Specifications

RIGOL
Programming Guide for DM3058
Return Format:
The query returns the enable state of DCV interval time measuring function.
24. [SENSe:]VOLTage[:DC]:IMPedance:AUTO
Command Format:
[SENSe:]VOLTage[:DC]:IMPedance:AUTO{ON|OFF}
[SENSe:]VOLTage[:DC]:IMPedance:AUTO?
Function:
The command sets or queries the enable state of DCV input impedance
measuring function.
Explanations:
[SENSe:]VOLTage[:DC]:IMPedance:AUTO {ON|OFF}: Enable or disable DCV
input impedance measuring function. The default is OFF. For all ranges, the
impedance of DCV is 10 MΩ.
[SENSe:]VOLTage[:DC]:IMPedance:AUTO?: Query the enable state of DCV input
impedance measuring function.
The command is available only for DCV. If DCV input impedance measuring
function is ON, for the range of 100 mV, 1 V and 10 V, input impedance will be
HI-Z (> 10 GΩ); for the range of 100 V and 1000 V, 10 MΩ.
The commands CONFigure:VOLTage:DC and MEASure:VOLTage:DC?can be
used to disable input impedance measuring function (OFF).
Return Format:
The query returns the enable state of DCV input impedance measuring function.
25. [SENSe:]VOLTage[:DC]:NPLC
Command Format:
[SENSe:]VOLTage[:DC]:NPLC{<PLCs>|MIN|MAX|DEF}
[SENSe:]VOLTage[:DC]:NPLC?[{MIN|MAX}]
Function:
The command sets or queries the interval time of DCV measurement which is