Specifications

Chapter 4 Calibration RIGOL
Service Guide for DM3058
4-5
1. Read Testing Equipments” and “Testing Conditions”.
2. Press and enable AC voltage measure, and then separately set the range as
20kΩ and the reading resolution as Slow.
3. Press , and select T/C PSW to input password and set SecrOff ,
then select Cal Enter ;
4. Connect the voltage output of calibration instrument with HI-LO of the meter.
Then, set the output of calibration instrument as Sine, 100 mV, 1 KHz. Next,
enable the output and press Gain to start calibrating until it is stable. The results
will display in the table of UnCal G .
5. After finishing above steps, select Save to store the parameters have been
calibrated and exit.
Table 4-2 AC Calibration Input Value
Function
Range
Middle
Calibration
[1]
Full Range
Calibration
AC Voltage
200 mV
1 kHz/100 mV
1 kHz/200 mV
2 V
1 kHz/1 V
1 kHz/2 V
20 V
1 kHz/10 V
1 kHz/20 V
200 V
1 kHz/100 V
1 kHz/200 V
750 V
1 kHz/375 V
1 kHz/750 V
AC Current
20 mA
1 kHz/10 mA
1 kHz/20 mA
200 mA
1 kHz/100 mA
1 kHz/200 mA
2 A
1 kHz/1 A
1 kHz/2 A
10 A
1 kHz/5 A
1 kHz/10 A
Frequency Calibration
As for Frequency, only Gain is required to be calibrated, the procedures are:
1. Read Testing Equipments” and “Testing Conditions”.
2. Press and enable frequency measure, then set the range as 2V, Manual.
3. Press , and select T/C PSW to input password; then, set SecrOff and
select CAL Enter .
4. Connect the voltage output of calibration instrument with HI-LO of the meter.