Specifications

RIGOL
Programming Guide for DM3058
5-75
[SENSe:]FRESistance:APERture:ENABled{ON|OFF}
[SENSe:]FRESistance:APERture:ENABled?
Function:
The command sets or queries the enable state of 4WR interval time measuring
function.
Explanations:
[SENSe:]FRESistance:APERture:ENABled {ON|OFF}: Enable or disable 4WR
interval time measuring function.
[SENSe:]FRESistance:APERture:ENABled?: Query the enable state of 4WR
interval time measuring function.
The commands ―CONFigure:FRESistance‖, ―MEASure:FRESistance?‖, ―[SENSe:]
FRESistance:NPLC‖ and ―[SENSe:]FRESistance:RESolution‖ can be used to
disable interval time measuring function (OFF).
2-wire and 4-wire resistance function both are influenced by the command. All
the command of 2WR and 4WR will have influence over the setting of resistance.
Return Format:
The command returns the enable state of 4WR interval time measuring function.
45. [SENSe:]FRESistance:NPLC
Command Format:
[SENSe:]FRESistance:NPLC{<PLCs>|MIN|MAX|DEF}
[SENSe:]FRESistance:NPLC?[{MIN|MAX}]
Function:
The command sets or queries the interval time of 4WR measurement, input value
and returned value are both multiples of PLC.
Parameters:
Name
Type
Range of Values
Default
<PLCs>
Discrete
{0.006|0.02|0.06|0.2|1|2|10|100}
1 PLC
Explanations:
[SENSe:]FRESistance:NPLC{<PLCs>|MIN|MAX|DEF}: Set the interval time of