Datasheet
Specications
DC Characteristics
Function
DC Voltage
DC Current
Resistance
[6]
Diode Test
Continuity Test
Range
[2]
200.0000mV
2.000000V
20.00000V
200.0000V
1000.000V
[4]
200.0000uA
2.000000mA
20.00000mA
200.0000mA
2.000000A
10.00000A
[5]
200.0000Ω
2.000000kΩ
20.00000kΩ
200.0000kΩ
1.000000MΩ
10.00000MΩ
100.0000MΩ
2.0000V
[7]
2000.0Ω
Test Current or
Burden Voltage
<0.03V
<0.25V
<0.07V
<0.7V
<0.12V
<0.6V
1mA
1mA
100uA
10uA
2uA
200nA
200nA || 10MΩ
1mA
1mA
24 Hour
[3]
T
CAL
℃
±1
℃
0.0020+ 0.0020
0.0015 + 0.0005
0.0020 + 0.0004
0.0020 + 0.0006
0.0020 + 0.0006
0.010 + 0.012
0.007 + 0.003
0.007 + 0.012
0.010 + 0.002
0.050 + 0.020
0.100 + 0.010
0.0030 + 0.0030
0.0020 + 0.0005
0.0020 + 0.0005
0.0020 + 0.0005
0.002 + 0.001
0.015 + 0.001
0.300 + 0.010
0.002 + 0.010
0.002 + 0.010
90 Day
T
CAL
℃
± 5
℃
0.0030 + 0.0025
0.0020 + 0.0006
0.0030 + 0.0005
0.0040 + 0.0006
0.0040 + 0.0010
0.040 + 0.015
0.030 + 0.003
0.030 + 0.015
0.030 + 0.003
0.080 + 0.020
0.120 + 0.010
0.008 + 0.004
0.008 + 0.001
0.008 + 0.001
0.008 + 0.001
0.010 + 0.001
0.030 + 0.001
0.800 + 0.010
0.008 + 0.020
0.008 + 0.020
1 Year
T
CAL
℃
± 5
℃
0.0040 + 0.0025
0.0035 + 0.0006
0.0040 + 0.0005
0.0050 + 0.0006
0.0055 + 0.0010
0.050 + 0.015
0.050 + 0.003
0.050 + 0.015
0.050 + 0.003
0.100 + 0.020
0.150 + 0.010
0.010 + 0.004
0.010 + 0.001
0.010 + 0.001
0.010 + 0.001
0.012 + 0.001
0.040 + 0.001
0.800 + 0.010
0.010 + 0.020
0.010 + 0.020
Temperature Coefcient
0
℃
to (T
CAL
℃
- 5
℃
)
(T
CAL
℃
+ 5
℃
) to 50
℃
0.0005 + 0.0005
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0001
0.0020 + 0.0030
0.0020 + 0.0005
0.0020 + 0.0020
0.0020 + 0.0005
0.0050 + 0.0010
0.0050 + 0.0020
0.0006 + 0.0005
0.0006 + 0.0001
0.0006 + 0.0001
0.0006 + 0.0001
0.0010 + 0.0002
0.0030 + 0.0004
0.1500 + 0.0002
0.0010 + 0.0020
0.0010 + 0.0020
Accuracy Specications: ± ( % of reading + % of range)
[1]
[1] Specications are for 90-minute warm-up and 100NPLC integration time.
For integration time <100NPLC, add the appropriate “RMS Noise Adder”
listed in the following table.
[2] 10% overrange on all ranges except DCV 1000V and DCI 10A range.
[3] Relative to calibration standards.
[4] For each additional volt over ± 500 V, add 0.03mV error.
[5] For continuous current > 7A DC or 7A AC RMS, 30 seconds ON and 30
seconds OFF.
[6] Specications are for 4–wire resistance measurement or 2–wire
resistance measurement using REL operation. Without REL operation,
add 0.2 Ω additional error in 2-wire resistance measurement.
[7] Accuracy specications for the voltage measured at the input terminal
only. 1 mA test current is typical. Variation in the current source will create
some variation in the voltage drop across a diode junction.
Performance Versus Integration Time – 50 Hz (60 Hz) Power-line Frequency
Integration Time
Number of Power line
Cycles
(NPLC)
0.006
0.02
0.06
0.2
1
2
10
100
Resolution
[1]
(ppm Range)
2.7
1.6
1
0.5
0.22
0.17
0.08
0.035
50Hz
10000
2500
833
250
50
25
5
0.5
60Hz
10000
3000
1000
300
60
30
6
0.6
DCV 20V
0.0006
0.0004
0.0003
0.0001
0
0
0
0
DCV 2V 200V
Resistance 2 kΩ
20 kΩ
0.0007
0.0004
0.0003
0.0002
0.0001
0
0
0
DCV 1000 V
DCI 2 mA
200 mA
0.0015
0.0008
0.0006
0.0003
0.0002
0.0001
0
0
DCV 200 mV
Resistance 200 Ω
DCI 10 A
0.0040
0.0025
0.0025
0.0015
0.0004
0.0003
0.0002
0
NMRR
[2]
(dB)
0
0
0
0
60
60
60
60
Readings/s
[3]
RMS Noise Adder
[4]
(% of Range)
[1] Typical value. Resolution is dened as the typical 20V range RMS noise (using auto zero “Once”).
[2] Normal mode rejection ratio for power-line frequency ± 0.1%. For power-line frequency ± 1%, subtract 20 dB. For ± 3%, subtract 30dB.
[3] Maximum rate for DCV, DCI, 2-wire resistance and 4-wire resistance functions.
[4] The basic DC accuracy specications include RMS noise at 100 NPLC. For <100 NPLC, add “RMS Noise Adder” to the basic DC accuracy specications.