User manual
Chapter 2 Command System RIGOL
DP700 Programming Guide 2-25
[:SOURce[<n>]]:CURRent:PROTection:TRIPped?
Syntax
[:SOURce[<n>]]:CURRent:PROTection:TRIPped?
Description
Queries whether the overcurrent protection (OCP) has occurred.
Parameter
Name Type Range Default
<n> Discrete {1} 1
Remarks
Overcurrent protection indicates that when the channel output current equals to the overcurrent
protection value, the internal OCP flag is set to 1, and a message is displayed on the screen, prompting
you that overcurrent protection occurs and the channel output is automatically disabled. At this time,
you can send the
:OUTPut:OCP:CLEAR or [:SOURce[<n>]]:CURRent:PROTection:CLEar command to
clear the internal OCP flag and the OCP prompt message.
You can also send the
:OUTPut:OCP:ALAR? or :OUTPut:OCP:QUES? command to query whether the
overcurrent protection has currently occurred.
Example
:CURRent:PROTection:TRIPped? /*Queries whether the overcurrent protection has currently occurred*/
Return Format
YES or NO
Related Commands
:OUTPut:OCP:CLEAR
[:SOURce[<n>]]:CURRent:PROTection:CLEar
:OUTPut:OCP:ALAR?
:OUTPut:OCP:QUES?
[:SOURce[<n>]]:VOLTage[:LEVel][:IMMediate][:AMPLitude]
Syntax
[:SOURce[<n>]]:VOLTage[:LEVel][:IMMediate][:AMPLitude] {<voltage>|MINimum|MAXimum}
[:SOURce[<n>]]:VOLTage[:LEVel][:IMMediate][:AMPLitude]? [MINimum|MAXimum]
Description
Sets the channel output voltage.
Queries the channel voltage setting value.
Parameter
Name Type Range Default
<n> Discrete {1} 1
<current> Real 0 V to 30 V 0 V
Remarks
For the setting command, MINimum indicates setting the channel output voltage to a minimum value;
MAXimum indicates setting the channel output voltage to a maximum value.
For the query command, MINimum indicates querying the minimum settable value of the channel
voltage; MAXimum indicates querying the maximum settable value of the channel voltage.
You can also send the
:APPLy command to set the channel output voltage and/or current.