Data Sheet

RIGOL DP1116A Power Supply Specifications
Specifications for DP1116A
Remote Sensing (Sense)
Each Sense lead can compensate up to 1V voltage drop for each load lead.
Command Processing Time
[2]
< 50 ms
Temperature Coefficient,± (output percentage + offset)
(The maximum varieties of Output/Read Back for each 1 change of temperature
after 30 minutes warm-up.)
Voltage
0.01% + 3 mV
Current
0.02% + 3 mA
Stability
[3]
, ± (output percentage + offset)
Voltage
0.02% + 1 mV
Current
0.1% + 1 mA
Voltage Programming Speed (1% of total variation range)
Rising
Full Load
50 ms
No Load
20 ms
Falling
Full Load
45 ms
No Load
400 ms
OVP/OCP
Accuracy ± (output
percentage + offset)
0.5% + 0.5 V/0.5% + 0.5 A
Activation Time
1.5 ms (OVP ≥ 3 V); < 10 ms (OVP < 3 V)
< 10 ms (OCP)
Mechanical
Dimension
235 mm (W) x 155 mm (H) x 384 mm (D)
Weight
11 kg
Power Supply
AC input
(50 Hz 60 Hz)
100 Vac ± 10%, 115 Vac ± 10%,
220 Vac ± 10%, 230 Vac ± 10% (up to 250 Vac)
Environment
Working Temperature
Full rating output: 0℃~40
At high temperature: the output current drops to 50%
at 55of maximum.
Cooling
Fan cooling
Product Regulation
CE, cTUVus
Remarks:
[1] Specifications are for one hour warm-up and at 25.
[2] The maximum time required for regulating the corresponding output when received APPLy