User manual

RIGOL Chapter 2 Command System
2-16 DP700 Programming Guide
:OUTPut:OCP:CLEAR
Syntax
:OUTPut:OCP:CLEAR [<channel>]
Description
Clears the internal OCP flag and the OCP prompt message.
Parameter
Name Type Range Default
<channel>
Discrete
DP711: {CH1|P30V}
DP712: {CH1|P50V}
CH1
Remarks
Before executing the command, ensure that you have located and resolved the problem for causing
the overcurrent protection. To resolve the problem, reduce the output current to below the overcurrent
protection value or increase the overcurrent protection value to above the output current.
You can send the [:SOURce[<n>]]:CURRent:PROTection:CLEar
command to clear the internal OCP
flag and the OCP prompt message.
You can send the :OUTPut:OCP:ALAR?, :OUTPut:OCP:QUES?, or
[:SOURce[<n>]]:CURRent:PROTection:TRIPped? command to query whether the overcurrent
protection has currently occurred.
Example
:OUTPut:OCP:CLEAR /*Clears the internal OCP flag and the OCP prompt message*/
Related Commands
[:SOURce[<n>]]:CURRent:PROTection:CLEar
:OUTPut:OCP:ALAR?
:OUTPut:OCP:QUES?
[:SOURce[<n>]]:CURRent:PROTection:TRIPped?
:OUTPut:OCP[:STATe]
Syntax
:OUTPut:OCP[:STATe] [<channel>,]<state>
:OUTPut:OCP[:STATe]? [<channel>]
Description
Enable or disable the overcurrent protection (OCP) function.
Queries the on/off state of the overcurrent protection (OCP) function.
Parameter
Name Type Range Default
<channel> Discrete
DP711: {CH1|P30V}
DP712: {CH1|P50V}
CH1
<state> Bool {ON|OFF} OFF
Remarks
Overcurrent protection indicates that when the channel output current equals to the overcurrent
protection value, the internal OCP flag is set to 1, and a message is displayed on the screen, prompting
you that overcurrent protection occurs and the channel output is automatically disabled. At this time,
you can send the
:OUTPut:OCP:CLEAR or [:SOURce[<n>]]:CURRent:PROTection:CLEar command to
clear the internal OCP flag and the OCP prompt message.
You can also send the [:SOURce[<n>]]:CURRent:PROTection:STATe command to enable or disable the