Specifications

RIGOL
Programming Guide for DM3058
5-85
The command sets or queries the NULL value of period measurement.
Parameters:
Name
Type
Range of Values
Default
<value>
Discrete
± 120% max range
0
Explanations:
[SENSe:]PERiod:NULL:VALue {<value>|MIN|MAX}: Set the NULL value of
period measurement. MIN is -120% max range of current function, MAX is max
range of current function, DEF = 0.
[SENSe:]PERiod:NULL:VALue?[{MIN|MAX}]: Query the NULL value of period
measurement.
Before sending the command, do run the command [SENSe:]PERiod:NULL
[:STATe]ON to enable NULL function.
Return Format:
The query returns the NULL value of period measurement.
60. [SENSe:]PERiod:VOLTage:RANGe:AUTO
Command Format:
[SENSe:]PERiod:VOLTage:RANGe:AUTO{ON|OFF}
[SENSe:]PERiod:VOLTage:RANGe:AUTO?
Function:
The command sets or queries the enable state of voltage autorange measuring
function in period measurement.
Explanations:
[SENSe:]PERiod:VOLTage:RANGe:AUTO{ON|OFF}: Enable or disable voltage
autorange measuring function in period measurement.
[SENSe:]PERiod:VOLTage:RANGe:AUTO?: Query the enable state of voltage
autorange measuring function in period measurement.
The max range of AC voltage is 750 V (rms), voltage RMS is determined by
waveform.
Return Format: