Specifications

RIGOL
Programming Guide for DM3058
5-51
Explanations:
[SENSe:]CURRent:AC:RANGe:AUTO {ON|OFF}: Enable or disable ACI
autorange function. If autorange function is ON, the measuring range will be
selected automatically according to input signal.
[SENSe:]CURRent:AC:RANGe:AUTO?: Query the enable state of ACI autorange
function.
Return Format:
The query returns the enable state of ACI autorange function.
6. [SENSe:]CURRent:AC:RANGe[:UPPer]
Command Format:
[SENSe:]CURRent:AC:RANGe[:UPPer]{<range>|MIN|MAX|DEF}
[SENSe:]CURRent:AC:RANGe[:UPPer]?[{MIN|MAX}]
Function:
The command sets or queries the range of ACI measurement.
Parameters:
Name
Type
Range of Values
Default
<range>
Discrete
{ 20 mA | 200 mA | 2 A | 10 A }
200 mA
Explanations:
[SENSe:]CURRent:AC:RANGe[:UPPer] {<range>|MIN|MAX|DEF}: Set the
range of ACI measurement. <range> may be 20 mA, 200 mA, 2 A or 10 A. MIN =
20 mA, MAX = 10 A. Unit is A, so 200 mA should be input in form of 0.2.
[SENSe:]CURRent:AC:RANGe[:UPPer]?[{MIN|MAX}]: Query the range of ACI
measurement.
Return Format:
The query returns current range of ACI measurement.
7. [SENSe:]CURRent[:DC]:APERture
Command Format:
[SENSe:]CURRent[:DC]:APERture{<seconds>|MIN|MAX|DEF}