Specifications
3
A/D Converter
8-bit resolution, all channels sample simultaneously
[2]
Volts/div Range
2mV/div~10V/div (at the input terminal connecting to BNC)
Maximum Input
Maximum input voltage on analog channel
CAT I 300Vrms, 1000Vpk; instantaneous overvoltage 1000Vpk
CAT II 100Vrms, 1000Vpk
RP2200 10:1:CAT II 300Vrms
RP3200 10:1:CAT II 300Vrms
RP3300 10:1:CAT II 300Vrms
Offset Range
± 40V (250mV/div~10V/div)
± 2V (2mV/div~245mV/div)
Analog Bandwidth
100MHz (DS1102D,DS1102E)
50MHz (DS1052D, DS1052E)
Single-shot Bandwidth
100MHz (DS1102D, DS1102E)
50MHz (DS1052D, DS1052E)
Selectable Analog
Bandwidth Limit (typical)
20MHz
Lower Frequency
Response (AC –3dB)
≤5Hz (at input BNC)
Rise Time at BNC (typical)
<3.5ns, <7ns, respectively at 100MHz, 50MHz
DC Gain Accuracy
2mV/div-5mV/div:
± 4% (In Normal or Average acquisition mode)
10mV/div-10V/div:
± 3% (In Normal or Average acquisition mode)
DC Measurement
Accuracy, Average
Acquisition Mode
When vertical displacement is zero, and N ≥16:
± (DC Gain Accuracy× reading+0.1div+1mV)
When vertical displacement is not zero, and N ≥16:
± [DC Gain Accuracy× (reading+vertical displacement)+(1% of
vertical displacement) + 0.2div]
When vertical scale is between 2mV/div and 245mV/div, add 2mV
more for setting value.
When vertical scale is between 250mV/div and 10V/div, add 50mV
more for setting value.
Delta Volts
Measurement Accuracy
(Average Acquisition Mode)
Under same setting and condition, the voltage difference (△ V)
between any two points in the waves coming from the average of
more than 16 waves have been acquired: ± (DC Gain
Accuracy× reading + 0.05 div)
Trigger
Trigger Sensitivity
0.1div~1.0div (adjustable)
Trigger Level Range
Internal
± 6 divisions from center of screen
EXT
± 1.2V
Trigger Level Accuracy (typical)
applicable for the signal of
rising and falling time ≥20ns
Internal
± (0.3div × V/div)(± 4 divisions from center of screen)
EXT
± (6% of setting + 200 mV)
Trigger Offset
In Normal mode: pre-trigger (memory depth/ 2*Sample rate),
delayed trigger 1s
In Slow Scan mode: pre-trigger 6div, delayed trigger 6div
Trigger Holdoff Range
500ns~1.5s
Set Level to 50% (typical)
When input signal frequency ≥50Hz
Edge Trigger
Edge trigger slope
Rising, Falling, Rising + Falling
Pulse WidthTrigger
Trigger Condition
(>, <, =) Positive pulse width, (>, <, =) Negative pulse width
Pulse Width Range
20ns ~10s