Bolts for the screen ground wire (M2*4 Pan head Torx recess machine screws): 1 ⑬ Bolts fixed on the keypad board (M3*6 Flush head Torx recess machine screws): 4 Disassemble steps: 1. 2. 3. 4. 5. 6.
Chapter 5 Disassembly and Assembly RIGOL Disassemble and Assemble the Mainboard ⑯ ⑯ ⑮ ⑮ ⑭ LA Board ⑭ Cooling Plate ⑮ ⑰ Mainboard ⑳ ⑳ ⑱ ⑱ ⑱ BNC Nut ⑲ ⑲ ⑲ Tank Body Figure 5-7 Disassemble and assemble the mainboard Parts Explanations: ⑭ Bolts on the LA board ((M3*6 Pan head Torx recess composite machine screws): 2 ⑮ Bolts on the cooling plate (M3*6 Pan head Torx recess composite machine screws): 3 ⑯ Bolts for fixing the mainboard (M3*6 Pan head Torx recess composite machine screws): 2 ⑰ Stud on the
Chapter 5 Disassembly and Assembly RIGOL ⑳ Bolts at LA terminal (0.1*0.3inch Hexagon Bolt in British system): 2 Disassemble steps: 1. 2. 3. 4. 5. 6. 7.
Chapter 6 Troubleshooting RIGOL Chapter 6 Troubleshooting 1. After the oscilloscopes is powered on, the screen remains dark (no display): (1) (2) (3) (4) 2. After the signal acquisition the waveform does not appear: (1) (2) (3) (4) (5) 3. Check the power cable connection. Ensure the power switch has been turned on. After above inspection, restart the oscilloscope. If the problem still remains, please ask RIGOL for help. Check if the probe is connecting with the signals.
RIGOL Chapter 6 Troubleshooting rotate the horizontal knob to increase horizontal resolution to improve the display. (2) Maybe the display Type is set to Vectors set it to Dots mode to improve the display.
Chapter 7 Replaceable Parts RIGOL Chapter 7 Replaceable Parts RIGOL provides some replaceable parts in order to maintain or update for users. Please see as the following figure and table. Note that the numbers in the figure and the table are associated with each other. Figure 7-1 DS1000E, DS1000D Exploded View Table 7-1 No.
Chapter 7 Replaceable Parts RIGOL 7-2 4 5 6 7 8 9 10 11 LCD Metallic Tank Body Rubber Button BNC Nut BNC Lock Spacer Encoder Keyboard Bolt 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 Fan Large Terminal Small Terminal BNC Cooling Plate LA Board LA Interface RS-232 Interface Power Jacket USB Host Interface Mainboard USB Device Interface Power Switch Cap Upper Metallic Cover Power Interface Power Supply Power Board Rear Metallic Cover Female ScrewLock Front Trestle Rear Trestle H
Chapter 8 Service & Support RIGOL Chapter 8 Service & Support This chapter contains the following topics: Warranty Care and Cleaning Contact Us Service Guide for DS1000E, DS1000D Series 8-1
RIGOL Chapter 8 Service & Support Warranty RIGOL warrants that its products mainframe and accessories will be free from defects in materials and workmanship within the warranty period. If a product proves defective within the respective period, RIGOL guarantees the free replacement or repair of products which are approved defective. To get repair service or obtain a copy of the whole warranty statement, please contact with your nearest RIGOL sales and service office.
Chapter 8 Service & Support RIGOL Care and Cleaning General Maintenance Do not store or leave the instrument in where the instrument will be exposed to direct sunlight for long periods of time. Caution To avoid damages to the instrument or probes, do not expose them to liquids which have causticity. Cleaning Clean the instrument and probes often based on its operating conditions require. To clean the exterior surface, perform the following steps: 1. Disconnect the instrument from all power sources. 2.
Chapter 8 Service & Support RIGOL Contact Us If you have any problem or requirement during using our products, please contact RIGOL Technologies, Inc. or the local distributors. Domestic: Please call Tel: (86-10) 8070 6688 Fax: (86-10) 8070 5070 Service & Support Hotline: 800 810 0002 9:00 am –5: 00 pm from Monday to Friday Or by e-mail: service@rigol.com Or mail to: RIGOL Technologies, Inc.
Index RIGOL Index Acquisition................................... 1-2 Big Knob ..................................... 7-1 BNC ............................................ 7-2 BNC Lock Spacer.......................... 7-2 BNC Nut ...................................... 7-2 Bolt ............................................. 7-2 Calibration ................................... 4-1 Calibration Interval....................... 1-5 Care and Cleaning ........................ 8-3 Cleaning .............................