Specifications
RIGOL
Programming Guide for DM3058
5-15
Parameters:
Name
Type
Range of Values
Default
<range>
Consecutive
Integer
20 Hz~1 MHz
switching to
autorange
<resolution>
Discrete
{MIN|MAX|DEF}
Rate: medium
Explanations:
<range> may be 20 Hz~1 MHz (unit is Hz). MAX = 1 000 000, MIN = 20, DEF is
autorange.
<resolution> (unit is Hz); in default setting, the rate is medium.
8. CONFigure:FRESistance
Command Format:
CONFigure:FRESistance□
[{<range>|AUTO|MIN|MAX|DEF} [,{<resolution>|MIN|MAX|DEF}]]
Function:
The command resets all the 4-wire resistance measuring parameters and trigger
parameters as defaults, and sets 4-wire resistance measuring function.
Parameters:
Name
Type
Range of Values
Default
<range>
Discrete
{200 Ω | 2 kΩ | 20 kΩ | 200 kΩ
| 2 MΩ | 10 MΩ | 100 MΩ}
2 kΩ
<resolution>
Discrete
{MIN|MAX|DEF}
Rate: medium
Explanations:
<range> may be 200 Ω (MIN), 2 kΩ, 20 kΩ, 200 kΩ, 2 MΩ, 10 MΩ, 100 MΩ (MAX),
AUTO (autorange) (unit is Ω). MAX = 100 000 000, MIN = 200.
<resolution> (unit is Ω); in default setting, the rate is medium.
9. CONFigure:PERiod
Command Format:
CONFigure:PERiod□[{<range>|MIN|MAX|DEF}[,{<resolution>|MIN|MAX|DEF
}]]