Quick Guide RIGOL Publication number QGA01106-1110 July 2009 DM3000 Series Digital Multimeter DM3061/2/4 DM3051/2/4 © 2008 RIGOL Technologies, Inc.
RIGOL © 2008 RIGOL Technologies, Inc. All Rights Reserved. RIGOL products are protected by patent laws in and outside of the P.R. China. Information in this publication replaces all previous corresponding material. RIGOL reserves the right to modify or change part of or all the specifications and pricing policies at company’s sole decision. NOTE: RIGOL is registered trademark of RIGOL Technologies, Inc.
RIGOL ENGLISH Safety Notice Review the following safety precautions carefully before operating the instrument to avoid any personal injuries or damages to the instrument and any products connected to it. The instrument should be serviced by qualified personnel only. To Avoid Fire or Personal Injury. Use Proper Power Cord. Use the power cord designed for the instrument as authorized in your country only. Ground the Instrument. The instrument is grounded through the grounding conductor of the power cord.
RIGOL Quick Guide for DM3000 series III ENGLISH the maximum measurable voltage. The limitation can be expressed as 1000 Vpk. 2) LO-ground protection limitation. LO terminal can safely “float” 500 Vpk relative to the ground. The maximum protection limitation of HI terminal relative to the ground is 1000 Vpk. Therefore, the sum of the “float” voltage and the measured voltage cannot exceed 1000 Vpk.
RIGOL ENGLISH Safety Terms and Symbols Terms in This Guide. These terms may appear in this manual: ! WARNING: Warning statements indicate the conditions or practices that could result in injury or loss of life. ! CAUTION: Caution statements indicate the conditions or practices that could result in damage to this product or other property. ! CAT I (1000V) IEC Measurement Category I. The maximum voltage can be measured by HI-LO terminal is 1000Vpk. ! CAT II (300V): IEC Measurement Category II.
RIGOL ENGLISH Contents Contents .................................................................................. V Necessary Inspection ................................................................ 1 Handle Adjustment ................................................................... 4 To Connect Power Cord ............................................................. 5 To Connect Test Lead................................................................ 7 To Connect Inspection Junction Box .........
RIGOL ENGLISH General Inspection 1. Inspect the shipping container for damage. Keep the damaged shipping container or cushioning material until the contents of the shipment have been checked for completeness and the instrument has been checked mechanically and electrically. 2. Inspect the instrument. In case there is any mechanical damage or defect, notify the RIGOL Sales Representative.
RIGOL ENGLISH I. Front Panel of the Instrument DM3000 Front Panel II.
RIGOL ENGLISH A Quick Guide Inspection Junction Box** A CD-ROM* Inspection Card Extended Line** NOTE*: The CD-ROM includes《User’s Guide》and Application Software. NOTE**: The inspection junction box and inspection card extended line are the specific accessories for DM3064/DM3054, and the inspection card extended line is optional. III. Optional Accessory RS232 Cable Kelvin Test Clip NOTE: All the accessories (standard and optional) are available by contacting your local RIGOL office.
RIGOL ENGLISH Handle Adjustment To adjust the handle position of DM3000 Multimeter, please grip the handle by the sides and pull it outward.
RIGOL ENGLISH To Connect Power Cord Before you connect the instrument to a power source, please select the voltage selector according to the power supply. Then, connect the power cord as shown in the following figure. To Connect Power Cord Turn on the power switch on the rear panel, then press the power key on the front panel to start up the Multimeter.
RIGOL ENGLISH Power Selection The Multimeter can operate on multiple power distribution standards and must be set up to operate on the line voltage that will power it. If the selected line voltage does not match the power that the Multimeter will be plugged into, the Multimeter’s linevoltage setting must be changed. The power selector is under the power switch on the rear panel.
RIGOL ENGLISH To Connect Test Lead 1. DC Voltage Measurement Connect test leads as shown in the following figure; red test lead to the HI Terminal, black test lead to the LO Terminal. DC Voltage 2. AC Voltage Measurement Connect test leads as shown in the following figure; red test lead to the HI Terminal, black test lead to the LO Terminal.
RIGOL ENGLISH 3. DC Current Measurement Connect test leads as shown in the following figure; red test lead to the Current Input I Terminal, black test lead to the LO Terminal. DC Current 4. AC Current Measurement Connect test leads as shown in the following figure; red test lead to the Current Input I Terminal, black test lead to the LO Terminal.
RIGOL Connect test leads as shown in the following figure; red test lead to the HI Terminal, black test lead to the LO Terminal. I Resistance 6. 4-Wire Resistance Measurement Connect test leads as shown in the following figure; red test lead to the HI Terminal, black test lead to the LO Terminal.
RIGOL ENGLISH 7. Capacitance Measurement Connect test leads with the Capacitance as shown in the following figure; red test lead to the positive pole, black test lead to the negative pole. Capacitance 8. Continuity Testing Connect test leads as shown in the following figure; red test lead to the HI Terminal, black test lead to the LO Terminal.
RIGOL I Forward Bias 10.Frequency/Period Measurement Connect test leads as shown in the following figure; red test lead to the HI Terminal, black test lead to the LO Terminal. AC Signal Quick Guide for DM3000 series 11 ENGLISH 9. Diode Testing Connect test leads as shown in the following figure; red test lead to the HI Terminal, black test lead to the LO Terminal.
RIGOL ENGLISH 11.Sensor Measurement Sensor Signal Voltage, Resistance and Frequency Sensor Sensor Current Sensor 12 Quick Guide for DM3000 series
RIGOL ENGLISH To Connect Inspection Junction Box Model DM3064 and DM3054 have a built-in 16-channels inspection module. Users can use this module to measure multi-channels alternately and do data acquisition. Measured signals are inputted to DM3000 by connecting the inspection junction box as shown in the following figure.
RIGOL ENGLISH To Connect USB, RS232, LAN and GPIB Ports DM3000 has plenty of I/O ports. To use any of the ports, follow the next instruction.
RIGOL ENGLISH Troubleshooting 1. When press the power switch, the Multimeter has blank screen with nothing displaying: (1) Check if the power is correctly connected. (2) Check if the main power switch on the rear panel has been turned on. (3) Check if the safety fuse has been blown, replace it if necessary. (4) Restart the instrument. (5) If it still can’t work properly, please contact RIGOL for help.
RIGOL ENGLISH Contact RIGOL If you have any problem or requirement occurs when using our products, please contact RIGOL Technologies, Inc. or the local distributors. In China: Please call Tel: (86-10) 8070 6688 Fax: (86-10) 8070 5070 Service & Support Hotline: 800 810 0002 9:00 am–5: 00 pm from Monday to Friday Or by e-mail: service@rigol.com Or mail to: RIGOL Technologies, Inc.