User manual

RIGOL Chapter 1 Quick Start
1-16 DP700 User’s Guide
(1) Connect a short across the positive terminal (+) and negative terminal (-) of
the channel output with an insulated test lead (18 AWG), and then press
On/Off to turn on the channel output. At this time, the actual output status
(actual output voltage (V), actual output current (A), and actual output
power (W)) on the left section of the interface is highlighted, and the actual
output mode is displayed. (Note that the actual output mode is determined
by the resistance of the test lead.)
(2) Set the output voltage to 2 V (For setting methods, refer to the descriptions
of the help information displayed at the bottom of the interface or refer to
the descriptions in "Parameter Setting Method") and ensure the output
mode to be CC.
(3) Set the output current to 0 A (For setting
methods, refer to the descriptions of the
help information at the bottom of the
interface or descriptions in "Parameter
Setting Method".), and check whether
the actual output current displayed on the
left section of the interface is approximately
0 A, the actual output voltage
approximately 0 V. Set the output current
to 5 A (max. rating value), and check
whether the actual output current is
approximately 5 A (the actual output
voltage indicates the voltage drop arising
from the test lead.)
(4) Press On/Off to turn off the channel output.
Voltage Setting
Value
Current Setting
Value