User manual

RIGOL Chapter 2 Command System
2-42 DP700 Programming Guide
:TIMEr:ENDState
Syntax
:TIMEr:ENDState <state>
:TIMEr:ENDState?
Description
Sets the end state of the timer.
Queries the end state selected for the timer.
Parameter
Name Type Range Default
<state> Discrete {OFF|LAST} OFF
Remarks
"End State" refers to the state of the instrument after it has completed outputting groups of voltage
and current values for a specified number of times (the total number of output groups) when "cycles"
is N[,<value>] (a finite value). When "cycles" is set to I (Infinite), "End State" is invalid.
<state>:
OFF: indicates that the output will be turned off automatically after the output is completed.
LAST: keeps the output state (voltage/current) of the last group after the output is completed.
Total Number of Output Groups for the Timer = Outp Groups x Cycles. Wherein, for "Outp Groups",
refer to the
:TIMEr:GROUPs command; for "Cycles", refer to the :TIMEr:CYCLEs command.
Example
:TIMEr:ENDState LAST /*Sets the end state of the timer to "LAST"*/
:TIMEr:ENDState? /*Queries the end state selected for the timer*/
Return Format
OFF or LAST
Related Commands
:TIMEr:CYCLEs
:TIMEr:GROUPs
:TIMEr:GROUPs
Syntax
:TIMEr:GROUPs <value>
:TIMEr:GROUPs?
Description
Sets the number of output groups.
Queries the number of output groups selected for the timer.
Parameter
Name Type Range Default
<value> Integer 1 to 2048 1
Remarks
"Outp Groups" indicates the number of groups of the preset voltage and current values that the
instrument outputs in each cycle.
Total Number of Output Groups for the Timer = Outp Groups x Cycles. Wherein, for "Outp Groups",
refer to
:TIMEr:GROUPs.
After the instrument has completed outputting operation for a specified number of times (total number