User manual
Chapter 1 Quick Start RIGOL
DP700 User’s Guide 1-15
Tip
After powering off the instrument, please wait for at least 2 s to power it on
again.
6. Perform the output check
Perform the output check to ensure that the power supply can output the rated
values and properly respond to the operations on the front panel.
Note: Before checking the current output, you should complete the check for
the voltage output first.
Check the voltage output: verify the basic voltage functions without a load.
(1) Press
to turn off the instrument.
(2) Ensure that no load is connected to the channel output terminals on the
front panel. Then, press to power on the instrument, and check
whether the voltage between the positive terminal (+) and negative
terminal (-) of the channel output is 0 V.
(3) Press On/Off to turn on the channel
output. At this time, the actual output
status (actual output voltage (V), actual
output current (A), and actual output
power (W)) on the left section of the
interface is highlighted, and the actual
output mode (CV) is displayed.
(4) Set the output voltage to 0 V (For setting
methods, refer to the descriptions of the
help information at the bottom of the
interface or descriptions in "Parameter
Setting Method".), and check whether
the actual output voltage displayed on the
left section of the interface is approximately
0 V, the actual output current
approximately 0 A. Set the output voltage
to 30 V (max. rating value), and check
whether the actual output voltage is
approximately 30 V, the actual output
current approximately 0 A.
(5) Press On/Off to turn off the channel output.
Check the current output: verify the basic current functions with a short
across the power supply’s output.
Voltage
Setting Value