10 DP700 Quick Guide
English
output voltage is approximately 30 V, the actual output current
approximately 0 A.
(5) Press On/Off to turn off the channel output.
Check the current output: verify the basic current functions with a
short across the power supply’s output.
(1) Connect a short across the positive terminal (+) and negative
terminal (-) of the channel output with an insulated test lead (18
AWG), and then press On/Off to turn on the channel output. At this
time, the actual output status (actual output voltage (V), actual
output current (A), and actual output power (W)) on the left section
of the interface is highlighted, and the actual output mode is
displayed. (Note that the actual output mode is determined by the
resistance of the test lead.)
(2) Set the output voltage to 2 V (For setting methods, refer to the
descriptions of the help information displayed below the interface or
refer to the descriptions on Page 11.) and ensure the output mode to
be CC.
(3) Set the output current to 0 A (For setting methods, refer to the
descriptions of the help information below the interface or
descriptions on Page 11.), and check whether the actual output
current displayed on the left section of the interface is approximately
0 A, the actual output voltage approximately 0 V. Set the output
current to 5 A (max. rating value), and check whether the actual
output current is approximately 5 A (the actual output voltage
indicates the voltage drop arising from the test lead).
(4) Press On/Off to turn off the channel output.