Specifications

RIGOL
Service Guide for DM3000 Series
3-7
2. DC Gain Test
In DC gain test, input the standard signal output from test equipment into the input
terminal of tested equipment; and compare the test results with error shown in the
table 3-4 and 3-5.
Table 3-4 DC Gain Error of the Standard Equipment (DM306X)
Test Signal
Function
Quick Test
Point
Range
Error (1 year)
2mA
DC Current
2.000000mA
± 2.52μA
20mA
Q
20.00000mA
± 16.2μA
200mA
200.0000 mA
± 172μA
1A
1.000000A
± 1.03mA
10A
[1]
10.00000A
± 35mA
200mV
DC Voltage
Q
200.0000mV
± 31μV
2V
2.000000V
± 170μV
20V
Q
20.00000V
± 1.8mV
200V
200.0000V
± 22.4mV
1000V
[2]
1000.000V
± 120mV
200.0000Ω
2-Wire
Ohms/
4-Wire
Ohms
[3]
200.0000Ω
± 62mΩ
2.000000kΩ
2.000000kΩ
± 340mΩ
20.00000kΩ
Q
20.00000kΩ
± 3.2Ω
200.0000kΩ
Q
200.0000kΩ
± 32Ω
1.000000MΩ
1.000000MΩ
± 160Ω
10.00000MΩ
10.00000 MΩ
± 6.2kΩ
100.0000MΩ
100.0000 MΩ
± 815kΩ
Notes:
[1] Continuous current more than DC 7A or AC RMS 7A should be 30 seconds Off after 30 seconds
On;
[2] Each additional volt over ± 500 VDC adds 0.02 mV of error;
[3] Errors in above table are corresponding to 2-wire resistance function in math Null. Without
Math Null, the error is ± 0.2 Ω additional errors in 2–wire resistance function;
[4] The reading resolution of multimeter is ;
[5] Q: Optional quick test point.