Specifications
RIGOL Chapter 3 Performance Test
Service Guide for DM3058
3-8
Table 3-4 AC Voltage Gain Error of the standard equipment
Function
[1]
Range
Input
signal
Quick
Test
Test limits (1 year)
Upper limit
Upper limit
200.000
mV
10 mV
100 kHz
Q
10.660 mV
9.340 mV
200 mV
20 Hz
203.200 mV
196.800 mV
200 mV
45 Hz
200.500 mV
199.500 mV
200 mV
20 kHz
Q
200.500 mV
199.500 mV
200 mV
50 kHz
202.100 mV
197.900 mV
200 mV
100 kHz
206.100 mV
193.900 mV
2.00000 V
100 mV
100 kHz
0.10660 V
0.09340 V
2 V
20 Hz
Q
2.03200 V
1.96800 V
2 V
45 Hz
Q
2.00500 V
1.99500 V
2 V
20 kHz
Q
2.00500 V
1.99500 V
2 V
50 kHz
Q
2.02100 V
1.97900 V
2 V
100 kHz
Q
2.06100 V
1.93900 V
20.0000 V
1 V
100 kHz
1.0660 V
0.9340 V
20 V
20 Hz
20.3200 V
19.6800 V
20 V
45 Hz
20.0500 V
19.9500 V
20 V
20 kHz
Q
20.0500 V
19.9500 V
20 V
50 kHz
20.2100 V
19.7900 V
20 V
100 kHz
20.6100 V
19.3900 V
200.000 V
10 V
100 kHz
10.660 V
9.340 V
200 V
20 Hz
203.200 V
196.800 V
200 V
45 Hz
200.500 V
199.500 V
200 V
20 kHz
Q
200.500 V
199.500 V
200 V
50 kHz
202.100 V
197.900 V
200 V
100 kHz
206.100 V
193.900 V
750.00 V
37.5 V
100 kHz
39.98 V
35.03 V
320 V
20 Hz
325.55 V
314.45 V
320 V
45 Hz
321.02 V
318.99 V
320 V
20 kHz
Q
321.02 V
318.99 V
320 V
50 kHz
323.58 V
316.43 V
320 V
100 kHz
329.98 V
310.03 V
750 V
1 kHz
Q
751.88 V
748.13 V
Remarks:
[1] The rate of reading is “Slow”.
[2] Continuous current more than DC 7 A or AC RMS 7 A should be 30 seconds Off after 30 seconds
On.
[3] Q denotes optional quick test point.
3. Verification procedure of AC Current Gain:
(1) Read “Testing Conditions” carefully before start.
(2) Connect the input terminal of the meter with calibration instrument.