Specifications
RIGOL
Service Guide for DM3000 Series
3-9
3. AC Voltage Gain Test
In AC voltage gain test, input the standard signal the test equipment outputs into
the input terminal of tested equipment; and compare the test results with the error
shown in the table 3-6 and 3-7.
Table 3-6 AC Voltage Gain Error of the standard equipment (DM306X)
Test
Signal
Input
Frequency
Quick Test
Point
Range
Error (1 year)
200mV
1kHz
Q
200.000mV
± 440μV
200mV
50kHz
200.000mV
± 460μV
200mV
300kHz
200.000mV
± 10.6mV
2V
1kHz
Q
2.00000V
± 3.6mV
2V
50kHz
2.00000V
± 9.4mV
2V
300kHz
2.00000V
±96mV
20V
1kHz
Q
20.0000V
± 36mV
20V
50kHz
20.0000V
± 94mV
20V
300kHz
20.0000V
±960mV
200V
1kHz
Q
200.000V
± 360mV
200V
50kHz
200.000V
± 940mV
200V
100kHz
200.000V
± 1.4V
750V
1kHz
Q
750.00V
± 1.35V
750V
10kHz
750.00V
± 1.35V
Notes:
[1] Set the AC filter as slow ( Meas Filter Slow );
[2] Set the reading resolution as 5½ ;
[3] Q: Optional quick test point.