Specifications

RIGOL
Programming Guide for DM3058
5-83
Function:
The command sets or queries the voltage range in frequency measurement.
Parameters:
Name
Type
Range of Values
Default
<range>
Discrete
{200 mV | 2 V | 20 V
| 200 V | 750 V}
20 V
Explanations:
[SENSe:]FREQuency:VOLTage:RANGe[:UPPer]{<range>|MIN|MAX|DEF}: Set
the voltage range in frequency measurement. <range> may be 200 mV, 2 V, 20 V,
200 V, 750 V, MIN = 200 mA, MAX = 750 V. Unit is V, so 200 mA should be input
in form of 0.2.
[SENSe:]FREQuency:VOLTage:RANGe[:UPPer]? [{MIN|MAX}]: Query the
voltage range in frequency measurement.
The max range of AC voltage is 750 V (rms), voltage RMS is determined by
waveform.
Return Format:
The query returns the voltage range in frequency measurement.
57. [SENSe:]PERiod:APERture
Command Format:
[SENSe:]PERiod:APERture{<seconds>|MIN|MAX|DEF}
[SENSe:]PERiod:APERture?[{MIN|MAX}]
Function:
The command sets or queries the interval time of period measurement.
Parameters:
Name
Type
Range of Values
Default
<seconds>
Consecutive
Real Number
400 ms~2 s
400 ms
Explanations:
[SENSe:]PERiod:APERture{<seconds>|MIN|MAX|DEF}: Set the interval time