Specifications
RIGOL
Service Guide for DM3000 Series
3-16
Capacitance Test
In capacitance test, input the standard signal from testing equipment into the
equipment to be tested; and compare the test results with the error shown in the table
3-16 and 3-17.
Table 3-16 Capacitance Test Error of the Standard Equipment (DM306X)
Test Signal
Quick Check
Range
Error (1 year)
2nF
2.000nF
± 90 pF
20nF
20.00nF
± 300pF
200nF
Q
200.0nF
± 3nF
2μF
2.000μF
± 30nF
20μF
Q
20.00μF
± 300nF
200μF
200.0μF
± 3μF
Notes:
[1] Specifications are for 60 minutes warm–up using Math Null;
[2] Q: Optional quick test point.
Table 3-17 Capacitance Test Error of the Standard Equipment (DM305X)
Test Signal
Quick Check
Range
Error (1 year)
4nF
4.000nF
± 180pF
40nF
40.00nF
± 600pF
400nF
Q
400.0nF
± 6nF
4μF
4.000μF
± 60nF
40μF
Q
40.00μF
± 600nF
200μF
200.0μF
± 3μF
Notes:
[1] Specifications are for 60 minutes warm–up using Math Null;
[2] Q: Optional quick test point.