Specifications
RIGOL Appendix Test Record Form
DS1000E/D Performance Verification Guide
2
DC Gain Accuracy Test:
Channel
Vertical
Scale
Test Result
Limit Pass/Fail
Vavg1 Vavg2
Calculation
Result
[1]
CH1
2 mV/div
≤4%
5 mV/div
10 mV/div
≤3%
20 mV/div
50 mV/div
100 mV/div
200 mV/div
500 mV/div
1 V/div
2 V/div
5 V/div
10 V/div
CH2
2 mV/div
≤4%
5 mV/div
10 mV/div
≤3%
20 mV/div
50 mV/div
100 mV/div
200 mV/div
500 mV/div
1 V/div
2 V/div
5 V/div
10 V/div
Note
[1]
: the calculation formula is |(Vavg1–Vavg2)–(V
out1
-V
out2
)|/Full Scale×100%; wherein, V
out1
and V
out2
are 3 and
-3 times of the current vertical scale respectively.