Specifications
Chapter 2 Performance Calibration Test RIGOL
DS1000E/D Performance Verification Guide
2-1
Chapter 2 Performance Calibration Test
This chapter introduces the performance verification test method of DS1000E/DS1000D series digital
oscilloscope by taking DS1102E as an example. You can perform the following test in any order. In
this manual, the test device used is Fluke 9500B. You can also use other devices that fulfill the
specification requirements for the test.
Recommended Devices List:
Device Specification Recommended Model
Oscilloscope
Calibrator
Output range of DC voltage:
1 MΩ: 1 mV to 200 V
50 MΩ:1 mV to 200 V
The rise time of fast edge signal: ≤ 150 ps
Fluke 9500B
Digital Multimeter
The resistance measurement accuracy is
higher than ±0.1% of reading
RIGOL DM3058/3068
Test Cable BNC (male) to Dual-banana Plug Cable --
Signal Generator Frequency Accuracy: ±1 ppm
RIGOL DG4162
Test Cable BNC (m)-BNC (m) cable --
Note:
1) Make sure that the oscilloscope passes the self-calibration before performing the performance
calibration test.
2) Make sure that the oscilloscope has been warmed up for 30 minutes before executing any of the
following tests.
3) Please reset the instrument to the factory setting before or after executing any of the following
tests.