Specifications
RIGOL Chapter 2 Performance Calibration Test
DS1000E/D Performance Verification Guide
2-8
23) Output a sine waveform with 1 MHz frequency and 3 Vpp amplitude from Fluke 9500B.
24) Repeat step 6).
25) Output a sine waveform with 20 MHz frequency and 3 Vpp amplitude from Fluke 9500B.
26) Repeat step 8), 9) and 10).
27) Output a sine waveform with 50 MHz frequency and 3 Vpp amplitude from Fluke 9500B.
28) Repeat step 12), 13) and 14).
29) Turn CH1 off. Test CH2 according to the above test steps and record the test results.