Specifications

RIGOL
Service Guide for DM3000 Series
3-10
Table 3-7 AC Voltage Gain Error of the Standard Equipment (DM305X)
Test
Signal
Input
Frequency
Quick Test
Point
Range
Error (1 year)
200mV
1kHz
Q
200.000mV
± 0.6mV
200mV
50kHz
200.000mV
± 4.4mV
200mV
100kHz
200.000mV
± 8.4mV
2V
1kHz
Q
2.00000V
± 6mV
2V
50kHz
2.00000V
± 22mV
2V
100kHz
2.00000V
± 44mV
20V
1kHz
Q
20.0000V
± 60mV
20V
50kHz
20.0000V
± 220mV
20V
100kHz
20.0000V
± 440V
200V
1kHz
Q
200.000V
± 600mV
200V
50kHz
200.000V
± 2.2V
200V
100kHz
200.000V
± 4.4V
750V
1kHz
Q
750.00V
± 2.25V
750V
10kHz
750.00V
± 2.25V
Notes:
[1] Set the AC filter as slow ( Meas Filter Slow );
[2] Set the reading resolution as ;
[3] Q: Optional quick test point.