Specifications
Chapter 2 Performance Test RIGOL
Service Guide for DM3068
2-9
Resistance
[2]
200.000Ω
short
±6mΩ
±8mΩ
±8mΩ
2.00000kΩ
short
±10mΩ
±20mΩ
±20mΩ
20.0000kΩ
short
Q
±100mΩ
±200mΩ
±200mΩ
200.000kΩ
short
±1Ω
±2Ω
±2Ω
1.00000MΩ
short
Q
±10Ω
±10Ω
±10Ω
10.0000MΩ
short
±100Ω
±100Ω
±100Ω
100.000MΩ
short
±10kΩ
±10kΩ
±10kΩ
Remarks:
[1] The integration time is set to 100PLC.
[2] Specifications are for 4-wire or 2-wire resistance measurement using “REL” operation. If
2-wire resistance measurement is used without REL operation, add 0.2 Ω additional errors.
[3] Q denotes the optional quick test point.