Specifications

RIGOL Chapter 3 Performance Test
Service Guide for DM3058
3-6
Zero Offset of the Standard Equipment (Continued)
Function
[1]
Range
Input
signal
Quick
Test
Test limits (1 year)
Upper
limit
Lower
limit
Resistance
[2]
200.000 Ω
short
Q
10 mΩ
-10 mΩ
2.00000 kΩ
short
60 mΩ
-60 mΩ
20.0000 kΩ
short
600 mΩ
-600 mΩ
200.000 kΩ
short
6 Ω
-6 Ω
2.00000 MΩ
short
Q
80 Ω
-80Ω
10.0000 MΩ
short
300 Ω
-300 Ω
100.000 MΩ
short
4 kΩ
-4 kΩ
Remarks:
[1] The rate of reading is Slow.
[2] Specifications are for REL operation under 4-waire or 2-wair resistance measuring. If REL
operation is disabling under 2-wair resistance measuring, 0.2 Ω additional errors are needed.
[3] Q denotes optional quick test point.
Gain Test
This test is used to inspect the accuracy of the device in full range, and which is
required only when regular offsets happened to function or range.
1. Verification procedure of DC Gain:
(1) Read Testing Conditions carefully before start.
(2) Connect the input terminal of the meter with calibration instrument.
(3) Test the functions and ranges listed in Table 3-3 item by item. And then set
the rate of reading as Slow and disable all the math operations. When
measure DC Voltage or DC Current, please shut off filter.
(4) Input signals listed in Table 3-3 using calibration instrument and compare the
test results with test limits listed. (Make sure that the calibration instrument
is in a state of output )
Table 3-3 DC Gain Errors of the Standard Equipment
Function
[1]
Range
Input
signal
Quick
Test
Test limits (1 year)
Upper limit
Lower limit
DC Voltage
200.000 mV
200 mV
200.038 mV
199.962 mV
200.000 mV
-200 mV
-199.962 mV
-200.038 mV
2.00000 V
2 V
Q
2.00036 V
1.99964 V
2.00000 V
-2 V
Q
-1.99964 V
-2.00036 V
20.0000 V
20 V
Q
20.0038 V
19.9962 V
20.0000 V
-20 V
Q
-19.9962 V
-20.0038 V
200.000 V
200 V
200.036 V
199.964 V