Explanation:
This command is invalide for Dual Display
Return Format:
The query returns the AC current in the form of 9.293791e-05, and unit is A.
15. :measure:current:AC
Command Format:
:measure:current:AC□{<range>|MIN|MAX|DEF}
Function:
The command sets the measuring range of AC current.
Parameters:
Name
Type
Range of Values
Default
<range>
discrete
{0|1|2|3}
1
Explanations:
:measure:current:AC?: Query the measuring range of AC current. Before sending
the command, please turn on AC current measuring function.
:measure:current:AC□<value>: Set the measuring range of AC current. At the
same time the manual mode is set.
The table 3-4 indicates the parameters and their meanings.
Table 3-4 The AC current parameters and their meanings
Parameter
Range
0
20 mA
1
200 mA
2
2 A
3
10 A
MIN
20 mA
MAX
10 A
DEF
200 mA
Example:
To set the measuring range of AC current as 10.0 A: