User manual

RIGOL Chapter 2 Front Panel Operations
2-6 DP700 User’s Guide
displayed at the bottom of the corresponding interface or refer to the
descriptions in "Parameter Setting Method".
Set the overvoltage protection
(1) In the main interface, press the Up/Down arrow key to switch the
parameter focus to "OVP Setting".
(2) Press OK to enable/disable the overvoltage protection function.
When "OVP" is indicated in yellow, it indicates that the overvoltage
protection function is enabled; when "OVP" is indicated in white, it indicates
that the overvoltage protection function is disabled.
(3) Use the num key or the knob to set a desired value for the overvoltage
protection. For setting methods, refer to the descriptions of the help
information displayed at the bottom of the corresponding interface or refer
to the descriptions in "Parameter Setting Method".
Set the overcurrent protection
(1) In the main interface, press the Up/Down arrow key to switch the
parameter focus to "OCP Setting".
(2) Press OK to enable/disable the overcurrent protection function.
When "OCP" is indicated in yellow, it indicates that the overcurrent
protection function is enabled; when "OCP" is indicated in white, it indicates
that the overcurrent protection function is disabled.
(3) Use the num key or the knob to set a desired value for the overcurrent
protection. For setting methods, refer to the descriptions of the help
information displayed at the bottom of the corresponding interface or refer
to the descriptions in "Parameter Setting Method".
4. Turn on the channel output
Press On/Off to turn on the channel output. Then, the actual output voltage
(V), the actual output current (A), and the actual output power (W) are
highlighted; the actual output mode (CC) is displayed.
If the CV mode is displayed, please increase the voltage setting value
appropriately.
WARNING
To avoid electric shock, ensure that the load is connected to the channel
output terminals properly before you turn on the channel output.
CAUTION
When the fan works abnormally, the channel output is disabled. Then, a
message is displayed, "Fan abnormality, output disabled."