Specifications

RIGOL
Programming Guide for DM3058
Function:
The command sets or queries the enable state of ACV NULL function.
Explanations:
[SENSe:]VOLTage:AC:NULL[:STATe]{ON|OFF}: Enable or disable ACV NULL
function.
[SENSe:]VOLTage:AC:NULL[:STATe]?: Query the enable state of ACV NULL
function.
The command [SENSe:]VOLTage:AC:NULL:VALue can be used to set NULL
value.
The commands CONFigure:CURRent:AC and MEASure:CURRent:AC? can be
used to disable NULL function.
Return Format:
The command returns the enable state of ACV NULL function.
18. [SENSe:]VOLTage:AC:NULL:VALue
Command Format:
[SENSe:]VOLTage:AC:NULL:VALue{<value>|MIN|MAX}
[SENSe:]VOLTage:AC:NULL:VALue?[{MIN|MAX}]
Function:
The command sets or queries the NULL value of ACV measurement.
Parameters:
Name
Type
Range of Values
Default
<value>
Discrete
± 120% max range
0
Explanations:
[SENSe:]VOLTage:AC:NULL:VALue{<value>|MIN|MAX}: Set the NULL value of
ACV measurement. MIN is -120% max range of current function, MAX is 120%
max function of current function, DEF = 0.
[SENSe:]VOLTage:AC:NULL:VALue?[{MIN|MAX}]: Query the NULL value of
ACV measurement.
Before sending the command, do run the command [SENSe:]VOLTage:AC:NULL
[:STATe]ON to enable NULL function.